PURPOSE: The purpose of the study was to evaluate the agreement between measurements by optical low-coherence interferometry (OLCI, Aladdin) and those by Scheimpflug imaging combined with partial coherence interferometry (Scheimpflug-PCI, Pentacam AXL) in cataract patients.Entities:
Keywords: Biometric parameters; Pentacam AXL; cataract surgery; optical low-coherence interferometry
Year: 2022 PMID: 36211306 PMCID: PMC9535919 DOI: 10.4103/sjopt.sjopt_50_21
Source DB: PubMed Journal: Saudi J Ophthalmol ISSN: 1319-4534
Mean axial length, anterior chamber depth, and K measured by optical low-coherence interferometry and Scheimpflug imaging combined with partial coherence interferometry
| Parameter | OLCI | Scheimpflug-PCI | Difference |
| 95% LoA | CC |
|---|---|---|---|---|---|---|
| AL (mm) | ||||||
| Mean±SD | 23.25±0.85 | 23.23±0.86 | +0.02 | <0.0001* | −0.05-0.10 | 0.9990 (<0.0001) |
| Range | 21.43-26.03 | 21.37-26.05 | ||||
| 95% CI | 23.09-23.41 | 23.06-23.39 | ||||
| ACD (mm) | ||||||
| Mean±SD | 3.06±0.37 | 3.15±0.44 | −0.09 | 0.0003* | −0.68-0.50 | 0.7386 (<0.0001) |
| Range | 1.73-3.90 | 2.07-4.89 | ||||
| 95% CI | 3.00-3.12 | 3.08-3.22 | ||||
| Mean K (D) | ||||||
| Mean±SD | 44.24±1.51 | 44.29±1.48 | −0.05 | 0.1074 | −0.87-0.76 | 0.9614 (<0.0001) |
| Range | 40.98-48.01 | 40.35-47.95 | ||||
| 95% CI | 44.00-44.47 | 44.06-44.52 | ||||
| Flat K (D) | ||||||
| Mean±SD | 43.59±1.60 | 43.67±1.59 | −0.08 | 0.0428* | −1.13-0.96 | 0.9445 (<0.0001) |
| Range | 38.91-47.43 | 37.70-47.60 | ||||
| 95% CI | 43.34-43.83 | 43.42-43.92 | ||||
| Steep K (D) | ||||||
| Mean±SD | 44.89±1.56 | 44.91±1.51 | −0.02 | 0.5845 | −0.94-0.90 | 0.9535 (<0.0001) |
| Range | 41.30-48.74 | 41.30-48.74 | ||||
| 95% CI | 44.65-45.13 | 44.68-45.14 | ||||
*Statistically significant. AL: Axial length, ACD: Anterior chamber depth, K: Keratometry, OLCI: Optical low-coherence interferometry, Scheimpflug-PCI: Scheimpflug imaging combined with partial coherence interferometry, LoA: Limits of agreement, CC: Correlation of coefficient, SD: Standard deviation, CI: Confidence interval, D: Diopters
Figure 1Bland–Altman plot for the mean axial length measured by the optical low-coherence interferometry and Scheimpflug-partial coherence interferometry devices (AL = axial length; OLCI = optical low-coherence interferometry; Scheimpflug-PCI = Scheimpflug imaging combined with partial coherence interferometry; SD = standard deviation)
Figure 2Bland–Altman plot for the mean anterior chamber depth measured by the optical low-coherence interferometry and Scheimpflug-partial coherence interferometry devices (ACD = anterior chamber depth; OLCI = optical low-coherence interferometry; Scheimpflug-PCI = Scheimpflug imaging combined with partial coherence interferometry; SD = standard deviation)
Figure 3Bland–Altman plot for the mean, steep, and flat keratometric values measured by the optical low-coherence interferometry and Scheimpflug-partial coherence interferometry devices (K = keratometry; OLCI = optical low-coherence interferometry; Scheimpflug-PCI = Scheimpflug imaging combined with partial coherence interferometry; SD = standard deviation)
Corneal astigmatism measured by optical low-coherence interferometry and Scheimpflug imaging combined with partial coherence interferometry
| Parameter | OLCI | Scheimpflug-PCI | Difference |
| 95% LoA | CC |
|---|---|---|---|---|---|---|
| Astigmatism magnitude (D) | ||||||
| Mean±SD | 1.30±0.96 | 1.24±0.92 | +0.06 | 0.1441 | −0.02–0.15 | 0.8210 (<0.0001) |
| Range | 0-5.48 | 0-5.60 | ||||
| 95% CI | 1.15-1.45 | 1.10-1.38 | ||||
| J0 | ||||||
| Mean±SD | −0.22±0.68 | −0.14±0.66 | −0.08 | 0.0087* | −0.13-0.02 | 0.8481 (<0.0001) |
| Range | −2.66-2.07 | −2.77–1.84 | ||||
| 95% CI | −0.32-−0.11 | −0.24-−0.04 | ||||
| J45 | ||||||
| Mean±SD | 0.06±0.39 | 0.04±0.38 | 0.02 | 0.4147 | −0.02-0.05 | 0.7905 (<0.0001) |
| Range | −0.79-2.11 | −0.97-2.09 | ||||
| 95% CI | 0–0.12 | −0.02-0.10 | ||||
*Statistically significant. OLCI: Optical low-coherence interferometry, Scheimpflug-PCI: Scheimpflug imaging combined with partial coherence interferometry, LoA: Limits of agreement, CC: Correlation of coefficient, SD: Standard deviation, CI: Confidence interval, D: Diopters, J0: Jackson cross-cylinder, axes at 90° and 180°, J45: Jackson cross-cylinder, axes at 45° and 135°