There is strong evidence that the IOL material is the factor having the greatest impact on posterior capsule opacification (PCO), anterior capsule opacification (ACO) development, and glistening formation after cataract surgery, even though there are other IOL features-such as haptic material and design and edge and optic design-that also have some influence. We reviewed the published literature describing the adverse events that are mainly related to the intraocular lens (IOL) material, such as PCO, ACO, and the subsequent capsule contraction, as well as glistening formation. The adverse events presented in this overview are the most common ones in clinical practice, and therefore, they are generally included in the clinical protocols for IOL evaluation.Entities:
Year: 2018 PMID: 29736282 PMCID: PMC5874989 DOI: 10.1155/2018/2687385
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Posterior capsule opacification (PCO) and neodymium:YAG (Nd:YAG) rates of several scientific articles published between 2000 and 2014.
| Authors | IOL model | IOL characteristics | Eyes ( | Study design | Follow-up | PCO rates (%) |
| Nd:YAG rates |
|
|---|---|---|---|---|---|---|---|---|---|
| Kucuksumer et al. [ | Acrysof MA60BM (Alcon) | 3-piece, hydrophobic acrylic | 32 | Contralateral & randomized study | 1 year | 93.7% grade 0 | <0.001 | 0% | — |
| MemoryLens | 3-piece, hydrophilic acrylic | 32 | 40.6% grade 0 | 0% | |||||
| Acrysof MA60BM (Alcon) | 3-piece, hydrophobic acrylic | 21 | Contralateral & randomized study | 3 years | 80.9% grade 0 | <0.05 | 0% | 0.46 | |
| MemoryLens | 3-piece, hydrophilic acrylic | 21 | 4.7% grade 0 | 19% | |||||
| Oner et al. [ | Acrysof MA30BA (Alcon) | 3-piece, rectangular-edge hydrophobic acrylic | 80 | Consecutive series | 16–22 months | 8.70% | <0.05 | 14.30% | — |
| Opsia-Agena 550 | Round-edge, PMMA | 77 | 24.70% | 26.30% | |||||
| Hayashi et al. [ | MZ60BD (Alcon) | Single-piece, PMMA | 90 | Prospective & randomized study | Up to 2 years | — | <0.0001 | 28.90% | — |
| SI-30NB (AMO) | Silicone | 83 | — | 14.40% | |||||
| Acrysof MA60BM (Alcon) | 3-piece, hydrophobic acrylic | 96 | — | 4.20% | |||||
| Schauersberger et al. [ | Acrysof (Alcon) | 3-piece, square-edge, hydrophobic acrylic | 25 | Prospective study | 3 years | 85% grade 0 | 0.323 | — | — |
| CeeOn 911A (AMO) | 3-piece, square-edge, silicone | 25 | 95% grade 0 | — | |||||
| Abhilakh Missier et al. [ | AcrySof MA30BA/MA60BM (Alcon) | 3-piece, hydrophobic acrylic | 107 | Retrospective & contralateral study | 3 years | 19.6% grade 0 | <0.0001 | 2.8% | <0.05 |
| AA4203VF (Staar) | Plate haptic, silicone | 107 | 6.5% grade 0 | 23.1% | |||||
| Bender et al. [ | 812A/Storz P497UV (Pharmacia) | PMMA | 63 | Retrospective study | 6 months | 21.9%/27.2% | <0.0001 | — | — |
| AcrySof MA30/SA31 (Alcon) | 3-piece/1-piece, hydrophobic acrylic | 33 | 10.5%/6.2% | ||||||
| SI-31 (AMO) | Silicone | 22 | 15.30% | ||||||
