Color light-emitting diode reflection topography: validation of keratometric repeatability in a large sample of wide cylindrical-range corneas.

Anastasios John Kanellopoulos1, George Asimellis2.   

Abstract

PURPOSE: To investigate repeatability of steep and flat keratometry measurements, as well as astigmatism axis in cohorts with normal range and regular astigmatic such as: eyes following laser-assisted in situ keratomileusis (LASIK) and normal population, as well as cohorts of high and irregular astigmatism such as keratoconic eyes, and keratoconic eyes following corneal collagen cross-linking, employing a novel corneal reflection topography device.
METHODS: Steep and flat keratometry and astigmatism axis measurement repeatability was investigated employing a novel multicolored-spot reflection topographer (Cassini) in four study groups, namely a post myopic LASIK-treated Group A, a keratoconus Group B, a post-CXL keratoconus Group C, and a control Group D of routine healthy patients. Three separate, maps were obtained employing the Cassini, enabling investigation of the intra-individual repeatability by standard deviation. Additionally we investigated in all groups,the Klyce surface irregularity indices for keratoconus, the SAI (surface asymmetry index) and the SRI (surface regularity index).
RESULTS: Flat keratometry repeatability was 0.74±0.89 (0.03 to 5.26) diopters (D) in the LASIK Group A, 0.88±1.45 (range minimum to maximum, 0.00 to 7.84) D in the keratoconic Group B, and 0.71±0.94 (0.02 to 6.23) D in the cross-linked Group C. The control Group D had flat keratometry repeatability 0.36±0.46 (0.00 to 2.71) D. Steep keratometry repeatability was 0.64±0.82 (0.01 to 4.81) D in the LASIK Group A, 0.89±1.22 (0.02 to 7.85) D in the keratoconic Group B, and 0.93±1.12 (0.04 to 5.93) D in the cross-linked Group C. The control Group D had steep keratometry repeatability 0.41±0.50 (0.00 to 3.51) D. Axis repeatability was 3.45±1.62° (0.38 to 7.78°) for the LASIK Group A, 4.12±3.17° (0.02 to 12.13°) for the keratoconic Group B, and 3.20±1.99° (0.17 to 8.61°) for the cross-linked Group C. The control Group D had axis repeatability 2.16±1.39° (0.05 to 5.86°). The SAI index measurement repeatability was 0.33±0.40 (0.01 to 2.31) in the post-LASIK Group A, 0.39±0.75 (0.00 to 7.15) in the keratoconic Group B, and 0.43±0.56 (0.05 to 3.50) in the keratoconus post-CXL Group C. The control group had SAI measurement repeatability of 0.26±0.30 (0.00 to 2.39). The SRI index repeatability was 0.22±0.17 (0.01 to 0.96) for post-LASIK Group A, 0.20±0.18 (0.00 to 1.07) in keratoconic Group B, and 0.13±0.09 (0.00 to 0.45) in the keratoconus post-CXL Group C. The control Group D had SRI measurement repeatability of 0.23±0.16 (0.00 to 0.75).
CONCLUSION: This novel corneal topography device appears to offer very high specificity in estimating corneal keratometry and specific corneal irregularity indices, even in topographically challenging corneas such as LASIK treated, keratoconic, and cross-linked.

Entities:  

Keywords:  Cassini topographer; LED topography; VU topographer; axis of astigmatism; form factor; point-source topography; steep meridian; surface asymmetry index; surface regularity index

