Corneal thickness measurements with the Orbscan Topography System and ultrasonic pachymetry.

V Yaylali1, S C Kaufman, H W Thompson.   

Abstract

PURPOSE: To compare corneal thickness measurements obtained with a new instrument, the Orbscan Topography System, with those obtained with the DGH ultrasonic pachymeter and to assess the agreement and repeatability of the two devices.
SETTING: LSU Eye Center, New Orleans, Louisiana, USA.
METHODS: Measurement agreement was assessed in 51 eyes of 26 normal volunteers using both Orbscan and ultrasonic pachymetry. Repeatability for the instruments was measured in 10 eyes of 5 additional volunteers. Corneal thicknesses were compared using the analysis of variance (ANOVA). The relationship between the devices was assessed by analysis of regression (ANOR).
RESULTS: In the measurement agreement experiment, the mean corneal thickness was 571.3 microm +/- 6.21 SEM with the Orbscan system and 543.3 +/- 7.49 microm with ultrasonic pachymetry; these values were significantly different (F test, ANOVA, P = .0048). In the repeatability experiment, the mean thickness was 561.1 +/- 8.42 microm with the Orbscan system and 537.4 +/- 5.84 microm with ultrasound pachymetry; these values were also significantly different (F test, ANOVA, P = .0003). Analysis of regression showed a significant linear regression between the values obtained with the devices (P = .0001, F test, ANOR).
CONCLUSIONS: In both studies, the Orbscan system obtained statistically significantly different and higher values for corneal thickness. Regression analysis suggests that over the range of values in this study, the two devices differ by a constant amount (intercept and slope). The nonzero intercept of this regression shows that the values from the devices differ and cannot be directly substituted for each other. We therefore conclude that in this study, Orbscan system measurements of corneal thickness were 23 to 28 microm greater than ultrasonic pachymeter measurements. Linear regression equations may be developed for the results of measurements from the two devices and used as a precise transformation factor for the values obtained with the two devices.

Entities:  

Mesh:

Year:  1997        PMID: 9423906     DOI: 10.1016/s0886-3350(97)80113-7

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  27 in total

1.  Evaluation of corneal thickness and topography in normal eyes using the Orbscan corneal topography system.

Authors:  Z Liu; A J Huang; S C Pflugfelder
Journal:  Br J Ophthalmol       Date:  1999-07       Impact factor: 4.638

2.  Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness.

Authors:  S Miglior; E Albe; M Guareschi; G Mandelli; S Gomarasca; N Orzalesi
Journal:  Br J Ophthalmol       Date:  2004-02       Impact factor: 4.638

3.  Posterior corneal topographic changes after partial flap during laser in situ keratomileusis.

Authors:  N Sharma; A Rani; R Balasubramanya; R B Vajpayee; R M Pandey
Journal:  Br J Ophthalmol       Date:  2003-02       Impact factor: 4.638

4.  Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis.

Authors:  K Kawana; T Tokunaga; K Miyata; F Okamoto; T Kiuchi; T Oshika
Journal:  Br J Ophthalmol       Date:  2004-04       Impact factor: 4.638

5.  The effects of corneal parameters on the assessment of endothelial cell density in the elderly eye.

Authors:  A Müller; J P Craig; C N Grupcheva; C N J McGhee
Journal:  Br J Ophthalmol       Date:  2004-03       Impact factor: 4.638

6.  Comparison of ultrasonic pachymetry with Orbscan in corneal haze.

Authors:  Rana Altan-Yaycioglu; Aysel Pelit; Yonca Aydin Akova
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2007-06-22       Impact factor: 3.117

7.  Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry.

Authors:  Hani S Al-Mezaine; Saleh A Al-Amro; Dustan Kangave; Abdulkareem Sadaawy; Taher A Wehaib; Saleh Al-Obeidan
Journal:  Int Ophthalmol       Date:  2007-09-26       Impact factor: 2.031

Review 8.  [Corneal topography. Analysis of the corneal surface].

Authors:  C Herrmann; U Ludwig; G Duncker
Journal:  Ophthalmologe       Date:  2008-02       Impact factor: 1.059

9.  Corneal pachymetry mapping with high-speed optical coherence tomography.

Authors:  Yan Li; Raj Shekhar; David Huang
Journal:  Ophthalmology       Date:  2006-05       Impact factor: 12.079

10.  Correlations between magnitude of refractive error and other optical components in Korean myopes.

Authors:  Sangkyu Lee; Boyun Kim; Tae-Hoon Oh; Hyun Seung Kim
Journal:  Korean J Ophthalmol       Date:  2012-09-24
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.