| Hydroview H61 (B&L) | Hydrophilic acrylic | 22 | 65.00% | ||||||
| 812A/Storz P497UV (Pharmacia) | 1-piece, PMMA | 63 | Retrospective study | 2 years | 32.9%/45.5% | — | — | ||
| AcrySof MA30/SA31 (Alcon) | 3-piece/1-piece, hydrophobic acrylic | 33 | 17.8%/18.4% | ||||||
| SI-31 (AMO) | Silicone | 22 | 17.80% | ||||||
| Hydroview H61 (B&L) | Hydrophilic acrylic | 22 | 62.20% | ||||||
| Prosdocimo et al. [ | CeeOn Edge (AMO) | 3-piece, sharp-edge, silicone | 40 | Prospective study | 18 months | 10% without VA lost | 0.0336 | 0% | — |
| Acrysof (Alcon) | 3-piece, square sharp-edge, hydrophobic acrylic | 38 | 29% without VA lost | 3% | |||||
| Auffarth et al. [ | — | Hydrophilic acrylic | 294 | Retrospective study | 3 years | 37% | <0.001 | 31.10% | <0.001 |
| — | PMMA | 384 | 28.30% | 19.30% | |||||
| — | Hydrophobic acrylic | 421 | 8.90% | 7.10% | |||||
| — | Silicone | 426 | 21.60% | 16.20% | |||||
| K. Hayashi and H. Hayashi [ | Hydroview H60M (B&L) | 3-piece, hydrophilic acrylic | 95 | Prospective randomized contralateral study | 24 months | — | — | 28% | <0.0001 |
| Acrysof MA60BM (Alcon) | 3-piece, hydrophobic acrylic | 95 | — | 2% | |||||
| Heatley et al. [ | Centerflex 570H (Rayner) | Single-piece, square-edge, hydrophilic acrylic | 53 | Prospective randomized contralateral study | 1 year | 50.30% | <0.001 | 2.60% | — |
| AcrySof SA60AT (Alcon) | Single-piece, square-edge, hydrophobic acrylic | 53 | 4.90% | 0% | |||||
| Kugelberg et al. [ | BL27 (B&L) | Single-piece, square-edge, hydrophilic acrylic | 60 | Prospective randomized study | 1 year | 18.20% | <0.001 | 3.50% | >0.05 |
| AcrySof SA60AT (Alcon) | Single-piece, square-edge, hydrophobic acrylic | 60 | 4.65% | 6.80% | |||||
| Hancox et al. [ | AcrySof SN60AT (Alcon) | 1-piece, hydrophobic acrylic | 36 | Prospective randomized contralateral study | 24 months | 8.83% | <0.0001 | — | — |
| AF-1 YA-60BB (Hoya) | Hydrophobic acrylic | 36 | 32.44% | ||||||
| Hayashi et al. [ | Acrysof MA60AC (Alcon) | 3-piece, round optic, hydrophobic acrylic | 45 | Prospective randomized contralateral study | 12 months | — | 0.5004 | 2.20% | >0.999 |
| AR40e (AMO) | 3-piece, round optic, hydrophobic acrylic | 45 | — | 2.20% | |||||
| Kohnen et al. [ | CeeOn Edge 911A (AMO) versus Acrysof MA60BM (Alcon) | Sharp-edge, silicone versus sharp-edge, hydrophobic acrylic | 139 | Prospective randomized contralateral study | 37 months | — | — | 2.1% versus 2.1% | |
| CeeOn Edge 911A (AMO) versus PhacoFlex SI40NB | Sharp-edge, silicone versus round optic edge, silicone | 108 | — | 5.7% versus 17% | |||||
| Kugelberg et al. [ | BL27 (B&L) | Hydrophlic acrylic | 60 | Prospective randomized study | 2 years | — | — | 42% | <0.001 |
| AcrySof SA60AT (Alcon) | Hydrophobic acrylic | 60 | — | 10% | |||||
| Boureau et al. [ | AcrySof SA60AT (Alcon) | 1-piece, sharp-edge, hydrophobic acrylic | 250 | Restrospective study | 2.9 years | 13.60% | <0.001 | 12% | <0.001 |