Year:  2015        PMID: 25709385      PMCID: PMC4332260          DOI: 10.2147/OPTH.S68371

Source DB:  PubMed          Journal:  Clin Ophthalmol        ISSN: 1177-5467


Introduction

Corneal topographers provide qualitative characteristics and quantitative metrics of the anterior corneal surface,1 transforming raw data into color-coded dioptric power-scale refractive and curvature maps. In addition, elevation maps can be derived by slope measurement data. Specific topography indices associated with corneal pathology have also been introduced by topography – analysis algorithms.2 While Placido-disc systems have been the mainstream of videokeratoscopy,3–6 certain restrictions exist with this technology,7 such as the limited ability to identify contour topographic changes.8 To override these limitations, topography systems with color-coded specular reflection forward ray-tracing have been proposed9 as alternatives to Placido-ring imaging,10 enabling, among other things, the proper imaging of radial, in addition to only contour topographic, changes. The “Cassini” (i-Optics, The Hague, the Netherlands) system is a novel topographer utilizing a multicolor (red, yellow, and green) spot pattern consisting of hundreds of light-emitting diode (LED) spots (up to 700) projected on the cornea. Image-processing software locates feature points in the reflected images and accounts for smearing and deformation in irregular corneas. The system has been recently introduced11 and has received US Food and Drug Administration approval for clinical topography. Due to their novelty, the clinical validation of such systems and their clinical implications have yet to be investigated and compared with those of established topographers. We have recently reported cases of forme fruste keratoconus12 and central corneal scar13 in which the Cassini device appeared to be more sensitive than established cornea imaging devices. The aim of the work reported here was to evaluate the repeatability measurement of this newly introduced device in a large number of LASIK-treated eyes, keratoconic eyes, and eyes treated with anterior-surface normalization and cross-linking. Specifically, we investigated the repeatability of measurement of the following parameters: flat and steep meridian keratometry, axis of astigmatism, form factor, surface regularity index (SRI), and surface asymmetry index (SAI).

Materials and methods

This observational, prospective study received approval by the ethics committee of our institution, adherent to the tenets of the Declaration of Helsinki. Informed written consent was obtained from each subject at the time of the first clinical visit.

Inclusion criteria

The study group consisted of 373 different patients. One eye was selected randomly from each patient to be included in the study. The following study groups were formed: Group A (nA=51 different eyes) consisted of eyes that had received myopic laser-assisted in situ keratomileusis (LASIK) treatment. Group B (nB=95) consisted of untreated keratoconic eyes; the keratoconus condition had been previously determined as being grade 1 or more on the Amsler–Kruemeich keratoconus severity scale, as determined by Scheimpflug imaging. Group C (nC=47) consisted of eyes treated with partial anterior-surface topography-guided excimer ablation and same-session high-fluence collagen cross-linking, in a procedure known as the “Athens Protocol”.14 In addition, a control Group D (nD=180) was formed consisting of subjects having no current or past ocular pathology other than refractive error, no other previous surgery, no epithelial defects, and no present irritation or dry-eye disorder, all confirmed by a complete ophthalmologic evaluation.

Instrumentation

The system employed in this study was the Cassini, a multicolored-spot reflection corneal topographer. The system projects approximately 700 LED point-sources onto the cornea and evaluates their reflection pattern, extending to up to 8.5 mm of the corneal diameter area as raw data. These spots are grouped in seven “septima” segments with a specific Cartesian array arrangement within each segment, as illustrated in Figure 1.
Figure 1

Raw data image as obtained by the Cassini multicolored-spot reflection topographer.

Notes: The septima sections are numbered sequentially, along with a cross indicating the prime orientation of the 100 spots within the section. The red dashed line indicates the steep axis; the blue dotted line, the flat axis.

The system consequently provides anterior-surface topography results, including axial and tangential curvature, refractive power, and elevation maps. The user may select from among various color-coded scales (Figure 2), including the American Style, the ISO (International Organization for Standardization [French]) 19980, the Oculus, and the topographic modeling system scale.15 The system calculates flat and steep keratometry (diopters [D]), axis orientation (°), and related astigmatism (D), and identifies the location of the corneal apex. Four topographic indices16 relating to surface asphericity, and three keratoconus indices – form factor, the SRI, and SAI (created by Dr Stephen Klyce)17 – are also provided. Acquisition quality was evaluated by four image quality indices. The system employed in the study ran on software version 1.2.0 (updated September 2013).
Figure 2

Axial curvature maps.

Notes: In this example from the keratoconic group C, the same axial map is illustrated in three different color scales, according to (A) the oculus, (B) the KlyceUSS, and (C) the Normalized ISO 19980 color scale.

Abbreviations: D, diopters; OD, optical density.

