Jorge Selem Haddad1,2, Eliza Barnwell1, Karolinne Maia Rocha1, Renato Ambrosio2,3, George O Waring Iv4.
Abstract
PURPOSE: To compare biometry measurements obtained by a partial interferometer biometer (IOLMaster 500) to the new Scheimpflug tomography with an integrated axial length biometer module (Pentacam AXL). PATIENTS AND METHODS: Cataract patients who underwent biometric measurements with the IOL Master 500 and the Pentacam AXL from July to November 2017 were enrolled in this study. Comparisons were performed for axial length (AL), keratometry (K), and anterior chamber depth (ACD). The Pearson correlation coefficient and the 95% limits of agreement (LoA) were calculated. Paired Student's t-tests and Bland-Altman plots were used to assess the differences between devices.Entities:
Keywords: IOLMaster 500; anterior chamber depth; axial length; biometry; partial coherence interferometry; pentacam AXL
Year: 2020 PMID: 32099321 PMCID: PMC7007795 DOI: 10.2147/OPTH.S238112
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Means and Standard Deviations of the Differences
| Measure | Pentacam | IOL Master | Mean Difference (95% CI) | Bias (95% Predictive Interval) | p valuea |
|---|---|---|---|---|---|
| K mean | 43.75 ± 2.09 (n=166) | 43.85 ± 2.16 (n=166) | −0.11 [−0.15; −0.06] | −0.11 [−0.69; 0.48] | <0.001 |
| K flat | 43.34 ± 2.08 (n=164) | 43.40 ± 2.18 (n=164) | −0.06 [−0.12; −0.01] | −0.06 [−0.81; 0.68] | 0.028 |
| K steep | 44.18 ± 2.15 (n=166) | 44.35 ± 2.20 (n=166) | −0.17 [−0.22; −0.11] | −0.17 [−0.88; 0.54] | <0.001 |
| AL | 24.03 ± 1.37 (n=163) | 24.03 ± 1.37 (n=163) | −0.002 [−0.009; 0.005] | −0.002 [−0.09; 0.086] | 0.558 |
| ACD | 3.24 ± 0.41 (n=166) | 3.19 ± 0.44 (n=166) | 0.05 [0.03; 0.07] | 0.05 [−0.25; 0.35] | <0.001 |
Note: aPaired Student’s t-test.
Abbreviations: AL, axial length; ACD, anterior chamber depth; CI, confidence interval.
Figure 1Bland-Altman plot for axial length according to Pentacam and IOL Master measurements.
Figure 2Bland-Altman plot for keratometry flat according to Pentacam and IOL Master measurements.
Figure 3Bland-Altman plot for keratometry mean according to Pentacam and IOL Master measurements.
Figure 5Bland-Altman plot for anterior chamber depth according to Pentacam and IOL Master measurements.