Literature DB >> 16863970

Anterior chamber depth measurements in phakic and pseudophakic eyes: Pentacam versus ultrasound device.

Gabor Nemeth1, Attila Vajas, Bence Kolozsvari, Andras Berta, Laszlo Modis.   

Abstract

PURPOSE: To compare anterior chamber depth (ACD) measurements with a new optical device with those taken with a standard ultrasound (US) device in emmetropic phakic and pseudophakic eyes.
SETTING: Department of Ophthalmology, Medical Health and Science Center, University of Debrecen, Debrecen, Hungary.
METHODS: Forty-two phakic and 42 pseudophakic patients with normal axial lengths (mean 22.91 mm +/- 1.21 [SD]) were enrolled in the study. The ACD was measured 3 times with Scheimpflug-based Pentacam (Oculus) and then 3 times with a standard A-scan US device (AL-2000, Tomey). The data were then analyzed.
RESULTS: In the phakic group, the mean ACD was 2.87 +/- 0.4 mm with the Pentacam and 2.89 +/- 0.49 mm with ultrasound A-scan (US) (P = .84). In the pseudophakic group, the mean ACD was 3.41 +/- 0.28 mm and 3.97 +/- 0.45 mm, respectively (P < .001). The correlation between measurements was significant in both the phakic and pseudophakic groups (r = .547/P < .001 and r = .404/P = .01, respectively).
CONCLUSIONS: In phakic eyes, ACD measured with the Pentacam and with US was the same. However, in pseudophakic eyes, the difference was significantly lower when the ACD was measured with the Pentacam. Therefore, in pseudophakic patients, further evaluation of ACD data with the Scheimpflug-based system is necessary.

Entities:  

Mesh:

Year:  2006        PMID: 16863970     DOI: 10.1016/j.jcrs.2006.02.057

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  11 in total

1.  Anterior chamber depth in normal subjects by rotating scheimpflug imaging.

Authors:  Matthew T Feng; Michael W Belin; Renato Ambrósio; Satinder P S Grewal; Wang Yan; Mohamed S Shaheen; Charles McGhee; Naoyuki Maeda; Tobias H Neuhann; H Burkhard Dick; Saleh A Alageel; Andreas Steinmueller
Journal:  Saudi J Ophthalmol       Date:  2011-04-30

2.  Comparison of central corneal thickness measurements by Orbscan II and Pentacam after corneal refractive surgery.

Authors:  Jumpei Matsuda; Osamu Hieda; Shigeru Kinoshita
Journal:  Jpn J Ophthalmol       Date:  2008-09-05       Impact factor: 2.447

3.  Repeatability and comparability of anterior segment biometry obtained by the Sirius and the Pentacam analyzers.

Authors:  Paola De la Parra-Colín; Manuel Garza-León; Tonatiuh Barrientos-Gutierrez
Journal:  Int Ophthalmol       Date:  2013-04-25       Impact factor: 2.031

Review 4.  Device interchangeability on anterior chamber depth and white-to-white measurements: a thorough literature review.

Authors:  Alberto Domínguez-Vicent; Cari Pérez-Vives; Teresa Ferrer-Blasco; Santiago García-Lázaro; Robert Montés-Micó
Journal:  Int J Ophthalmol       Date:  2016-07-18       Impact factor: 1.779

5.  [Precision of a new device for biometric measurements in pseudophakic eyes].

Authors:  A L Hildebrandt; G U Auffarth; M P Holzer
Journal:  Ophthalmologe       Date:  2011-08       Impact factor: 1.059

Review 6.  Quantitative evaluation of anterior segment parameters in the era of imaging.

Authors:  Syril Dorairaj; Jeffrey M Liebmann; Robert Ritch
Journal:  Trans Am Ophthalmol Soc       Date:  2007

7.  Central corneal thickness and anterior chamber depth measurement by Sirius(®) Scheimpflug tomography and ultrasound.

Authors:  J Jorge; Jl Rosado; Ja Díaz-Rey; Jm González-Méijome
Journal:  Clin Ophthalmol       Date:  2013-02-26

8.  Anterior segment parameters using Pentacam and prediction of corneal endothelial cell loss after cataract surgery.

Authors:  Yang Kyeung Cho; Hwa Seok Chang; Tae Yoon La; Donghyun Ji; Hyunkyung Kim; Jin A Choi; Man Soo Kim
Journal:  Korean J Ophthalmol       Date:  2010-10-05

9.  Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster.

Authors:  Roma P Patel; Rahul T Pandit
Journal:  J Ophthalmol       Date:  2012-01-19       Impact factor: 1.909

10.  Evaluation of Anterior Segment Parameter Changes Using the Sirius after Uneventful Phacoemulsification.

Authors:  Ali Şimşek; Burak Bilgin; Musa Çapkın; Şemsettin Bilak; Mete Güler; Ali Hakim Reyhan
Journal:  Korean J Ophthalmol       Date:  2016-07-21