Comparability and repeatability of corneal astigmatism measurements using different measurement technologies.

Nienke Visser1, Tos T J M Berendschot, Frenne Verbakel, John de Brabander, Rudy M M A Nuijts.   

Abstract

PURPOSE: To determine the comparability and repeatability of corneal astigmatism measurements obtained with different devices and determine the interobserver variability of a new automated keratometer.
SETTING: University Eye Clinic Maastricht, the Netherlands.
DESIGN: Prospective cohort study.
METHODS: The right eye of healthy subjects was examined with the following 6 devices: IOLMaster (automated keratometry), Lenstar (automated keratometry), SMI Reference Unit 3 (automated keratometry), Javal (manual keratometry), KR-1W (corneal topography), and Pentacam (Scheimpflug imaging). An experienced operator obtained 3 repeated measurements. An inexperienced operator obtained additional measurements with the SMI Reference Unit 3. Astigmatism vector analysis was used to determine the comparability, repeatability, and interobserver variability.
RESULTS: Corneal astigmatism vectors measured by automated, manual, or simulated keratometry were comparable except for the Pentacam equivalent keratometry (K) (P<.001, repeated-measures analysis of variance [ANOVA]). The mean difference between the equivalent K and other K values was 0.18 to 0.29 diopter (D) (P<.05, Hotelling trace multivariate ANOVA). The mean differences between automated, manual, and simulated keratometry were small (≤0.12 D). The within-subject standard deviation ranged from 0.05 D @ 21 degrees (KR-1W) to 0.18 D @ 23 degrees (Lenstar). The SMI Reference Unit showed small mean differences and comparable repeatability between the experienced operator and the inexperienced operator.
CONCLUSIONS: Vector analysis showed comparable corneal astigmatism measurements using automated, manual, and simulated keratometry. Pentacam equivalent K values were not comparable with those of the other keratometers. The repeatability of astigmatism magnitudes was acceptable; however, the repeatability of astigmatism meridians was moderate. The SMI Reference Unit showed good interobserver variability.
Copyright © 2012 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22999600     DOI: 10.1016/j.jcrs.2012.05.036

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  32 in total

1.  Clinical utility of ocular residual astigmatism and topographic disparity vector indexes in subclinical and clinical keratoconus.

Authors:  David P Piñero; Rafael J Pérez-Cambrodí; Roberto Soto-Negro; Pedro Ruiz-Fortes; Alberto Artola
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-09-07       Impact factor: 3.117

2.  The effects of ocular magnification on Spectralis spectral domain optical coherence tomography scan length.

Authors:  Irene Ctori; Stephen Gruppetta; Byki Huntjens
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-01-10       Impact factor: 3.117

3.  A novel Hartman Shack-based topography system: repeatability and agreement for corneal power with Scheimpflug+Placido topographer and rotating prism auto-keratorefractor.

Authors:  Gaurav Prakash; Dhruv Srivastava; Sounak Choudhuri
Journal:  Int Ophthalmol       Date:  2015-03-28       Impact factor: 2.031

4.  Comparison of two types of visual quality analyzer for the measurement of high order aberrations.

Authors:  Jing Hao; Lin Li; Fang Tian; Hong Zhang
Journal:  Int J Ophthalmol       Date:  2016-02-18       Impact factor: 1.779

5.  Repeatability of Ophtha Top topography and comparison with IOL-Master and LenstarLS900 in cataract patients.

Authors:  Sha-Sha Yu; Hui Song; Xin Tang
Journal:  Int J Ophthalmol       Date:  2017-11-18       Impact factor: 1.779

6.  Accuracy of two devices and three different calculation methods for predicting residual astigmatism after intraocular lens implantation.

Authors:  Kendall Bicknell; Christopher Helpert; Kourtney Dwyer; Brett McKnight; Sonali Singh; Samuel Fulcher; Derrick Fung
Journal:  Proc (Bayl Univ Med Cent)       Date:  2020-03-03

7.  Pseudolentogenic astigmatic effect of multifocal intraocular lenses: non-corneal ocular residual astigmatism (N-CORA) as a new parameter in astigmatic change analysis.

Authors:  Andreas Frings; Johannes Steinberg; Vasyl Druchkiv; Stephan J Linke; Toam Katz
Journal:  Int Ophthalmol       Date:  2016-09-24       Impact factor: 2.031

8.  Ocular residual astigmatism (ORA) in pre-cataract eyes prior to and after refractive lens exchange.

Authors:  Toam Katz; Johannes Steinberg; Vasyl Druchkiv; Stephan J Linke; Andreas Frings
Journal:  Int Ophthalmol       Date:  2016-09-13       Impact factor: 2.031

9.  Quantitative single and multi-surface clinical corneal topography utilizing optical coherence tomography.

Authors:  Ryan P McNabb; Anthony N Kuo; Joseph A Izatt
Journal:  Opt Lett       Date:  2013-04-15       Impact factor: 3.776

10.  Development of a program for toric intraocular lens calculation considering posterior corneal astigmatism, incision-induced posterior corneal astigmatism, and effective lens position.

Authors:  Youngsub Eom; Dongok Ryu; Dae Wook Kim; Seul Ki Yang; Jong Suk Song; Sug-Whan Kim; Hyo Myung Kim
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-08-19       Impact factor: 3.117

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.