Data collection and analysis

Three successive acquisitions, centered on the pupil, were performed in each case (eye), assuring that each acquisition had at least 75% coverage, as reported by the quality factor. In each acquisition, the data collected were: flat and steep keratometry (in diopters, D), axis of steep keratometry (in degrees, °), form factor, and the indices SRI and SAI (both unit-less). The repeatability of measurement was assessed by the standard deviation of the three values of each parameter investigated as obtained by the three consecutive acquisitions performed in each eye. Statistical analysis was performed using Minitab software (v 16.2.3; MiniTab Ltd, Coventry, UK).

Results

Post-myopic LASIK Group A, of the 51 eyes included, there were 24 male (47%) and 27 female (53%). Mean subject age at the time of examination was 35.05±8.11 (24 to 49) years of age. Mean value of flat keratometry was 38.20±4.50 (23.93 to 45.57) D, of steep keratometry was 39.50±4.23 (29.28 to 47.04) D, of steep axis was 87.78±42.09° (4.42 to 179.75°), of form factor was 2.62±2.18 (0.07 to 10.61), of SRI was 1.60±0.78 (−0.11 to 3.52), and of SAI was 1.01±0.98 (0.12 to 6.31). Descriptive statistics of these values as well as of those from Groups B, C, and control Group D are reported in Table 1.
Table 1

Descriptive statistics of flat and steep keratometry (diopters [D]), axis (°), form factor, surface regularity index (SRI), and surface asymmetry index (SAI) measurements for the control group and the three study groups

GroupK flat (D)K steep (D)Steep axis (°)Form factor (k)SRISAI
Control
 Mean42.8344.3591.850.960.910.61
 SD2.022.1727.150.630.620.65
 Min32.3633.480.15−0.20−0.46−0.11
 Max50.9454.01179.866.303.015.38
Group A (LASIK)
 Mean38.2039.5087.782.621.601.01
 SD4.504.2342.092.180.780.98
 Min23.9325.284.420.07−0.110.12
 Max45.5747.04179.7510.613.526.31
Group B (keratoconus)
 Mean42.4545.7193.852.961.882.96
 SD3.903.6546.781.960.712.89
 Min31.0039.491.15−0.05−0.260.10
 Max53.9858.07179.7910.473.6020.47
Group C (Athens Protocol)
 Mean43.2946.3995.512.422.093.02
 SD4.224.7239.431.910.632.33
 Min34.7935.4313.00−0.840.130.18
 Max53.7557.52179.237.273.2510.31

Abbreviations: LASIK, laser-assisted in situ keratomileusis; max, maximum; min, minimum; SD, standard deviation; K, keratometry; D, diopters.