| AR40e (AMO) | 3-piece, sharp-edge, acrylic hydrophobic | 254 | 26.80% | 25.20% | |||||
| XL-Stabi (Zeiss) | 1-piece, sharp-edge, acrylic hydrophilic | 263 | 52.90% | 51% | |||||
| Ronbeck et al. [ | 809C (Pharmacia) | Round-edge, PMMA | 54 | Prospective randomized study | 5 years | 100% | 54% | <0.05 | |
| SI-40NB (AMO) | Round-edge, silicone | 48 | 12% | 29% | |||||
| Acrysof MA60BM (Alcon) | Sharp-edge, acrylic hydrophobic | 50 | 18% | 8% | |||||
| Vock et al. [ | Acrysof MA60BM (Alcon) | 3-piece, sharp-edge, hydrophobic acrylic | 98 | Retrospective study | 10 years | 91% grade 0 | 0.000073 | 42% | 0.007 |
| SI-30NB/SI-40NB (AMO) | 3-piece, round optic edge, silicone | 44 | 61% grade 0 | 18% | |||||
| Vock et al. [ | CeeOn 911A (AMO) | 3-piece, sharp-edge, silicone | 22 | Randomized and contralateral study | 6 years | — | 0.0016 | 9% | 0.01 |
| Acrysof MA60BM (Alcon) | 3-piece, sharp-edge, hydrophobic acrylic | 22 | — | 27.20% | |||||
| Gauthier et al. [ | AcrySof ReSTOR (Alcon) | Hydrophobic acrylic | 160 | Retrospective bilateral study | 2 years | — | — | 8.80% | 0.0001 |
| Acri.Lisa (Zeiss) | Hydrophilic acrylic with hydrophobic surface | 152 | — | 37.20% | |||||
| Iwase et al. [ | AcrySof SA60AT | 1-piece, sharp-edge, hydrophobic acrylic | 63 | Prospective randomized and contralateral study | 2 years | — | <0.001 | 2% | <0.01 |
| Meridian HP60M (B&L) | 1-piece, double square-edge, hydrophilic acrylic | 63 | — | 13% | |||||
| Vasavada et al. [ | Acrysof IQ SN60WF (Alcon) versus C-flex 570C (Rayner) | Hydrophobic acrylic versus hydrophilic acrylic | 66 | Prospective randomized and contralateral study | 3 years | — | 0 | 0% versus 12.9% | 0.04 |
| Acrysof IQ SN60WF (Alcon) versus Akreos Adapt (B&L) | Hydrophobic acrylic versus hydrophilic acrylic | 62 | — | 0 | 0% versus 16.1% | 0.02 | |||
| Zemaitiene and Jasinskas [ | Acrysof MA30BA (Alcon) | 3-piece, hydrophobic acrylic | 31 | Prospective randomized study | 3 years | — | 0.995 | 9% | NS |
| Acrysof SA30AL (Alcon) | 1-piece, hydrophobic acrylic | 31 | — | 3.10% | |||||
| CeeOn 911A (AMO) | 3-piece, silicone | 30 | — | 0% | |||||
| Chang et al. [ | Acrysof SA60AT (Alcon) | 1-piece, hydrophobic acrylic | 40 | Prospective randomized study | 5–7 years | — | 0.5535 | 22% | >0.05 |
| Sensar AR40e (AMO) | 3-piece, hydrophobic acrylic | 40 | — | 10% | |||||
| Leydolt et al. [ | iMics1 NY-60 (Hoya) | 1-piece, sharp-edge optimized, hydrophobic acrylic | 100 | Prospective, randomized and contralateral study | 3 years | — | <0.001 | 35.60% | 0.001 |
| Acrysof SN60WF (Alcon) | 1-piece, sharp-edge, hydrophobic acrylic | 100 | — | 13.70% | |||||
| Morgan-Warren and Smith [ | Hoya FY60AD (HOYA) | 3-piece, round-edge, hydrophobic acrylic | 315 | Retrospective comparative study | 2 years | — | — | 8.90% | 0.03 |
| Hoya PY60AD (HOYA) | 3-piece, sharp-edge, hydrophobic acrylic | 254 | — | 4.30% | |||||
| Acrysof SN60WF (Alcon) | 1-piece, sharp hydrophobic acrylic | 696 | — | 1.40% | 0.007 | ||||
