Equilibrium water content and glistenings in acrylic intraocular lenses.

Akira Miyata1, Shigeo Yaguchi.   

Abstract

PURPOSE: To investigate the possible relationship between temperature and water absorption in acrylic intraocular lenses (IOLs).
SETTING: Laboratory setting, Miyata Eye Clinic, Hiroshima, Japan.
METHODS: The equilibrium water content (amount of water absorbed per weight of the resin x 100) in 2 hydrophobic acrylic IOLs (AcrySof MA60BM [Alcon] and Sensar AR40 [AMO]) was calculated at 30 degrees C, 40 degrees C, and 50 degrees C. The 2 IOLs were also subjected to 3 changes in temperature: from 37 degrees C to 35 degrees C, 39 degrees C to 35 degrees C, and 41 degrees C to 35 degrees C. They were incubated in physiological saline at the higher temperature for 2 hours and at the lower temperature for 30 days before being examined for glistening formation.
RESULTS: The water content was higher in the AR40 IOL than in the MA60BM IOL at all temperatures. A temperature-dependent increase in water content was seen in both IOLs, greater in the MA60BM. With a temperature change from 37 degrees C to 35 degrees C, glistening formation was not observed in either IOL. With a temperature change from 39 degrees C to 35 degrees C, glistenings were observed in the MA60BM IOL, and with a temperature change from 41 degrees C to 35 degrees C, they were observed in both IOLs.
CONCLUSION: The change in the equilibrium water content caused by temperature changes between 30 degrees C and 40 degrees C is an important factor in glistening formation, and thus an IOL featuring less temperature-dependent water absorption is less likely to form glistenings.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15313305     DOI: 10.1016/j.jcrs.2003.12.038

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  15 in total

1.  Analysis of glistenings in hydrophobic acrylic intraocular lenses on visual performance.

Authors:  Lei Xi; Yi Liu; Feng Zhao; Chuan Chen; Bing Cheng
Journal:  Int J Ophthalmol       Date:  2014-06-18       Impact factor: 1.779

2.  Factors associated with changes in posterior corneal surface following photorefractive keratectomy.

Authors:  Achia Nemet; Michael Mimouni; Igor Vainer; Tzahi Sela; Igor Kaiserman
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2021-06-07       Impact factor: 3.117

3.  Visual results and complications of primary intraocular lens implantation in infants aged 6 to 12 months.

Authors:  Yi Lu; Ying-Hong Ji; Yi Luo; Yong-Xiang Jiang; Man Wang; Xu Chen
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-02-17       Impact factor: 3.117

4.  Effects of intraocular lens glistenings on visual function: a prospective study and presentation of a new glistenings grading methodology.

Authors:  Nick Stanojcic; David P S O'Brart; Nick Maycock; Chistopher C Hull
Journal:  BMJ Open Ophthalmol       Date:  2019-04-01

5.  Changes in Opacification of Hydrophobic Acrylic Intraocular Lenses According to Temperature and Hydration.

Authors:  Jung Youb Kang; Ju Hwan Song; Sang Joon Lee
Journal:  Clin Ophthalmol       Date:  2020-10-14

6.  Evolution of visual performance in 70 eyes implanted with the Tecnis(®) ZMB00 multifocal intraocular lens.

Authors:  Carlos Palomino Bautista; David Carmona González; Alfredo Castillo Gómez
Journal:  Clin Ophthalmol       Date:  2012-03-21

Review 7.  Biomaterial Influence on Intraocular Lens Performance: An Overview.

Authors:  Cari Pérez-Vives
Journal:  J Ophthalmol       Date:  2018-03-15       Impact factor: 1.909

8.  Evaluation of in vitro glistening formation in hydrophobic acrylic intraocular lenses.

Authors:  Brett E Thomes; Thomas A Callaghan
Journal:  Clin Ophthalmol       Date:  2013-07-25

Review 9.  Clinical properties of a novel, glistening-free, single-piece, hydrophobic acrylic IOL.

Authors:  Mark Packer; Mohan Rajan; Edoardo Ligabue; Peter Heiner
Journal:  Clin Ophthalmol       Date:  2014-02-21

10.  Glistening formation in a new hydrophobic acrylic intraocular lens.

Authors:  Timur M Yildirim; Hui Fang; Sonja K Schickhardt; Qiang Wang; Patrick R Merz; Gerd U Auffarth
Journal:  BMC Ophthalmol       Date:  2020-05-06       Impact factor: 2.209