Agreement among 3 methods to measure corneal thickness: ultrasound pachymetry, Orbscan II, and Visante anterior segment optical coherence tomography.

Emmy Y M Li1, Shaheeda Mohamed, Christopher K S Leung, Srinivas K Rao, Arthur C K Cheng, Carol Y L Cheung, Dennis S C Lam.   

Abstract

PURPOSE: To assess the agreement of central and paracentral corneal thickness measurements between ultrasound pachymetry (USP), Orbscan II, and Visante anterior segment optical coherence tomography (ASOCT).
DESIGN: Observational cross-sectional study. PARTICIPANTS: Seventy eyes of 70 subjects.
METHODS: Each subject underwent Orbscan II (using an acoustic equivalent correction factor of 0.89), ASOCT, and USP examination. Bland-Altman plots were used to evaluate agreement between instruments. MAIN OUTCOME MEASURES: Central and paracentral corneal thickness measurements by the 3 methods and agreement, as evaluated by 95% limits of agreement (LOA).
RESULTS: The mean measurements of average central corneal thickness by USP, Orbscan II, and ASOCT were 553.5+/-30.26 microm, 553.22+/-25.47 microm, and 538.79+/-26.22 microm, respectively. There was high correlation between instruments: USP with ASOCT (r = 0.936, P<0.001), USP with Orbscan II (r = 0.900, P<0.001) for central corneal thickness measurements, and Orbscan II with ASOCT for average paracentral 2- to 5-mm measurements (r = 0.947, P<0.001). The mean differences (and upper/lower LOA) for central corneal thickness measurements were 0.31+/-13.34 microm (26.44/-25.83) between USP and Orbscan II, 14.74+/-10.84 microm (36.0/-6.51) between USP and ASOCT, and 14.44+/-9.14 microm (32.36/-3.48) between Orbscan II and ASOCT. The average mean difference (and upper/lower LOA) between Orbscan II and ASOCT for paracentral 2- to 5-mm corneal thickness measurements was 10.35+/-8.67 microm (27.35/-6.65).
CONCLUSION: Anterior segment optical coherence tomography underestimated corneal thickness compared with that measured with USP. Anterior segment optical coherence tomography had better agreement with the gold standard USP, as compared with Orbscan II. However, important discrepancies among instruments exist. Clinicians should be aware that corneal thickness measurements are influenced by the method of measurement and that, although highly correlated, these instruments should not be used interchangeably for the assessment of corneal thickness.

Entities:  

Mesh:

Year:  2007        PMID: 17507097     DOI: 10.1016/j.ophtha.2007.02.017

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  25 in total

1.  Pachymetric mapping with Fourier-domain optical coherence tomography.

Authors:  Yan Li; Maolong Tang; Xinbo Zhang; Camila H Salaroli; Jose L Ramos; David Huang
Journal:  J Cataract Refract Surg       Date:  2010-05       Impact factor: 3.351

2.  Measurements of anterior segment parameters using three different non-contact optical devices in keratoconus patients.

Authors:  Ahmet Taylan Yazıcı; Gökhan Pekel; Ercüment Bozkurt; Yusuf Yıldırım; Evre Pekel; Ahmet Demirok; Omer Faruk Yılmaz
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

3.  Meta-analysis of Pentacam vs. ultrasound pachymetry in central corneal thickness measurement in normal, post-LASIK or PRK, and keratoconic or keratoconus-suspect eyes.

Authors:  Wenjing Wu; Yan Wang; Lulu Xu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-11-12       Impact factor: 3.117

4.  Difference in normal corneal thickness and curvature between Mongolian and Han nationalities.

Authors:  Hai-Xia Zhao; Li Zhang; Wen-Ying Guan
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

5.  Central corneal thickness determination in corneal edema using ultrasound pachymetry, a Scheimpflug camera, and anterior segment OCT.

Authors:  David Kuerten; Niklas Plange; Eva Charlotte Koch; Antonis Koutsonas; Peter Walter; Matthias Fuest
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-04-21       Impact factor: 3.117

6.  Intra-examiner repeatability and agreement of corneal pachymetry map measurement by time-domain and Fourier-domain optical coherence tomography.

Authors:  Jehn-Yu Huang; Melike Pekmezci; Stephanie Yaplee; Shan Lin
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-03-30       Impact factor: 3.117

7.  [Difference and distance between the central and thinnest points of the cornea: impact of refractive state, age and ocular side].

Authors:  J Steinberg; C Kohl; T Katz; G Richard; S J Linke
Journal:  Ophthalmologe       Date:  2014-04       Impact factor: 1.059

8.  Repeatability and agreement of central corneal thickness measurement with non-contact methods: a comparative study.

Authors:  Tarannum Mansoori; Nagalla Balakrishna
Journal:  Int Ophthalmol       Date:  2017-04-22       Impact factor: 2.031

9.  Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter.

Authors:  Hasan Ali Bayhan; Seray Aslan Bayhan; Izzet Can
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

10.  Comparison of central corneal thickness and anterior chamber depth measurements using three imaging technologies in normal eyes and after phakic intraocular lens implantation.

Authors:  Muriël Doors; Lars P J Cruysberg; Tos T J M Berendschot; John de Brabander; Frenne Verbakel; Carroll A B Webers; Rudy M M A Nuijts
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-04-30       Impact factor: 3.117

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.