Comparative study of central corneal thickness measurement with slit-lamp optical coherence tomography and visante optical coherence tomography.

Haitao Li1, Christopher Kai Shun Leung, Lee Wong, Carol Yim Lui Cheung, Chi Pui Pang, Robert Neal Weinreb, Dennis Shun Chiu Lam.   

Abstract

PURPOSE: To evaluate the repeatability and reproducibility of central corneal thickness (CCT) measurements obtained by 2 anterior segment optical coherence tomography (OCT) imaging systems and to examine their agreements with ultrasound pachymetry.
DESIGN: Observational cross-sectional study. PARTICIPANTS: Fifty eyes from 50 healthy normal subjects were recruited.
METHODS: In one randomly selected eye in each subject, CCT was measured by slit-lamp OCT (SLOCT), Visante OCT, and ultrasound pachymetry. For anterior segment OCT measurements, both automatic and manual CCTs were obtained. Twenty-five of the 50 subjects were invited for 2 more visits within a week to evaluate repeatability and reproducibility of CCT measurement. MAIN OUTCOME MEASURES: Central corneal thickness measurement obtained by the 3 methods and their agreements. Intrasession and intersession within-subject standard deviation (S(w)), precision (1.96xS(w)), coefficient of variation (CV(w)) (100xS(w)/overall mean), and intraclass correlation coefficient (ICC) were calculated to evaluate repeatability and reproducibility.
RESULTS: Good repeatability and reproducibility were found for both automatic and manual CCT measurements obtained by SLOCT and Visante OCT. For intrasession repeatability, CV(w) and ICC values ranged between 0.9% and 1.2% and 0.96 and 0.98, respectively. For intersession reproducibility, the respective CV(w) and ICC values ranged between 1.2% and 1.4% and 0.94 and 0.96. Although no significant difference was found between automatic/manual SLOCT measurements and ultrasound pachymetry, automatic Visante OCT CCT (535.7+/-30.2 microm) was significantly less than CCT with ultrasound pachymetry (550.3+/-31.14 microm) (P<0.001). In contrast, manual Visante OCT measurement (558.8+/-32.8 microm) was slightly higher than ultrasound pachymetry (P<0.001). Nevertheless, SLOCT and Visante OCT measurement of CCT had 95% limits of agreement comparable to that of ultrasound pachymetry. The best agreement was observed in the manual SLOCT measurement (95% limits of agreement between -15.5 and 11.7 microm).
CONCLUSIONS: Both SLOCT and Visante OCT automatic and manual CCT measurements were reliable and showed comparable agreement with ultrasound pachymetry. Although the 2 anterior segment OCT imaging systems have similar design and working principles, clinicians should be aware of the differences in CCT measurement between the 2 anterior segment OCTs.

Entities:  

Mesh:

Year:  2007        PMID: 17916376     DOI: 10.1016/j.ophtha.2007.07.006

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  33 in total

1.  Pachymetric mapping with Fourier-domain optical coherence tomography.

Authors:  Yan Li; Maolong Tang; Xinbo Zhang; Camila H Salaroli; Jose L Ramos; David Huang
Journal:  J Cataract Refract Surg       Date:  2010-05       Impact factor: 3.351

2.  Evaluation of time domain and spectral domain optical coherence tomography in the measurement of diabetic macular edema.

Authors:  Farzin Forooghian; Catherine Cukras; Catherine B Meyerle; Emily Y Chew; Wai T Wong
Journal:  Invest Ophthalmol Vis Sci       Date:  2008-05-30       Impact factor: 4.799

3.  Intra-examiner repeatability and agreement of corneal pachymetry map measurement by time-domain and Fourier-domain optical coherence tomography.

Authors:  Jehn-Yu Huang; Melike Pekmezci; Stephanie Yaplee; Shan Lin
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-03-30       Impact factor: 3.117

4.  Repeatability and agreement of central corneal thickness measurement with non-contact methods: a comparative study.

Authors:  Tarannum Mansoori; Nagalla Balakrishna
Journal:  Int Ophthalmol       Date:  2017-04-22       Impact factor: 2.031

5.  Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter.

Authors:  Hasan Ali Bayhan; Seray Aslan Bayhan; Izzet Can
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

6.  In-vivo Three-dimensional Characteristics of Bowman's Layer and Endothelium/Descemet's Complex Using Corneal Microlayer Tomography in Healthy Subjects.

Authors:  Taher K Eleiwa; Amr Elsawy; Zeba A Syed; Vatookarn Roongpoovapatr; Ahmed M Sayed; Sonia H Yoo; Mohamed Abou Shousha
Journal:  Curr Eye Res       Date:  2020-02-16       Impact factor: 2.424

7.  Relationship between photoreceptor outer segment length and visual acuity in diabetic macular edema.

Authors:  Farzin Forooghian; Paul F Stetson; Scott A Meyer; Emily Y Chew; Wai T Wong; Catherine Cukras; Catherine B Meyerle; Frederick L Ferris
Journal:  Retina       Date:  2010-01       Impact factor: 4.256

8.  Validation of optical coherence tomography-based crystalline lens thickness measurements in children.

Authors:  Bret M Lehman; David A Berntsen; Melissa D Bailey; Karla Zadnik
Journal:  Optom Vis Sci       Date:  2009-03       Impact factor: 1.973

9.  Repeatability of layered corneal pachymetry with the artemis very high-frequency digital ultrasound arc-scanner.

Authors:  Dan Z Reinstein; Timothy J Archer; Marine Gobbe; Ronald H Silverman; D Jackson Coleman
Journal:  J Refract Surg       Date:  2009-11-16       Impact factor: 3.573

10.  Comparison of central corneal thickness and anterior chamber depth measurements using three imaging technologies in normal eyes and after phakic intraocular lens implantation.

Authors:  Muriël Doors; Lars P J Cruysberg; Tos T J M Berendschot; John de Brabander; Frenne Verbakel; Carroll A B Webers; Rudy M M A Nuijts
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-04-30       Impact factor: 3.117

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.