Comparison of anterior segment parameters and axial length measurements performed on a Scheimpflug device with biometry function and a reference optical biometer.

Maria Muzyka-Woźniak1, Adam Oleszko2,3.   

Abstract

PURPOSE: To compare measurements of axial length (AL), corneal curvature (K), anterior chamber depth (ACD) and white-to-white (WTW) distance on a new device combining Scheimpflug camera and partial coherence interferometry (Pentacam AXL) with a reference optical biometer (IOL Master 500). To evaluate differences between IOL power calculations based on the two biometers.
METHODS: Ninety-seven eyes of 97 consecutive cataract or refractive lens exchange patients were examined preoperatively on IOL Master 500 and Pentacam AXL units. Comparisons between two devices were performed for AL, K, ACD and WTW. Intraocular lens (IOL) power targeting emmetropia was calculated with SRK/T and Haigis formulas on both devices and compared.
RESULTS: There were statistically significant differences between two devices for all measured parameters (P < 0.05), except ACD (P = 0.36). Corneal curvature measured with Pentacam AXL was significantly flatter then with IOL Master. The mean difference in AL was clinically insignificant (0.01 mm; 95% LoA 0.16 mm). Pentacam AXL yielded higher IOL power in 75% of eyes for Haigis formula and in 62% of eyes for SRK/T formula, with a mean difference within ± 0.5 D for 72 and 86% of eyes, respectively.
CONCLUSIONS: There were statistically significant differences between AL, K and WTW measurements obtained with the compared biometers. Flatter corneal curvature measurements on Pentacam AXL necessitate formulas optimisation for Pentacam AXL.

Entities:  

Keywords:  Axial lengths; IOL calculation; Keratometry; Optical biometry; Scheimpflug device; White-to white distance

Mesh:

Year:  2018        PMID: 29700651     DOI: 10.1007/s10792-018-0927-x

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  6 in total

1.  Ocular biometry through fully refocused steady-state magnetic resonance imaging sequence: reliability and agreement with the IOLMaster® 500.

Authors:  Lorenzo Ismael Perez-Sanchez; Julia Gutierrez-Vazquez; Maria Satrustegui-Lapetra; Francisco Ferreira-Manuel; Juan Jose Arevalo-Manso; Juan Jesus Gomez-Herrera; Juan Jose Criado-Alvarez
Journal:  Int Ophthalmol       Date:  2021-02-23       Impact factor: 2.031

2.  Comparison of Anterior Segment Parameters and Axial Length Using Two Scheimpflug Devices with Integrated Optical Biometers.

Authors:  Chantaka Supiyaphun; Sasivimol Rattanasiri; Passara Jongkhajornpong
Journal:  Clin Ophthalmol       Date:  2020-10-23

3.  Lenstar LS 900 versus Pentacam-AXL: analysis of refractive outcomes and predicted refraction.

Authors:  Henrique Aragão Arruda; Joana M Pereira; Arminda Neves; Maria João Vieira; Joana Martins; João C Sousa
Journal:  Sci Rep       Date:  2021-01-14       Impact factor: 4.379

4.  Evaluation and comparison of a novel Scheimpflug-based optical biometer with standard partial coherence interferometry for biometry and intraocular lens power calculation.

Authors:  Ziyang Wang; Wenli Yang; Dongjun Li; Wei Chen; Qi Zhao; Yifeng Li; Rui Cui; Lin Shen; Junfang Xian
Journal:  Exp Ther Med       Date:  2021-02-05       Impact factor: 2.447

5.  Inter-device measurement variability of vital data parameters for keratorefractive and cataract refractive surgery.

Authors:  Piotr Kanclerz; Katarzyna Przewłócka; Xiaogang Wang
Journal:  Ther Adv Ophthalmol       Date:  2021-09-20

6.  Comparison of optical low coherence interferometry and Scheimpflug imaging combined with partial coherence interferometry biometers in cataract eyes.

Authors:  Moonjung Kim; Eui S Han
Journal:  Saudi J Ophthalmol       Date:  2022-08-29
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.