Bernardo T Lopes1,2, Cynthia J Roberts3, Ahmed Elsheikh4, Riccardo Vinciguerra5, Paolo Vinciguerra6,7,8, Sven Reisdorf9, Stefanie Berger9, Robert Koprowski10, Renato Ambrósio1,2.
Abstract
PURPOSE: To assess the repeatability and reproducibility of dynamic corneal response parameters measured by the Corvis ST (Oculus, Wetzlar, Germany).Entities:
Year: 2017 PMID: 28676837 PMCID: PMC5476874 DOI: 10.1155/2017/8515742
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Corvis ST repeatability and reproducibility IOP and DCR indices.
| Mean | Repeatability | Reproducibility | |||||
|---|---|---|---|---|---|---|---|
|
| CV | CR |
| CV | CR | ||
| IOP (mmHg) | 14.73 | 0.98 | 0.066 | 2.7146 | 1.12 | 0.076 | 3.1024 |
| bIOP(mmHg) | 14.57 | 0.89 | 0.061 | 2.4653 | 1.05 | 0.072 | 2.9085 |
| DA Max (mm) | 1.09 | 0.04 | 0.038 | 0.1108 | 0.06 | 0.057 | 0.1662 |
| DefA Max (mm) | 0.91 | 0.03 | 0.037 | 0.0831 | 0.04 | 0.053 | 0.1108 |
| DAratio in 2 mm | 4.29 | 0.13 | 0.032 | 0.3601 | 0.23 | 0.054 | 0.6371 |
| DAratio in 1 mm | 1.57 | 0.019 | 0.012 | 0.05263 | 0.025 | 0.012 | 0.06925 |
| Integrated radius (mm−1) | 8.17 | 0.26 | 0.031 | 0.7202 | 0.31 | 0.038 | 0.8587 |
| Maximum inverse radius (mm−1) | 0.139 | 0.003 | 0.024 | 0.00831 | 0.006 | 0.046 | 0.01662 |
| A1 velocity (m/s) | 0.15 | 0.008 | 0.054 | 0.02216 | 0.012 | 0.079 | 0.03324 |
IOP: intraocular pressure; bIOP: biomechanical-corrected IOP; DA Max: maximum deformation amplitude; DefA Max: maximum deflection amplitude; DAratio: deformation amplitude ratio; integrated radius: integrated sum of inverse radius between the first and second applanation events; maximum inverse radius: inverse concave radius at the highest concavity moment; A1 velocity: speed of the corneal apex at applanation.