| Literature DB >> 32336075 |
Bodin Khwannimit, Rattina Jomsuriya.
Abstract
BACKGROUND/AIM: FloTrac/Vigileo is a non-calibrated arterial pressure waveform analysis for cardiac index (CI) monitoring. The aim of our study was to compare the CI measured by the 4th generation of FloTrac with PiCCO in septic shock patients.Entities:
Keywords: arterial waveform analysis; cardiac output; hemodynamic monitoring; measurement technique; pulse contour analysis
Year: 2020 PMID: 32336075 PMCID: PMC7379425 DOI: 10.3906/sag-1909-58
Source DB: PubMed Journal: Turk J Med Sci ISSN: 1300-0144 Impact factor: 0.973
Patients clinical characteristic (n = 31)
| Men, n (%) | 17 (54.8) |
|---|---|
| Age (years) | 56.3 ± 18.5 |
| Body weight (kg) | 60.6 ± 10.1 |
| Height (cm) | 161.8 ± 8.2 |
| Body surface area (kg/m2) | 1.64 ± 0.16 |
| APACHE II | 26.5 ± 8.1 |
| SOFA | 9.8 ± 3.7 |
| ICU length of stay (days) | 7.7 ± 5.4 |
| Community acquired infection, n (%) | 20 (64.5) |
| Site of infection, n (%) | |
| - Respiratory tract infection | 13 (41.9) |
| - Digestive system | 7 (22.6) |
| - Primary bacteremia | 4 (12.9) |
| - Others* | 7 (22.5) |
| Combined vasoactive, n (%) | 13 (41.9) |
| Norepinephrine (ug/kg/min) | 0.33 ± 0.18 |
| Dopamine (ug/kg/min) | 11.5 ± 10.3 |
APACHE, acute physiology and chronic health evaluation; ICU, intensive care unit; SOFA, sequential organ failure assessment
Summary of the comparison between cardiac index measured by the 4th generation of FloTrac with PiCCO.
| Correlation, r | Bias(L/min/m2) | Limits of agreement(L/min/m2) | Percentageerror (%) | Concordance (%) | |
|---|---|---|---|---|---|
| CIv-CItd (n = 156) | 0.62 | 0.14 | –1.62–1.91 | 47.4 | 93.6 |
| CIv-CIp (n = 352) | 0.63 | -0.16 | –1.45–1.79 | 44.8 | 85.4 |
CIv, cardiac index obtained by FloTrac.CIp, cardiac index measured by pulse contour analysis of PiCCO.
Hemodynamic parameters before and after increase dose of norepinephrine (n = 56).
| Before | After | P value | |
|---|---|---|---|
| HR (/min) | 105.9 ± 24.9 | 115.4 ± 27.2 | < 0.001 |
| SBP (mmHg) | 104 (98–113) | 123.5 (110.25–140) | < 0.001 |
| DBP (mmHg) | 55 (52–62) | 66.5 (61–70.75) | < 0.001 |
| MAP (mmHg) | 68.4 ± 8.5 | 86.5 ± 11.0 | < 0.001 |
| CVP (mmHg) | 14 (12–16.7) | 15 (13–18) | < 0.001 |
| SVRI (dyne-s-m2/cm5) | 1,530.3 ± 410.6 | 1,508.3 ± 453.2 | <0.001 |
| CIv (L/min/m2) | 3.2 ± 0.7 | 3.9 ± 0.8 | < 0.001 |
| CIp (L/min/m2) | 3.4 ± 0.9 | 3.9 ± 0.8 | 0.003 |
| SVV (%) | 13 (7–22.2) | 9.5 (6.2–19) | 0.04 |
| NE (ug/kg/min) | 0.20 ± 0.16 | 0.27 ± 0.17 | < 0.001 |
CIv, cardiac index obtained by FloTrac; CIp, cardiac index measured by pulse contour analysis of PiCCO; CVP, central venous pressure; DBP, diastolic blood pressure; HR, heart rate; MAP, mean arterial pressure; NE, norepinephrine; SBP, systolic blood pressure; SVRI, systemic vascular resistance index; SVV, stroke volume variation.
Hemodynamic parameters before and after volume expansion (n = 16).
| Before | After | P-value | |
|---|---|---|---|
| HR (/min) | 107.9 ± 24.2 | 104.6 ± 23.3 | 0.01 |
| SBP (mmHg) | 100.6 ± 22.5 | 110.6 ± 23.8 | < 0.001 |
| DBP (mmHg) | 54 ± 8.6 | 57 ± 9.3 | < 0.001 |
| MAP (mmHg) | 67.9 ± 13.4 | 75.6 ± 14.5 | < 0.001 |
| CVP (mmHg) | 13.7 ± 3.1 | 15.2 ± 3.8 | 0.05 |
| SVRI (dyne-s-m2/cm5) | 1,728.1 ± 884.7 | 1,817.6 ± 965.9 | 0.29 |
| CIv (L/min/m2) | 2.7 ± 0.9 | 3.1 ± 0.9 | 0.0001 |
| CIp (L/min/m2) | 2.9 ± 1.1 | 3.2 ± 1.0 | 0.003 |
| SVV (%) | 18.5 ± 8.0 | 15.1 ± 8.2 | 0.06 |
CIv, cardiac index obtained by FloTrac; CIp, cardiac index measured by pulse contour analysis of PiCCO; CVP, central venous pressure; DBP, diastolic blood pressure; HR, heart rate; MAP, mean arterial pressure; NE, norepinephrine; SBP, systolic blood pressure; SVRI, systemic vascular resistance index; SVV, stroke volume variation.