Repeatability and reproducibility of corneal curvature measurements using the Pentacam and Keratron topography systems.

Takushi Kawamorita1, Nanami Nakayama, Hiroshi Uozato.   

Abstract

PURPOSE: To compare the repeatability and reproducibility of corneal curvature measurements using the Pentacam eye scanner (Oculus Optikgeräte GmbH) and Keratron corneal topographer (Optikon 2000 SpA).
METHODS: Axial topography maps were used to acquire measurements. Twenty-six eyes of 26 healthy patients were measured to determine repeatability and reproducibility. Another 10 eyes of 10 healthy patients were included in a parallel study. Three measurements per eye were performed. Repeatability was assessed via the coefficient of variation. Reproducibility was assessed using Bland-Altman plots. Linear correlations were used to determine the agreement between devices. A P value < .05 was considered statistically significant.
RESULTS: The majority of coefficient of variations for both devices were within 1%. The coefficient of variation of the Pentacam was higher in the superior cornea (P < .01). The mean difference (95% limits of agreement) in the flattest meridian between examiners was -0.03 +/- 0.27 diopters (D) (range: -0.56 to +0.49 D) for the Pentacam and -0.08 +/- 0.21 D (range: -0.50 to +0.33 D) for the Keratron. The mean difference (95% limits of agreement) in the steepest meridian was -0.10 +/- 0.26 D (range: -0.60 to +0.41 D) for the Pentacam and -0.11 +/-0.22 D (range: -0.53 to +0.31 D) for the Keratron. The mean axial power for the central 3.0 mm of the Pentacam was statistically significantly lower than that of the Keratron (P < .01).
CONCLUSIONS: Although statistically significant differences were noted, both devices provided repeatable and reproducible corneal measurements centrally. Pentacam repeatability outcomes indicate that superior corneal measurements should be interpreted with caution.

Entities:  

Mesh:

Year:  2009        PMID: 19603622     DOI: 10.3928/1081597X-20090512-08

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  12 in total

1.  Keratometry with five different techniques: a study of device repeatability and inter-device agreement.

Authors:  Shiva Mehravaran; Soheila Asgari; Sara Bigdeli; Ashkan Shahnazi; Hassan Hashemi
Journal:  Int Ophthalmol       Date:  2014-02-23       Impact factor: 2.031

2.  Scheimpflug camera combined with placido-disk corneal topography and optical biometry for intraocular lens power calculation.

Authors:  Ahmet Kirgiz; Kurşat Atalay; Havva Kaldirim; Kubra Serefoglu Cabuk; Mehmet Orcun Akdemir; Muhittin Taskapili
Journal:  Int Ophthalmol       Date:  2016-09-01       Impact factor: 2.031

3.  Characteristics of Pupil Offset in Young Asian Adults With Mild-Moderate and High Myopia.

Authors:  Shengshu Sun; Zhanglin Liu; Yuan Wu; Xiaowen Sun; Shaozhen Zhao; Yue Huang
Journal:  Transl Vis Sci Technol       Date:  2022-06-01       Impact factor: 3.048

4.  Comparison of two Nd:YAG laser posterior capsulotomy: cruciate pattern vs circular pattern with vitreous strand cutting.

Authors:  Jin-Soo Kim; Jung Yeol Choi; Ji-Won Kwon; Won Ryang Wee; Young Keun Han
Journal:  Int J Ophthalmol       Date:  2018-02-18       Impact factor: 1.779

5.  Precision and agreement of corneal power measurements obtained using a new corneal topographer OphthaTOP.

Authors:  Jinhai Huang; Giacomo Savini; Hao Chen; Fangjun Bao; Yuanguang Li; Haisi Chen; Weicong Lu; Ye Yu; Qinmei Wang
Journal:  PLoS One       Date:  2015-01-05       Impact factor: 3.240

6.  Repeated Same-Day Versus Single Tomography Measurements of Keratoconic Eyes for Analysis of Disease Progression.

Authors:  Wolf Wonneberger; Bertil Sterner; Ulrika MacLean; Margareta Claesson; Madeleine Zetterberg
Journal:  Cornea       Date:  2018-04       Impact factor: 2.651

7.  Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability.

Authors:  David P Piñero; Alberto López-Navarro; Inmaculada Cabezos; Dolores de Fez; María T Caballero; Vicent J Camps
Journal:  J Ophthalmol       Date:  2017-01-04       Impact factor: 1.909

8.  The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging.

Authors:  Yune Zhao; Ding Chen; Giacomo Savini; Qing Wang; Hongfang Zhang; Yili Jin; Benhao Song; Rui Ning; Jinhai Huang; Chenyang Mei
Journal:  Eye Vis (Lond)       Date:  2020-06-09

9.  Comparison of Anterior Segment Biometric Measurements between Pentacam HR and IOLMaster in Normal and High Myopic Eyes.

Authors:  Jing Dong; Maolong Tang; Yaqin Zhang; Yading Jia; Haining Zhang; Zhijie Jia; Xiaogang Wang
Journal:  PLoS One       Date:  2015-11-17       Impact factor: 3.240

10.  Algorithm for Correcting the Keratometric Error in the Estimation of the Corneal Power in Keratoconus Eyes after Accelerated Corneal Collagen Crosslinking.

Authors:  David P Piñero; Vicente J Camps; Esteban Caravaca-Arens; Dolores de Fez; Francisco J Blanes-Mompó
Journal:  J Ophthalmol       Date:  2017-10-22       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.