Literature DB >> 12106730

Glistenings in foldable intraocular lenses.

Daniele Tognetto1, Lisa Toto, Giorgia Sanguinetti, Giuseppe Ravalico.   

Abstract

PURPOSE: To evaluate the formation of glistenings in foldable intraocular lenses (IOLs) of various materials after cataract surgery.
SETTING: University Eye Clinic of Trieste, Trieste, Italy.
METHODS: In this prospective study, 273 patients were randomized to have implantation of 1 of 7 types of foldable IOLs in the capsular bag after phacoemulsification. The IOLs were the CeeOn Edge 911A (Pharmacia & Upjohn Co.), ACR6D (Corneal), AcrySof (Alcon), SI-40NB (AMO), Hydroview H60M (Storz), Sensar (AMO), and Stabibag (Ioltech). The presence of glistenings in the IOL optic was assessed by slitlamp evaluation 7, 30, 90, 180, 360, and 720 days after surgery. The glistenings were graded at each control.
RESULTS: Glistenings were observed in all IOL groups. The percentage of patients with glistenings ranged from 40.0% to 67.5%, with the percentage increasing up to 90 days after surgery and then becoming stable in most groups. The AcrySof group had a continuous increase over time. The mean grade of glistenings increased up to 180 days after surgery and then became stable in most groups. The AcrySof and 911A groups had a continuous increase over time.
CONCLUSIONS: Glistening formation was observed in 7 different foldable IOLs. The AcrySof group had a higher percentage and a greater density of glistenings.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12106730     DOI: 10.1016/s0886-3350(02)01353-6

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  22 in total

1.  Intraoperative clouding of a posterior chamber intraocular lens.

Authors:  Pallavi Tyagi; Nimish Shah; Mohamad Jabir
Journal:  Int Ophthalmol       Date:  2011-12-03       Impact factor: 2.031

2.  Preoperative opacification of acrylic intraocular lenses in storage.

Authors:  C M Pratt; S Barton; E McGonigle; M Kishi; P J S Foot
Journal:  J Mater Sci Mater Med       Date:  2007-04       Impact factor: 3.896

3.  Effects of glistenings in intraocular lenses.

Authors:  Marrie van der Mooren; Luuk Franssen; Patricia Piers
Journal:  Biomed Opt Express       Date:  2013-07-11       Impact factor: 3.732

4.  Spontaneous resolution of early postoperative intraocular lens opacification in a patient with uveitis.

Authors:  J Wong; A Jap; S-P Chee
Journal:  Eye (Lond)       Date:  2011-09-16       Impact factor: 3.775

5.  Analysis of glistenings in hydrophobic acrylic intraocular lenses on visual performance.

Authors:  Lei Xi; Yi Liu; Feng Zhao; Chuan Chen; Bing Cheng
Journal:  Int J Ophthalmol       Date:  2014-06-18       Impact factor: 1.779

6.  Polymethyl methacrylate intraocular lens opacification 20 years after cataract surgery: A case report in a tertiary eye hospital in Saudi Arabia.

Authors:  Abdullah G Al-Otaibi; Elham S Al-Qahtani
Journal:  Saudi J Ophthalmol       Date:  2011-12-19

7.  Post-DSAEK optical changes: a comprehensive prospective analysis on the role of ocular wavefront aberrations, haze, and corneal thickness.

Authors:  Holly B Hindman; Krystel R Huxlin; Seth M Pantanelli; Christine L Callan; Ramkumar Sabesan; Steven S T Ching; Brooke E Miller; Tim Martin; Geunyoung Yoon
Journal:  Cornea       Date:  2013-12       Impact factor: 2.651

8.  Refractive index and its impact on pseudophakic dysphotopsia.

Authors:  Bryce R Radmall; Anne Floyd; Zack Oakey; Randall J Olson
Journal:  Clin Ophthalmol       Date:  2015-07-20

9.  Evaluation of in vitro glistening formation in hydrophobic acrylic intraocular lenses.

Authors:  Brett E Thomes; Thomas A Callaghan
Journal:  Clin Ophthalmol       Date:  2013-07-25

10.  Safety and effectiveness of a glistening-free single-piece hydrophobic acrylic intraocular lens (enVista).

Authors:  Mark Packer; Luther Fry; Kevin T Lavery; Robert Lehmann; James McDonald; Louis Nichamin; Brian Bearie; Jon Hayashida; Griffith E Altmann; Omid Khodai
Journal:  Clin Ophthalmol       Date:  2013-09-24
View more