Evaluation of a novel Scheimpflug-based non-contact tonometer in healthy subjects and patients with ocular hypertension and glaucoma.

Lukas Reznicek1, Daniel Muth, Anselm Kampik, Aljoscha S Neubauer, Christoph Hirneiss.   

Abstract

BACKGROUND: To evaluate the agreement of intraocular pressure (IOP) and central corneal thickness (CCT) measurements obtained with the non-contact tonometer Corvis Scheimpflug Technology (Corvis ST, OCULUS, Wetzlar, Germany) versus Goldmann applanation tonometry (GAT) and ultrasound-based corneal pachymetry (US-CCT).
METHODS: Eye healthy participants, patients with ocular hypertension (OHT) and patients with open-angle glaucoma were included in this prospective study. In each participant, GAT, US-CCT and measurements with Corvis ST were obtained (Corvis-IOP and Corvis-CCT). Accuracy and repeatability were tested by correlation and regression analyses, Bland-Altman plots and assessment of intraclass correlation coefficients.
RESULTS: A consecutive series of 188 right study eyes of 188 participants (142 eyes with glaucoma, 10 eyes with OHT and 36 control eyes) were included in this prospective study. The mean GAT of all included was 14.5±4.8 mm Hg compared with mean Corvis-IOP of 15.4±5.6 mm Hg (Spearman's r=0.75, p<0.0001). Mean US-CCT was 544.56±40.0 µm compared with Corvis-CCT of 545.2±46.5 µm (Pearson's r=0.78, p<0.0001). Bland-Altman plots of all included eyes as well as subgroup analyses revealed good agreement of the IOP and CCT measurement techniques. High intraclass correlation coefficient values in 17 patients with repeated measurements revealed very good repeatability (0.942 and 0.937 for Corvis-IOP and Corvis-CCT, respectively). Corvis-IOP but not GAT showed a trend of dependence on CCT.
CONCLUSIONS: Obtaining CCT and measuring IOP with the Corvis ST reveals very good repeatability and good accuracy in healthy subjects and patients with OHT and glaucoma when compared with standardised US pachymetry or GAT.

Entities:  

Keywords:  Glaucoma; Intraocular pressure

Mesh:

Year:  2013        PMID: 23969314     DOI: 10.1136/bjophthalmol-2013-303400

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  30 in total

1.  Intraocular pressure measurement with Corvis ST in comparison with applanation tonometry and Tomey non-contact tonometry.

Authors:  Jan Luebke; L Bryniok; M Neuburger; J F Jordan; D Boehringer; T Reinhard; T Wecker; A Anton
Journal:  Int Ophthalmol       Date:  2019-04-09       Impact factor: 2.031

2.  Comparative analysis of biomechanically corrected intraocular pressure with corneal visualization Scheimpflug technology versus conventional noncontact intraocular pressure.

Authors:  Jiaonan Ma; Yan Wang; Weiting Hao; Vishal Jhanji
Journal:  Int Ophthalmol       Date:  2019-08-20       Impact factor: 2.031

3.  IOP measurement in silicone oil tamponade eyes by Corvis ST tonometer, Goldmann applanation tonometry and non-contact tonometry.

Authors:  Yang Zhang; Lin Zheng; Ailing Bian; Qi Zhou
Journal:  Int Ophthalmol       Date:  2017-04-25       Impact factor: 2.031

4.  Corneal biomechanical properties in 3 corneal transplantation techniques with a dynamic Scheimpflug analyzer.

Authors:  Naoyuki Maeda; Ryotaro Ueki; Mutsumi Fuchihata; Hisataka Fujimoto; Shizuka Koh; Kohji Nishida
Journal:  Jpn J Ophthalmol       Date:  2014-09-05       Impact factor: 2.447

5.  Evaluation of Corvis ST tonometer with the updated software in glaucoma practice.

Authors:  Ioannis Halkiadakis; Vasilios Tzimis; Alexandros Gryparis; Ioannis Markopoulos; Vasiliki Konstadinidou; Elias Zintzaras; Michalis Tzakos
Journal:  Int J Ophthalmol       Date:  2022-03-18       Impact factor: 1.779

6.  Coupled biomechanical response of the cornea assessed by non-contact tonometry. A simulation study.

Authors:  Miguel Á Ariza-Gracia; Jesús F Zurita; David P Piñero; José F Rodriguez-Matas; Begoña Calvo
Journal:  PLoS One       Date:  2015-03-17       Impact factor: 3.240

7.  Comparison of intraocular pressure measurement with Scheimpflug-based noncontact tonometer with and without hydrogel contact lenses.

Authors:  Mukesh Kumar; Rohit Shetty; Chaitra Jayadev; Debarun Dutta; Maneck D Nicolsan; Sriharsha Nagaraj; Rajesh S Kumar
Journal:  Indian J Ophthalmol       Date:  2015-04       Impact factor: 1.848

8.  Ocular surface displacement with and without contact lenses during non-contact tonometry.

Authors:  Ulfah Rimayanti; Yoshiaki Kiuchi; Shohei Uemura; Joji Takenaka; Hideki Mochizuki; Makoto Kaneko
Journal:  PLoS One       Date:  2014-04-29       Impact factor: 3.240

9.  [Pachymetry and intraocular pressure measurement by corneal visualization Scheimpflug technology (Corvis ST): A clinical comparison to the gold standard].

Authors:  J Steinberg; J Mehlan; A Frings; V Druchkiv; G Richard; T Katz; S J Linke
Journal:  Ophthalmologe       Date:  2015-09       Impact factor: 1.059

10.  Evaluation of corneal deformation analyzed with Scheimpflug based device in healthy eyes and diseased ones.

Authors:  Michele Lanza; Michela Cennamo; Stefania Iaccarino; Carlo Irregolare; Miguel Rechichi; Mario Bifani; Ugo Antonello Gironi Carnevale
Journal:  Biomed Res Int       Date:  2014-06-23       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.