The keratoconic Group B consisted of 95 eyes (55.7% male; eye laterality 47.3% right). Mean patient age was 30.72±7.98 (17 to 45) years of age. Mean value of flat keratometry was 42.45±3.90 (31.00 to 53.98) D, of steep keratometry was 45.71±3.65 (39.49 to 58.07) D, of steep axis was 93.85±46.78° (1.15 to 179.79°), of form factor was 2.96±1.96 (−0.05 to 10.47), of SRI was 1.88±0.71 (−0.26 to 3.60), and of SAI was 2.96±2.89 (0.10 to 20.47). The cross-linked Group C consisted of 47 eyes, belonging to 29 male and 11 female patients; 25 eyes were right and 22 left. Mean patient age was 27.45±7.46 (17 to 49) years of age. Mean value of flat keratometry was 43.29±4.22 (34.79 to 53.75) D, of steep keratometry was 46.39±4.72 (35.43 to 57.52) D, of steep axis was 92.51±39.43° (13.00 to 179.23°), of form factor was 2.42±1.91 (−0.84 to 7.27), of SRI was 2.09±0.63 (0.13 to3.25), and of SAI was 3.02±2.33 (0.18 to 10.31). The control Group D consisted of 180 eyes, belonging to 55 male and 52 female patients; 92 eyes were right and 88 left. Mean patient age was 40.04±16.86 (16 to 61) years of age. Mean value of flat keratometry was 42.83±2.02 (32.36 to 50.94) D, of steep keratometry was 44.35±2.17 (33.48 to 54.01) D, of steep axis was 91.85±27.15° (0.15 to 179.86°), of form factor was 0.96±0.63 (−0.20 to 6.30), of SRI was 0.91±0.62 (−0.46 to 3.01), and of SAI was 0.61±0.65 (−0.11 to 5.38). Specifically, flat keratometry repeatability was 0.74±0.89 (0.03 to 5.26) D in the LASIK Group A, 0.88±1.45 (0.00 to 7.84) D in the keratoconic Group B, and 0.71±0.94 (0.02 to 6.23) D in the cross-linked Group C. The control Group D had flat keratometry repeatability of 0.36±0.46 (0.00 to 2.71) D. Steep keratometry repeatability was 0.64±0.82 (0.01 to 4.81) D in the LASIK Group A, 0.89±1.22 (0.02 to 7.85) D in the keratoconic Group B, and 0.93±1.12 (0.04 to 5.93) D in Group C. The control Group D had steep keratometry repeatability of 0.41±0.50 (0.00 to 3.51) D. Axis repeatability was 3.45±1.62° (0.38 to 7.78°) for the LASIK Group A, 4.12±3.17° (0.02 to 12.13°) for the keratoconic Group B, and 3.20±1.99° (0.17 to 8.61°) for the Athens Protocol Group C. The control Group D had axis repeatability of 2.16±1.39° (0.05 to 5.86°). Figure 3 presents box plots illustrating steep and flat keratometry and axis measurement repeatability. The repeatability results are also reported in Table 2.
Figure 3

Box plots illustrating steep and flat keratometry and axis measurement repeatability.

Notes: Group A, laser-assisted in situ keratomileusis (LASIK); Group B, keratoconus (KCN); Group C, Athens Protocol (AP); and Group D, control. Numerical indicators correspond to median values. (A) Comparative flat keratometry repeatabilty in all groups; (B) comparative steep keratometry repeatability in all groups; (C) comparative axis measurement repeatability in all groups.

Abbreviation: D, diopters.

Table 2

Repeatability of flat and steep keratometry (diopters [D]), axis (°), form factor, surface regularity index (SRI), and surface asymmetry index (SAI) measurements for the control group and the three study groups

GroupK flat (D)K steep (D)Steep axis (°)Form factor (k)SRISAI
Control
 Mean0.360.412.160.120.230.26
 SD0.460.501.390.150.160.30
 Min0.000.000.050.000.000.00
 Max2.713.515.861.060.752.39
Group A (LASIK)
 Mean0.740.643.450.290.220.33
 SD0.890.821.620.320.170.40
 Min0.030.010.380.000.010.01
 Max5.264.817.781.690.962.31
Group B (keratoconic)
 Mean0.880.894.120.420.200.39
 SD1.451.223.170.420.180.75
 Min0.000.020.020.000.000.00
 Max7.847.8512.132.031.077.15
Group C (Athens Protocol)
 Mean0.710.933.200.460.130.43
 SD0.941.121.990.500.090.56
 Min0.020.040.170.020.000.05
 Max6.235.938.611.850.453.50

Abbreviations: LASIK, laser-assisted in situ keratomileusis; max, maximum; min, minimum; SD, standard deviation; K, keratometry; D, diopters.

The form factor measurement repeatability was 0.29±0.32 (0.00 to 1.69) in the LASIK Group A, 0.42±0.42 (0.00 to 2.03) in the keratoconic Group B, and 0.46±0.50 (0.02 to 1.85) in the cross-linked Group C. The control Group D had form factor repeatability of 0.12±0.15 (0.00 to 1.06) (Figure 4). The SAI index measurement repeatability was 0.33±0.40 (0.01 to 2.31) in the LASIK Group A, 0.39±0.75 (0.00 to 7.15) in the keratoconic Group B, and 0.43±0.56 (0.05 to 3.50) in the cross-linked Group C. The control group had SAI measurement repeatability of 0.26±0.30 (0.00 to 2.39). The SRI index repeatability was 0.22±0.17 (0.01 to 0.96) for LASIK Group A, 0.20±0.18 (0.00 to 1.07) in keratoconic Group B, and 0.13±0.09 (0.00 to 0.45) in the cross-linked Group C. The control Group D had SRI measurement repeatability of 0.23±0.16 (0.00 to 0.75).
Figure 4

Box plots illustrating form factor, surface asymmetry index (SAI), and surface regularity index (SRI) measurement repeatability.

