Fellow-eye comparison of posterior capsule opacification with AcrySof SN60AT and AF-1 YA-60BB blue-blocking intraocular lenses.

Joanne Hancox1, David Spalton, Georgia Cleary, James Boyce, Mayank A Nanavaty, Sri Thyagarajan, John Marshall.   

Abstract

PURPOSE: To evaluate intraindividual differences in posterior capsule opacification (PCO) and visual performance of AcrySof SN60AT (Alcon Laboratories) and AF-1 YA-60BB (Hoya Corp.) intraocular lenses (IOLs).
SETTING: Ophthalmology Department, St. Thomas' Hospital, London, United Kingdom.
METHODS: In this prospective single-surgeon standardized-surgical-procedure fellow-eye comparison, an AcrySof SN60AT or an AF-1 YA-60BB IOL was randomized to the first eye of 36 patients and fellow-eye surgery was performed within 4 to 6 weeks. Follow-up was at 1, 3, 6, 12, and 24 months. Best distance- corrected logMAR visual acuity was measured at 100% and 9% contrast. Contrast sensitivity was measured using the Functional Acuity Contrast Test (FACT) on the Optec 3500 instrument (Stereo Optical Company, Inc.). Color vision was assessed with the Farnsworth Munsell 100-hue test. After pupil dilation, digital retroillumination photographs were taken and the percentage area of PCO was calculated using POCO software.
RESULTS: Posterior capsule opacification was significantly greater in the AF-1 YA-60BB group than in the AcrySof SN60AT group at all time points (P<.0001), with the difference greatest at 24 months. At 24 months, 100% contrast visual acuity was significantly better in the AcrySof SN60AT group than in the AF-1 YA-60BB group (P = .0313); 9% contrast visual acuity was significantly better in the AcrySof SN60AT group from 6 months onward. There was no significant difference between groups in color vision or contrast sensitivity. Electron microscopy showed the AcrySof SN60AT IOL has a much sharper posterior edge profile.
CONCLUSIONS: The AcrySof SN60AT IOL had better PCO performance and thus visual performance than the AF-1 YA-60BB IOL. This is attributable to the differences in optic edge design.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18721708     DOI: 10.1016/j.jcrs.2008.05.024

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  6 in total

1.  Academy IRIS® Registry Analysis of Incidence of Laser Capsulotomy Due to Posterior Capsule Opacification After Intraocular Lens Implantation.

Authors:  Jeffrey D Horn; Bret L Fisher; Daniel Terveen; Helene Fevrier; Mohinder Merchea; Xiaolin Gu
Journal:  Clin Ophthalmol       Date:  2022-06-01

2.  Intraocular lens-edge design and material factors contributing to posterior-capsulotomy rates: comparing Hoya FY60aD, PY60aD, and AcrySof SN60WF.

Authors:  Peter J Morgan-Warren; Jm Alaric Smith
Journal:  Clin Ophthalmol       Date:  2013-08-22

Review 3.  Biomaterial Influence on Intraocular Lens Performance: An Overview.

Authors:  Cari Pérez-Vives
Journal:  J Ophthalmol       Date:  2018-03-15       Impact factor: 1.909

Review 4.  An overview of the influence and design of biomaterial of the intraocular implant of the posterior capsule opacification.

Authors:  Razvan Vladimir Nanu; Emil Ungureanu; Sinziana Luminita Instrate; Alexandra Vrapciu; Roxana Cozubas; Laura Carstocea; Liliana Mary Voinea; Radu Ciuluvica
Journal:  Rom J Ophthalmol       Date:  2018 Jul-Sep

5.  Effect of AcrySof versus other intraocular lens properties on the risk of Nd:YAG capsulotomy after cataract surgery: A systematic literature review and network meta-analysis.

Authors:  Howard Thom; Frank Ender; Saisudha Samavedam; Caridad Perez Vivez; Subhajit Gupta; Mukesh Dhariwal; Jan de Haan; Derek O'Boyle
Journal:  PLoS One       Date:  2019-08-19       Impact factor: 3.240

6.  Effect of hydrophobic acrylic versus hydrophilic acrylic intraocular lens on posterior capsule opacification: meta-analysis.

Authors:  Ying Li; Jiaxing Wang; Zhuo Chen; Xin Tang
Journal:  PLoS One       Date:  2013-11-05       Impact factor: 3.240