Accuracy and Trending Ability of the Fourth-Generation FloTrac/Vigileo System in Patients With Low Cardiac Index.

Kohshi Hattori1, Takuma Maeda2, Tetsuhito Masubuchi1, Atsushi Yoshikawa1, Keigo Ebuchi1, Kuniko Morishima1, Masataka Kamei1, Kenji Yoshitani1, Yoshihiko Ohnishi1.   

Abstract

OBJECTIVES: To determine the accuracy and trending ability of the fourth-generation FloTrac/Vigileo in patients with low cardiac index by comparing FloTrac/Vigileo-derived cardiac index with that measured by 3-dimensional echocardiography.
DESIGN: Prospective clinical study.
SETTING: Cardiac surgery operating room in a single cardiovascular center. PARTICIPANTS: Twenty-five patients undergoing elective cardiac resynchronization therapy lead implantation.
INTERVENTIONS: FloTrac/Vigileo-derived cardiac index and 3-dimensional echocardiography-derived cardiac index were determined simultaneously before and after phenylephrine bolus and cardiac resynchronization therapy using 3-dimensional echocardiography-derived cardiac index as the reference method.
MEASUREMENTS AND MAIN RESULTS: Cardiac index measured by the fourth-generation FloTrac/Vigileo had a wide limit of agreement with that measured by 3-dimensional echocardiography, with a percentage error of 59.1%. The tracking ability of the unit after both phenylephrine administration and cardiac resynchronization therapy were measured by concordance rate, and both were below the acceptable limit (72.7% and 85%, respectively).
CONCLUSIONS: The degree of accuracy of the fourth-generation FloTrac/Vigileo in patients with low cardiac index was not acceptable, and high systemic vascular resistance in patients with low cardiac index may have contributed to this inaccuracy. The tracking ability of the fourth-generation FloTrac/Vigileo after phenylephrine administration or cardiac resynchronization therapy was below acceptable limits.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  cardiac output; echocardiography; heart failure; measurement technique; monitoring; ultrasound

Mesh:

Substances:

Year:  2016        PMID: 27612931     DOI: 10.1053/j.jvca.2016.06.016

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  8 in total

1.  Accuracy of the ClearSight™ system in patients undergoing abdominal aortic aneurysm surgery.

Authors:  Miho Sumiyoshi; Takuma Maeda; Erika Miyazaki; Naoshi Hotta; Hitoshi Sato; Eisuke Hamaguchi; Hiroko Kanazawa; Yoshihiko Ohnishi; Masataka Kamei
Journal:  J Anesth       Date:  2019-03-23       Impact factor: 2.078

2.  The accuracy and trending ability of cardiac index measured by the fourth-generation FloTrac/Vigileo system™ and the Fick method in cardiac surgery patients.

Authors:  Takuma Maeda; Eisuke Hamaguchi; Naoko Kubo; Akira Shimokawa; Hiroko Kanazawa; Yoshihiko Ohnishi
Journal:  J Clin Monit Comput       Date:  2018-11-07       Impact factor: 2.502

3.  Accuracy and trending ability of the fourth-generation FloTrac/Vigileo System™ in patients undergoing abdominal aortic aneurysm surgery.

Authors:  Takuma Maeda; Kohshi Hattori; Miho Sumiyoshi; Hiroko Kanazawa; Yoshihiko Ohnishi
Journal:  J Anesth       Date:  2018-04-03       Impact factor: 2.078

Review 4.  Accuracy of non-invasive and minimally invasive hemodynamic monitoring: where do we stand?

Authors:  Issa Pour-Ghaz; Theodore Manolukas; Nathalie Foray; Joel Raja; Aranyak Rawal; Uzoma N Ibebuogu; Rami N Khouzam
Journal:  Ann Transl Med       Date:  2019-09

5.  Evaluation of the use of the fourth version FloTrac system in cardiac output measurement before and after cardiopulmonary bypass.

Authors:  Sheng-Yi Lin; An-Hsun Chou; Yung-Fong Tsai; Su-Wei Chang; Min-Wen Yang; Pei-Chi Ting; Chun-Yu Chen
Journal:  J Clin Monit Comput       Date:  2017-10-16       Impact factor: 2.502

Review 6.  Cardiac Output Monitoring by Pulse Contour Analysis, the Technical Basics of Less-Invasive Techniques.

Authors:  Jörn Grensemann
Journal:  Front Med (Lausanne)       Date:  2018-03-06

Review 7.  Pulmonary Artery Catheter Monitoring in Patients with Cardiogenic Shock: Time for a Reappraisal?

Authors:  Maurizio Bertaina; Alessandro Galluzzo; Nuccia Morici; Alice Sacco; Fabrizio Oliva; Serafina Valente; Fabrizio D'Ascenzo; Simone Frea; Pierluigi Sbarra; Elisabetta Petitti; Silvia Brach Prever; Giacomo Boccuzzi; Paola Zanini; Matteo Attisani; Francesco Rametta; Gaetano Maria De Ferrari; Patrizia Noussan; Mario Iannaccone
Journal:  Card Fail Rev       Date:  2022-04-26

8.  Comparison the accuracy and trending ability of cardiac index measured by the fourth-generation of FloTrac with the PiCCO device in septic shock patients.

Authors:  Bodin Khwannimit; Rattina Jomsuriya
Journal:  Turk J Med Sci       Date:  2020-04-27       Impact factor: 0.973

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.