Notes: Group A, laser-assisted in situ keratomileusis (LASIK); Group B, keratoconus (KCN); Group C, Athens Protocol (AP); and Group D, control. Numerical indicators correspond to median values. (A) Comparative form factor repeatability in all groups; (B) comparative SAI repeatability in all groups; (C) comparative SRI repeatability in all groups.

Discussion

There are many advantages justifying the widespread use of Placido-ring based topography systems, such as the noncontact measurement of the single-shot capture, which reduces motion artifacts. However, there is a number of limitations associated with Placido topography, including the skew-ray error,18,19 lower data reliability at the corneal center, and susceptibility to error in areas of abrupt corneal elevation changes.20,21 Specifically, when reconstructing the anterior corneal surface, the numerical algorithms employed in commercially available Placido-disc topographers neglect skew-ray reflections,22 leading to inaccuracy in reconstructing non-rotationally symmetric corneas.23 The new modality introduced by the Cassini topographer aims to avoid these two critical limitations. The work presented here is an investigation of this novel device which is currently under evaluation by our team. We chose to focus on a very specific metric, that of the repeatability of measurement in various groups of several clinical settings. We have already reported12,13 early clinical case reports of this device in specific challenging cases in which other established devices have failed to properly image the cornea. In the present study, we observed the distributions and repeatability of keratometric, axis, and three qualitative indices associated with anterior corneal surface. The distributions measured in the control group were of the same range reported in the literature,24 the keratometries encountered in the keratoconic- and Athens Protocol-treated eyes were also within the same range reported in previous work by our team employing Placido topography21 and Scheimpflug imaging technologies.25 The anterior-surface indices were also within the range of SAI and SRI measurements reported in keratoconic eyes.26 The keratometric repeatability was under 0.5 D in the control group, 0.7–0.9 D in the keratoconic and cross-linked eyes, and 0.65–0.75 D in the LASIK-treated eyes. Axis measurement repeatability was under 2.5° in the control group, in the range of under 4° in the keratoconic and cross-linked eyes, and approximately 3.5° in the LASIK-treated eyes. Obviously, the sensitivity to centration in the keratoconic and LASIK-treated eyes could be a reason for the larger fluctuation among values, in comparison to the control group. In the LASIK-treated group, the surface analysis was also complex compared with the spherical corneas encountered in the control group, due to the central flattening as a result of the excimer ablation performed. Comparison with the values reported in the literature suggests that the findings herein are quite comparable to established corneal-screening devices such as standard manual keratometry,27,28 Placido topography,29 the Orbscan II (Bausch & Lomb, Rochester, NY, USA), the Pentacam (Oculus, Weltzlar, Germany), the Galilei (Ziemer Ophthalmic Systems AG, Port, Switzerland),30 and the Sirius (CSO Costruzione Strumenti Oftalmici, Scandicci, Italy).31 For example, the standard deviation of the test–retest reliability in the size of corneal astigmatism was 0.293 D as reported in automated keratometry measurements.32

Conclusion

The Cassini demonstrated extreme measurement precision in terms of sensitivity and specificity. It appears to precisely measure irregular corneas, both in keratometry and axis. Perhaps more importantly it appears to precisely measure normal and even post-myopic LASIK corneas, expected to have minimal astigmatism which poses a challenging diagnostic measurement.
  30 in total

1.  The Universal Standard Scale: proposed improvements to the American National Standards Institute (ANSI) scale for corneal topography.

Authors:  Michael K Smolek; Stephen D Klyce; Jeffery K Hovis
Journal:  Ophthalmology       Date:  2002-02       Impact factor: 12.079

Review 2.  Evaluation of optical quality of corneas using corneal topographers.

