Comparison of a supra-sternal cardiac output monitor (USCOM) with the pulmonary artery catheter.

O Thom1, D M Taylor, R E Wolfe, J Cade, P Myles, H Krum, R Wolfe.   

Abstract

BACKGROUND: Cardiac output (CO) cannot reliably be estimated by clinical examination. We aimed to measure the agreement between CO measurements using a supra-sternal Doppler monitor (USCOM, Coffs Harbour, Australia) and the pulmonary artery catheter (PAC).
METHODS: The study was conducted in the intensive care unit of a tertiary teaching hospital. All patients with PAC in situ were eligible. Simultaneous CO readings were taken when clinically indicated. Investigators and clinicians were blinded to each other's results. The CO values used were the mean of three consecutive supra-sternal Doppler readings for patients with a sinus rhythm and seven for atrial fibrillation, and the mean of three thermodilution curves with acceptable form and values within 10% of each other for the PAC. Agreement was measured using both the paired t-test to calculate bias and limits of agreement and the intraclass correlation (ICC) coefficient.
RESULTS: Ninety-four subjects were enrolled. From 89 subjects, 250 paired comparisons were obtained. USCOM monitor readings were unobtainable in five patients. Mean supra-sternal Doppler CO was 5.5 litre min(-1). Bias was -0.09 litre min(-1) and levels of agreement were +/- 2.92 litre min(-1) when compared with PAC. ICC was 0.46 (95% CI 0.36-0.56), and mean percentage difference was 19 (IQR 6-31)%.
CONCLUSIONS: In our subjects, there was poor agreement between CO measurements done with the supra-sternal Doppler monitor and PAC.

Entities:  

Mesh:

Year:  2009        PMID: 19864307     DOI: 10.1093/bja/aep296

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  15 in total

1.  In vitro evaluation of an ultrasonic cardiac output monitoring (USCOM) device.

Authors:  Shaun D Gregory; Helena Cooney; Sara Diab; Chris Anstey; Ogilvie Thom; John F Fraser
Journal:  J Clin Monit Comput       Date:  2015-03-07       Impact factor: 2.502

Review 2.  [Meta-analyses on measurement precision of non-invasive hemodynamic monitoring technologies in adults].

Authors:  G Pestel; K Fukui; M Higashi; I Schmidtmann; C Werner
Journal:  Anaesthesist       Date:  2018-06       Impact factor: 1.041

3.  Cardiac index measurements by transcutaneous Doppler ultrasound and transthoracic echocardiography in adult and pediatric emergency patients.

Authors:  H Bryant Nguyen; Daryl P Banta; Gail Stewart; Tommy Kim; Ramesh Bansal; James Anholm; William A Wittlake; Stephen W Corbett
Journal:  J Clin Monit Comput       Date:  2010-06-20       Impact factor: 2.502

4.  Capability of a new paediatric oesophageal Doppler monitor to detect changes in cardiac output during testing of external pacemakers after cardiac surgery.

Authors:  Thilo Fleck; Stephan Schubert; Brigitte Stiller; Matthias Redlin; Peter Ewert; Nicole Nagdyman; Felix Berger
Journal:  J Clin Monit Comput       Date:  2011-11-12       Impact factor: 2.502

Review 5.  Cardiac output monitoring: an integrative perspective.

Authors:  Jamal A Alhashemi; Maurizio Cecconi; Christoph K Hofer
Journal:  Crit Care       Date:  2011-03-22       Impact factor: 9.097

6.  Pulmonary Artery Catheter (PAC) Accuracy and Efficacy Compared with Flow Probe and Transcutaneous Doppler (USCOM): An Ovine Cardiac Output Validation.

Authors:  Robert A Phillips; Sally G Hood; Beverley M Jacobson; Malcolm J West; Li Wan; Clive N May
Journal:  Crit Care Res Pract       Date:  2012-05-09

Review 7.  Perioperative cardiovascular monitoring of high-risk patients: a consensus of 12.

Authors:  Jean-Louis Vincent; Paolo Pelosi; Rupert Pearse; Didier Payen; Azriel Perel; Andreas Hoeft; Stefano Romagnoli; V Marco Ranieri; Carole Ichai; Patrice Forget; Giorgio Della Rocca; Andrew Rhodes
Journal:  Crit Care       Date:  2015-05-08       Impact factor: 9.097

8.  The correlation between the first heart sound and cardiac output as measured by using digital esophageal stethoscope under anaesthesia.

Authors:  Young Duck Shin; Kyoung Hoon Yim; Sang Hi Park; Yong Wook Jeon; Jin Ho Bae; Tae Soo Lee; Myoung Hwan Kim; Young Jin Choi
Journal:  Pak J Med Sci       Date:  2014-03       Impact factor: 1.088

9.  Influence of different positions on hemodynamics derived from noninvasive transcutaneous Doppler ultrasound.

Authors:  Cangel Pui-Yee Chan; Pui-Ling Cheung; Mandy Man Tse; Nandini Agarwal; Sangeeta Narain; Stewart Siu-Wa Chan; Brendan E Smith; Colin A Graham; Timothy H Rainer
Journal:  Physiol Rep       Date:  2013-09-20

10.  Continuous stroke volume estimation from aortic pressure using zero dimensional cardiovascular model: proof of concept study from porcine experiments.

Authors:  Shun Kamoi; Christopher Pretty; Paul Docherty; Dougie Squire; James Revie; Yeong Shiong Chiew; Thomas Desaive; Geoffrey M Shaw; J Geoffrey Chase
Journal:  PLoS One       Date:  2014-07-17       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.