Comparing immersion ultrasound with partial coherence interferometry for intraocular lens power calculation.

Julio Narváez1, D Hunter Cherwek, R Doyle Stulting, Rhonda Waldron, Grenith J Zimmerman, Izak F Wessels, George O Waring.   

Abstract

BACKGROUND AND
OBJECTIVE: To compare the utility and accuracy of high-precision immersion ultrasound, partial coherence interferometry, and the IOL Master system (Carl Zeiss Meditec, Dublin, CA) in a tertiary care referral center. PATIENTS AND METHODS: Comparative clinical study in a tertiary care, multispecialty, university practice. Participants were consecutive patients scheduled for cataract surgery at Emory Eye Center. Patients underwent biometry using immersion ultrasound, partial coherence interferometry, and the IOL Master system.
RESULTS: Twenty-two percent of eyes undergoing surgery could not be measured using interferometry. In those eyes measurable by both methods, there was no difference in measurements nor postoperative refractive outcome. Each method had a high repeatability (intraclass correlation coefficient > 0.99) with a high intraclass correlation between methods (intraclass correlation coefficient = 0.99).
CONCLUSIONS: When it can be used, interferometry is equivalent to immersion ultrasonography regarding biometric accuracy and precision. However, it cannot replace ultrasonography, especially for eyes with dense media opacity.

Entities:  

Mesh:

Year:  2008        PMID: 18254348     DOI: 10.3928/15428877-20080101-08

Source DB:  PubMed          Journal:  Ophthalmic Surg Lasers Imaging        ISSN: 1542-8877


  11 in total

1.  Comparison of optical low-coherence reflectometry and applanation ultrasound biometry on intraocular lens power calculation.

Authors:  Mirjana Bjeloš Rončević; Mladen Bušić; Ivan Cima; Biljana Kuzmanović Elabjer; Damir Bosnar; Daliborka Miletić
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-09-18       Impact factor: 3.117

2.  Intraobserver and interobserver repeatability of ocular components measurement in cataract eyes using a new optical low coherence reflectometer.

Authors:  Mirjana Bjeloš Rončević; Mladen Bušić; Ivan Cima; Biljana Kuzmanović Elabjer; Damir Bosnar; Daliborka Miletić
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-10-28       Impact factor: 3.117

3.  Refractive outcomes comparison between the Lenstar LS 900® optical biometry and immersion A-scan ultrasound.

Authors:  Palanyraj Naicker; Siva Sundralingam; Mohammadreza Peyman; Azida Juana; Nor Fadhilah Mohamad; Maung Maung Win; Angela Loo; Visvaraja Subrayan
Journal:  Int Ophthalmol       Date:  2014-07-15       Impact factor: 2.031

4.  Partial coherence interferometry versus immersion ultrasonography for axial length measurement in children.

Authors:  Phoebe D Lenhart; Amy K Hutchinson; Michael J Lynn; Scott R Lambert
Journal:  J Cataract Refract Surg       Date:  2010-12       Impact factor: 3.351

5.  Comparison of intraocular lens power prediction using immersion ultrasound and optical biometry with and without formula optimization.

Authors:  Gabor Nemeth; Attila Nagy; Andras Berta; Laszlo Modis
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-04-13       Impact factor: 3.117

6.  Agreement of IOL power and axial length obtained by IOLMaster 500 vs IOLMaster 500 with Sonolink connection.

Authors:  Sabong Srivannaboon; Chareenun Chirapapaisan; Patchara Nantasri; Mathinee Chongchareon; Pratuangsri Chonpimai
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-01-12       Impact factor: 3.117

7.  Comparison of anterior segment measurements obtained by three different devices in healthy eyes.

Authors:  Carmen Lopez de la Fuente; Ana Sanchez-Cano; Francisco Segura; Isabel Pinilla
Journal:  Biomed Res Int       Date:  2014-06-02       Impact factor: 3.411

8.  Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts.

Authors:  Jinhai Huang; Na Liao; Giacomo Savini; Fangjun Bao; Ye Yu; Weicong Lu; Qingjie Hu; Qinmei Wang
Journal:  J Ophthalmol       Date:  2014-10-23       Impact factor: 1.909

9.  Comparison of two optical biometers in intraocular lens power calculation.

Authors:  Sheng Hui; Lu Yi
Journal:  Indian J Ophthalmol       Date:  2014-09       Impact factor: 1.848

10.  Accuracy of optical biometry combined with Placido disc corneal topography for intraocular lens power calculation.

Authors:  Giacomo Savini; Kenneth J Hoffer; Piero Barboni; Nicole Balducci; Domenico Schiano-Lomoriello; Pietro Ducoli
Journal:  PLoS One       Date:  2017-02-23       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.