Authors:  Naoyuki Maeda
Journal:  Cornea       Date:  2002-10       Impact factor: 2.651

Review 3.  Measuring the cornea: the latest developments in corneal topography.

Authors:  Tracy Swartz; Lisa Marten; Ming Wang
Journal:  Curr Opin Ophthalmol       Date:  2007-07       Impact factor: 3.761

4.  Clinical evaluation of keratometry and computerised videokeratography: intraobserver and interobserver variability on normal and astigmatic corneas.

Authors:  C H Karabatsas; S D Cook; J Papaefthymiou; P Turner; J M Sparrow
Journal:  Br J Ophthalmol       Date:  1998-06       Impact factor: 4.638

5.  Keratoconus management: long-term stability of topography-guided normalization combined with high-fluence CXL stabilization (the Athens Protocol).

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  J Refract Surg       Date:  2014-02       Impact factor: 3.573

6.  Comparison and repeatability of keratometric and corneal power measurements obtained by Orbscan II, Pentacam, and Galilei corneal tomography systems.

Authors:  Alexandra Z Crawford; Dipika V Patel; Charles N J McGhee
Journal:  Am J Ophthalmol       Date:  2013-03-28       Impact factor: 5.258

7.  Comparison of keratometric and topographic cylinder and axis measurements on normal corneas with low astigmatism.

Authors:  C H Karabatsas; I Papaefthymiou; I M Aslanides; D Z Chatzoulis
Journal:  Eur J Ophthalmol       Date:  2005 Jan-Feb       Impact factor: 2.597

8.  [Simulation of toric intraocular lens results: the effect of repeatability and increments of automated keratometry].

Authors:  Yoshihiko Ninomiy; Yayoi Kanazawa; Yoshinao Kojima; Naoyuki Maeda
Journal:  Nippon Ganka Gakkai Zasshi       Date:  2013-08

9.  Clinical Correlation between Placido, Scheimpflug and LED Color Reflection Topographies in Imaging of a Scarred Cornea.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  Case Rep Ophthalmol       Date:  2014-10-01

10.  Forme Fruste Keratoconus Imaging and Validation via Novel Multi-Spot Reflection Topography.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  Case Rep Ophthalmol       Date:  2013-10-25
View more
  6 in total

1.  Placido disk-based topography versus high-resolution rotating Scheimpflug camera for corneal power measurements in keratoconic and post-LASIK eyes: reliability and agreement.

Authors:  Rachele R Penna; Ugo de Sanctis; Martina Catalano; Luca Brusasco; Federico M Grignolo
Journal:  Int J Ophthalmol       Date:  2017-03-18       Impact factor: 1.779

2.  Validation of corneal topographic and aberrometric measurements obtained by color light-emitting diode reflection topography in healthy eyes.

Authors:  David P Piñero; Ainhoa Molina-Martín; Vicent J Camps; Dolores de Fez; María Teresa Caballero
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2019-09-03       Impact factor: 3.117

3.  Effect of the posterior corneal surface on total corneal astigmatism in patients with age-related cataract.

Authors:  Ying-Ying Jin; Zhen Zhou; Xiao-Yong Yuan; Hui Song; Xin Tang
Journal:  Int J Ophthalmol       Date:  2018-06-18       Impact factor: 1.779

4.  Scheimpflug vs Scanning-Slit Corneal Tomography: Comparison of Corneal and Anterior Chamber Tomography Indices for Repeatability and Agreement in Healthy Eyes.

Authors:  Anastasios John Kanellopoulos
Journal:  Clin Ophthalmol       Date:  2020-09-04

5.  A novel color-LED corneal topographer to assess astigmatism in pseudophakic eyes.

Authors:  Tiago B Ferreira; Filomena J Ribeiro
Journal:  Clin Ophthalmol       Date:  2016-08-13

6.  Comparability and repeatability of different methods of corneal astigmatism assessment.

Authors:  Tiago B Ferreira; Filomena J Ribeiro
Journal:  Clin Ophthalmol       Date:  2017-12-20
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.