Marcus Ang1, Wesley Chong, Huiqi Huang, Wan Ting Tay, Tien Yin Wong, Ming-Guang He, Tin Aung, Jodhbir S Mehta.
Abstract
OBJECTIVE: To compare anterior segment parameters measured using a semi-automatic software (Zhongshan Angle Assessment Program, ZAP) applied to anterior segment optical coherence tomography (AS-OCT) images, with commonly used instruments.Entities:
Mesh:
Year: 2013 PMID: 23750265 PMCID: PMC3672133 DOI: 10.1371/journal.pone.0065559
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Characteristics of anterior segment parameters in study population.
| Characteristics | Sub-category |
| ACD (mm) | CCT (µm) | K-reading (D) | |||||||||||
| AS-OCT (ZAP) | IOLMaster | AS-OCT (ZAP) | Ultrasound Pachymetry | AS-OCT (ZAP) | Auto Refractor | |||||||||||
| Mean | (SD) | Mean | (SD) | Mean | (SD) | Mean | (SD) | Mean | (SD) | Mean | (SD) | |||||
| All persons | 1069 | 3.29 | (0.35) | 3.13 | (0.38) | 560.75 | (35.07) | 544.00 | (35.03) | 46.79 | (2.72) | 43.98 | (1.53) | |||
| Age | 40–49 | 301 | 3.42 | (0.33) | 3.27 | (0.37) | 566.58 | (37.31) | 552.81 | (37.48) | 46.45 | (2.26) | 43.92 | (1.49) | ||
| 50–59 | 363 | 3.31 | (0.34) | 3.17 | (0.35) | 561.79 | (34.39) | 544.94 | (34.20) | 46.50 | (2.36) | 43.91 | (1.50) | |||
| 60–69 | 281 | 3.18 | (0.36) | 3.02 | (0.37) | 558.30 | (33.36) | 538.81 | (32.58) | 47.21 | (3.05) | 44.04 | (1.56) | |||
| ≥70 | 124 | 3.14 | (0.31) | 2.93 | (0.36) | 549.10 | (32.15) | 531.60 | (31.02) | 47.54 | (3.57) | 44.20 | (1.62) | |||
|
| <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | 0.255 | ||||||||||
| Gender | Male | 544 | 3.35 | (0.35) | 3.19 | (0.38) | 559.22 | (34.70) | 542.78 | (35.43) | 46.48 | (2.55) | 43.69 | (1.50) | ||
| Female | 525 | 3.22 | (0.34) | 3.07 | (0.37) | 562.33 | (35.42) | 545.27 | (34.60) | 47.12 | (2.86) | 44.28 | (1.50) | |||
|
| <0.001 | <0.001 | 0.147 | 0.245 | <0.001 | <0.001 | ||||||||||
| Race | Chinese | 320 | 3.27 | (0.33) | 3.22 | (0.34) | 567.85 | (33.43) | 552.82 | (35.14) | 47.32 | (2.14) | 43.80 | (1.54) | ||
| Indian | 496 | 3.28 | (0.37) | 3.07 | (0.40) | 561.81 | (34.00) | 538.83 | (32.96) | 47.31 | (2.86) | 44.11 | (1.54) | |||
| Malay | 253 | 3.32 | (0.34) | 3.13 | (0.37) | 549.69 | (36.58) | 542.98 | (36.77) | 45.11 | (2.42) | 43.95 | (1.47) | |||
|
| 0.291 | <0.001 | <0.001 | <0.001 | <0.001 | 0.017 | ||||||||||
based on analysis of variance or independent-samples t-test where appropriate.
ACD: anterior chamber depth; CCT: central corneal thickness; K: keratometry; D: diopters; AS-OCT: anterior segment optical coherence tomography; ZAP: Zhongshan Angle Assessment Program; SD: standard deviation; n: number of subjects.
Comparison of anterior segment parameters measured by anterior segment optical coherence tomography and various modalities.
| Anterior segment parameters | Mean | Difference | Bias |
| 95% LOA | 95% CI of LOA | |||||
| Mean | (SD) | Lower LOA | Upper LOA | ||||||||
| ACD | 3.21 | 0.16 | (0.20) | 4.88% | <0.001 | (−0.24, | 0.55) | (−0.26, | −0.22) | (0.53, | 0.57) |
| CCT (µm) | 552.37 | 16.75 | (17.35) | 3.03% | <0.001 | (−17.26, | 50.76) | (−19.06, | −15.46) | (48.96, | 52.56) |
| K-reading (D) | 45.39 | 2.81 | (2.36) | 6.19% | <0.001 | (−1.81, | 7.43) | (−2.06, | −1.57) | (7.19, | 7.68) |
based on paired-samples t-test.
ACD measurements by ZAP have CCT measurements by ZAP added to allow comparison with ACD measurements by IOLMaster.
ACD: anterior chamber depth; CCT: central corneal thickness; K: keratometry; D: diopters; SD: standard deviation; LOA: limits of agreement; CI: confidence interval; ZAP: Zhongshan Angle Assessment Program.
Figure 1Bland-Altman plot of anterior chamber depth (ACD) measurements, with zero-line (blue line), mean difference (red line) and 95% limits of agreement (green dotted lines), comparing AS-OCT (ZAP) and IOLMaster.
X-axis units = mm/Y-axis units = mm.
Figure 2Bland-Altman plot of central corneal thickness (CCT) measurements, with zero-line (blue line), mean difference (red line) and 95% limits of agreement (green dotted lines), comparing AS-OCT (ZAP) and Ultrasound Pachymetry.
X-axis units = µm/Y-axis units = µm.
Figure 3Bland-Altman plot of keratometric (K)-readings with the difference between the K-readings of the two methods regressed on the average of K-readings with its 95% prediction interval.
X-axis units = D/Y-axis units = D.
Correlations of anterior segment parameters measured by anterior segment optical coherence tomography and various modalities.
| Anterior segment parameters | Measurement modality | Mean | (SD) | Pearson Correlation |
|
| ACD | AS-OCT (ZAP) | 3.29 | (0.35) | 0.85 | <0.001 |
| IOL Master | 3.13 | (0.38) | |||
| CCT (µm) | AS-OCT (ZAP) | 560.75 | (35.07) | 0.88 | <0.001 |
| Ultrasound Pachymetry | 544 | (35.03) | |||
| K-reading (D) | AS-OCT (ZAP) | 46.79 | (2.72) | 0.50 | <0.001 |
| Auto Refractor | 43.98 | (1.53) |
ACD measurements by ZAP have CCT measurements by ZAP added to allow comparison with ACD measurements by IOLMaster.
ACD: anterior chamber depth; AS-OCT: anterior segment optical coherence tomography; ZAP: Zhongshan Angle Assessment Program; CCT: central corneal thickness; K: keratometry; D: diopters; SD: standard deviation.