| Literature DB >> 21457508 |
Jamal A Alhashemi1, Maurizio Cecconi, Christoph K Hofer.
Abstract
Entities:
Mesh:
Year: 2011 PMID: 21457508 PMCID: PMC3219410 DOI: 10.1186/cc9996
Source DB: PubMed Journal: Crit Care ISSN: 1364-8535 Impact factor: 9.097
Factors affecting selection of cardiac output monitoring devices
| Factor groups | Examples |
|---|---|
| Institution | Type of institution |
| Availability of monitoring techniques | |
| Level of standardization | |
| Potential of integration into existing monitoring systems | |
| Level of experience | |
| Devices | Invasiveness |
| Handling | |
| Technical limitations | |
| Validity, accuracy & repeatability | |
| Availability of additional hemodynamic information | |
| Patient | Severity of specific diseases |
| Heart rhythm | |
| Contraindications | |
| Type of intervention | |
| Type of treatment protocol |
Figure 1Overview of cardiac output monitoring techniques. PAC: pulmonary artery catheter.
Overview of hemodynamic monitoring techniques
| Additional variables | |||||||
|---|---|---|---|---|---|---|---|
| Groups | Examples | Features | Invasiveness | Continuous CO | Static | Dynamic | SvO2/ScvO2 |
| Vigilance™ | Right heart catheterization | ■ ■ ■ | Response time up to 12 minutes | CVP PCWP | - | Specific catheter for continuous measurement available | |
| Calibrated | PiCCO | Thermistor-tipped arterial catheter Central venous line | ■ ■ | Response time 3 seconds | CVP GEDV EVLW | SVV PPV | Specific catheter for continuous Measurement available |
| LiDCO | Lithium dilution set | ■□ | Beat-by-beat | - | SVV PPV | - | |
| EV1000™/ | Thermistor-tipped | ■ ■ | NA | CVP | SVV | Specific catheter | |
| VolumeView™* | arterial catheter | GEDV | for continuous | ||||
| Central venous line | EVLW | Measurement available | |||||
| Uncalibrated | FloTrac/Vigileo™ | Specific arterial pressure sensor | ■ | Response time 20 seconds | - | SVV | Specific catheter for continuous Measurement available |
| LiDCO | Regular arterial line | ■ | Beat-by-beat | - | SVV PPV | - | |
| PulsioFlex™* | Regular arterial line Specific sensor | ■ | NA | - | SVV PPV | Specific catheter for continuous Measurement available | |
| PRAM MostCare® | Specific arterial kit | ■ | Beat-by-beat | - | SVV PPV | - | |
| Nexfin™HD | Specific pressure | □ | Beat-by-beat | - | - | - | |
| sensors | |||||||
| TE | CardioQ™ | Esophageal | ■ | Limitation: | - | - | - |
| Flowprobe | probe | ||||||
| positioning | |||||||
| TT | USCOM™ | Flowprobe | ■ | Intermittent | - | - | - |
| Partial CO2 rebreathing | NiCO™ | Rebreating loop | ■ | Up-date every 3' | - | - | - |
| Dye dilution | DDG analyzer® | Specific sensor | ■ | Intermittent | - | - | - |
| Endotracheal bioimpedance | ECOM™ | Specific endotracheal tube, arterial line | ■ | Continuous | - | - | - |
| Thoracic/whole body bioimpedance | BioZ® | Specific electrodes | □ | Continuous | - | - | - |
| Thoracic bioreactance | NICOM™ | Specific electrodes | □ | Continuous | - | SVV | - |
CO: cardiac output; CVP: central venous pressure; EVLW: extravascular lung water; GEDV: global end-diastolic volume; NA: technical specifications not yet available; PAC: pulmonary artery catheter; PAOP: pulmonary artery occlusion pressure; PPV: pulse pressure variation; SvO2: mixed venous oxygen saturation; ScvO2; central venous oxy gen saturation; SVV; stroke volume variation; TE: transesophageal; TT: transthoracic; *not yet available.
Figure 2Integrative concept for the use of cardiac output monitoring devices. ED: emergency department; HD: hemodynamic; ICU: intensive care unit; OR: operating room; PAC: pulmonary artery catheter.