Posterior capsule opacification in rabbit eyes implanted with 1-piece and 3-piece hydrophobic acrylic intraocular lenses.

Liliana Werner1, Nick Mamalis, Andrea M Izak, Suresh K Pandey, Brandon L Davis, Chistian D Nilson, Christopher Weight, David J Apple.   

Abstract

PURPOSE: To evaluate the outcome of posterior capsule opacification (PCO) after implantation in rabbit eyes of currently available 3-piece and 1-piece hydrophobic acrylic intraocular lenses (IOLs) with square optic edges.
SETTING: John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA.
METHODS: The 3-piece designs evaluated were the AR40e (Advanced Medical Optics Inc.) and the MA60AC (Alcon, Inc.); the 1-piece designs were the SA60AT and the SA30AT (Alcon, Inc.). Nine lenses of each type were implanted in a randomized manner by the same surgeon in 18 Dutch Belted pigmented rabbits. After a follow-up of 3 weeks, the rabbits were killed and analyses of the enucleated eyes were performed from the posterior or Miyake-Apple view. The intensity of central PCO, peripheral PCO, and Soemmering's ring formation was scored from 0 to 4. The area of Soemmering's ring formation was also scored from 0 to 4 based on the number of quadrants involved. Other parameters analyzed were capsulorhexis coverage of the IOL anterior surface, IOL centration, fixation, and presence of striae. Results from the posterior view were complemented by histopathologic evaluation of the eyes.
RESULTS: No statistically significant difference was found between the 4 groups of IOLs in the parameters analyzed from the posterior view. When cell ingrowth occurred with the 1-piece designs, causing peripheral and central PCO formation, it was more likely to start at the optic-haptic junctions, as observed during the clinical follow-up with slitlamp examination and confirmed by gross and histopathologic analyses of the enucleated eyes.
CONCLUSIONS: The square, truncated optic edge is the most important IOL design feature for PCO prevention. The optic-haptic junctions of the 1-piece designs appear to be sites where the barrier effect of the truncated optic edge is less effective.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15899460     DOI: 10.1016/j.jcrs.2004.06.088

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  5 in total

1.  Influence of super-hydrophobic silicone rubber substrate on the growth and differentiation of human lens epithelial cells.

Authors:  Liuxueying Zhong; Jiayi Jin; Danying Zheng; Weisheng Guan; Yue Guo; Anfu Chen; Yujiang Peng; Qianying Gao; Yongxin Zheng; Hanxiong Huang
Journal:  J Mater Sci Mater Med       Date:  2018-11-13       Impact factor: 3.896

2.  Findings of an experimental study in a rabbit model on posterior capsule opacification after implantation of hydrophobic acrylic and hydrophilic acrylic intraocular lenses.

Authors:  Nikolaos Trakos; Elli Ioachim; Elena Tsanou; Miltiadis Aspiotis; Konstantinos Psilas; Chris Kalogeropoulos
Journal:  Clin Ophthalmol       Date:  2008-12

3.  Square-edge intraocular lenses and epithelial lens cell proliferation: implications on posterior capsule opacification in an in vitro model.

Authors:  Rita Mencucci; Eleonora Favuzza; Carlotta Boccalini; Jean-Jacques Gicquel; Laura Raimondi
Journal:  BMC Ophthalmol       Date:  2015-01-19       Impact factor: 2.209

Review 4.  Biomaterial Influence on Intraocular Lens Performance: An Overview.

Authors:  Cari Pérez-Vives
Journal:  J Ophthalmol       Date:  2018-03-15       Impact factor: 1.909

5.  Time Course of Lens Epithelial Cell Behavior in Rabbit Eyes following Lens Extraction and Implantation of Intraocular Lens.

Authors:  Daijiro Kurosaka; Toshiyasu Imaizumi; Junya Kizawa
Journal:  J Ophthalmol       Date:  2021-01-16       Impact factor: 1.909

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.