| Bourdiol Ducasse et al. [ | Acrysof SN60WF (Alcon) | 1-piece, square-edge, hydrophobic acrylic | 126 | Retrospective study | 2-3 years | — | — | 10.30% | |
| Akreos AO-MI60 (B&L) | 1-piece, square-edge, hydrophilic acrylic | 89 | — | 36% | |||||
| Hoya YA-60BB (Hoya) | 3-piece, square-edge, hydrophobic acrylic | 85 | — | 24.90% | |||||
| Cullin et al. [ | AcrySof SN60 (Alcon) | Hydrophobic acrylic | 375 | Retrospective study | 41.5 months | — | — | 7.50% | <0.001 |
| Akreos Adapt (B&L) | Hydrophilic acrylic | 350 | — | 17.70% | |||||
| Tecnis Acryl Z9003 (AMO) | Hydrophobic acrylic | 801 | — | 3.70% | |||||
| Fong et al. [ | AcrySof SA60AT (Alcon) | 1-piece, square-edge, hydrophobic acrylic | 1016 | Prospective cohort study | 3 years | 34.40% | — | — | — |
| MA50BM (Alcon) | 3-piece, square-edge, hydrophobic acrylic | 67 | 50.80% | — | |||||
| Sensar AR40e (AMO) | 3-piece, round-edge, hydrophobic acrylic | 156 | 38.50% | — | |||||
| Akreos Adapt (B&L)/Quatrix (Croma) | Square-edge, hydrophilic acrylic | 101 | 64.40% | — | |||||
| Ronbeck and Kugelberg [ | 809C (Pharmacia) | Round-edge, HSM PMMA | 61 | Prospective randomized study | 12 years | — | — | 57.40% | |
| SI-40NB (AMO) | round-edge, silicone | 59 | — | 28.80% | |||||
| Acrysof MA60BM (Alcon) | Sharp-edge, hydrophobic acrylic | 59 | — | 32.20% |
Impact of glistenings on visual function with several models of AcrySof intraocular lenses (IOLs).
| Author | IOL model | Eyes ( | Follow-up | Results |
|---|---|---|---|---|
| Moreno-Montanes et al. [ | AcrySof MA30BA | 129 | 20 ± 11 months | No impact on visual acuity |
| Colin et al. [ | AcrySof SN60AT, SN60WF, SA60AT, or MA | 260 | 0–86 months | No impact on visual acuity |
| Colin et al. [ | AcrySof SN60WF | 111 | ±24 months | No impact on visual acuity |
| Monestam and Behndig [ | AcrySof MA60BM | 103 | >10 years | No impact on visual acuity at high and low contrast |
| Colin and Orignac [ | Several AcrySof | 97 | 18 months | No impact on visual acuity or contrast sensitivity |
| Hayashi et al. [ | AcrySof versus SI30NB versus PMMA | 35 | >10 years | No impact on visual acuity and aberrometry |
| Chang et al. [ | AcrySof SA60AT versus Sensar AR40e | 80 | 5–7 years | No impact on visual acuity or contrast sensitivity |
| Chang and Kugelberg [ | AcrySof SA60AT & BL27 | 120 | 9 years | No impact on visual acuity or contrast sensitivity |
| Waite et al. [ | AcrySof SA & SN60 | 20 | 36 months | No impact on visual acuity, aberrometry, and contrast sensitivity but at high spatial frequencies |
| Gunenc et al. [ | AcrySof MA30BA & MA60BA | 91 | 7–24 months | Decreased contrast sensitivity at high spatial frequencies |
| Xi et al. [ | AcrySof SA60AT | 120 | 2 years | No impact on visual acuity and contrast sensitivity but at high spatial frequencies |
| Christiansen et al. [ | AcrySof MA30BA & MA60BM | 42 | 6–46 months | Decreased visual acuity but not glare and contrast sensitivity |