Quantitative megavoltage radiation therapy dosimetry using the storage phosphor KCl: Eu2+.

Zhaohui Han1, Joseph P Driewer, Yuanshui Zheng, Daniel A Low, H Harold Li.   

Abstract

This work, for the first time, reports the use of europium doped potassium chloride (KCl:Eu2+) storage phosphor for quantitative megavoltage radiation therapy dosimetry. In principle, KCl:Eu2+ functions using the same photostimulatated luminescence (PSL) mechanism as commercially available BaFBr0.85I0.15:Eu2+ material that is used for computed radiography (CR) but features a significantly smaller effective atomic number--18 versus 49--making it a potentially useful material for nearly tissue-equivalent radiation dosimetry. Cylindrical KCl:Eu2+ dosimeters, 7 mm in diameter and 1 mm thick, were fabricated in-house. Dosimetric properties, including radiation hardness, response linearity, signal fading, dose rate sensitivity, and energy dependence, were studied with a laboratory optical reader after irradiation by a linear accelerator. The overall experimental uncertainty was estimated to be within +/-2.5%. The findings were (1) KCl:Eu2+ showed satisfactory radiation hardness. There was no significant change in the stimulation spectra after irradiation up to 200 Gy when compared to a fresh dosimeter, indicating that this material could be reused at least 100 times if 2 Gy per use was assumed, e.g., for patient-specific IMRT QA. (2) KCl:Eu2+ exhibited supralinear response to dose after irradiation from 0 to 800 cGy. (3) After x ray irradiation, the PSL signal faded with time and eventually reached a fading rate of about 0.1 % /h after 12 h. (4) The sensitivity of the dosimeter was independent of the dose rate ranging from 15 to 1000 cGy/min. (5) The sensitivity showed no beam energy dependence for either open x ray or megavoltage electron fields. (6) Over-response to low-energy scattered photons was comparable to radiographic film, e.g., Kodak EDR2 film. By sandwiching dosimeters between low-energy photon filters (0.3 mm thick lead foils) during irradiation, the over-response was reduced. The authors have demonstrated that KCl:Eu2+ dosimeters have many desirable dosimetric characteristics that make the material conducive to radiation therapy dosimetry. In the future, a large-area KCl:Eu2+-based CR plate with a thickness of the order of a few microns, created using modern thin film techniques, could provide a reusable, quantitative, high-resolution two-dimensional dosimeter with minimal energy dependence.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19746808      PMCID: PMC2724175          DOI: 10.1118/1.3171687

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  28 in total

1.  The use of radiographic film for linear accelerator stereotactic radiosurgical dosimetry.

Authors:  J L Robar; B G Clark
Journal:  Med Phys       Date:  1999-10       Impact factor: 4.071

Review 2.  Intensity-modulated radiotherapy: current status and issues of interest.

Authors: 
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-11-15       Impact factor: 7.038

3.  Comparative evaluation of Kodak EDR2 and XV2 films for verification of intensity modulated radiation therapy.

Authors:  Nesrin Dogan; Leonid B Leybovich; Anil Sethi
Journal:  Phys Med Biol       Date:  2002-11-21       Impact factor: 3.609

4.  Investigation of Kodak extended dose range (EDR) film for megavoltage photon beam dosimetry.

Authors:  Indrin J Chetty; Paule M Charland
Journal:  Phys Med Biol       Date:  2002-10-21       Impact factor: 3.609

5.  Dosimetry of therapeutic photon beams using an extended dose range film.

Authors:  Jacqueline Esthappan; Sasa Mutic; William B Harms; James F Dempsey; Daniel A Low
Journal:  Med Phys       Date:  2002-10       Impact factor: 4.071

Review 6.  The physics of computed radiography.

Authors:  J A Rowlands
Journal:  Phys Med Biol       Date:  2002-12-07       Impact factor: 3.609

7.  Evaluation of Kodak EDR2 film for dose verification of intensity modulated radiation therapy delivered by a static multileaf collimator.

Authors:  X R Zhu; P A Jursinic; D F Grimm; F Lopez; J J Rownd; M T Gillin
Journal:  Med Phys       Date:  2002-08       Impact factor: 4.071

8.  Dosimetric performance of an enhanced dose range radiographic film for intensity-modulated radiation therapy quality assurance.

Authors:  Arthur J Olch
Journal:  Med Phys       Date:  2002-09       Impact factor: 4.071

9.  TG-69: radiographic film for megavoltage beam dosimetry.

Authors:  Sujatha Pai; Indra J Das; James F Dempsey; Kwok L Lam; Thomas J Losasso; Arthur J Olch; Jatinder R Palta; Lawrence E Reinstein; Dan Ritt; Ellen E Wilcox
Journal:  Med Phys       Date:  2007-06       Impact factor: 4.071

10.  Intensity-modulated radiotherapy.

Authors:  Steven A Leibel; Zvi Fuks; Michael J Zelefsky; Suzanne L Wolden; Kenneth E Rosenzweig; Kaled M Alektiar; Margie A Hunt; Ellen D Yorke; Linda X Hong; Howard I Amols; Chandra M Burman; Andrew Jackson; Gikas S Mageras; Thomas LoSasso; Laura Happersett; Spiridon V Spirou; Chen-Shou Chui; C Clifton Ling
Journal:  Cancer J       Date:  2002 Mar-Apr       Impact factor: 3.360

View more
  10 in total

1.  Theoretical and empirical investigations of KCl : Eu2+ for nearly water-equivalent radiotherapy dosimetry.

Authors:  Yuanshui Zheng; Zhaohui Han; Joseph P Driewer; Daniel A Low; H Harold Li
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

2.  Evaluation of the absorbed dose to the breast using radiochromic film in a dedicated CT mammotomography system employing a quasi-monochromatic x-ray beam.

Authors:  Dominic J Crotty; Samuel L Brady; D'Vone C Jackson; Greta I Toncheva; Colin E Anderson; Terry T Yoshizumi; Martin P Tornai
Journal:  Med Phys       Date:  2011-06       Impact factor: 4.071

3.  Radiation hardness of the storage phosphor europium doped potassium chloride for radiation therapy dosimetry.

Authors:  Joseph P Driewer; Haijian Chen; Andres Osvet; Daniel A Low; H Harold Li
Journal:  Med Phys       Date:  2011-08       Impact factor: 4.071

4.  The role of activator concentration and precipitate formation on optical and dosimetric properties of KCl:Eu(2+) storage phosphor detectors.

Authors:  Rachael A Hansel; Zhiyan Xiao; Yanle Hu; Olga Green; Deshan Yang; H Harold Li
Journal:  Med Phys       Date:  2013-09       Impact factor: 4.071

5.  Temporal signal stability of KCl:Eu(2+) storage phosphor dosimeters.

Authors:  Zhiyan Xiao; Rachael Hansel; Haijian Chen; Dongsu Du; Deshan Yang; H Harold Li
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

6.  X ray storage performance of KCl:Eu2+ with high cumulated dose.

Authors:  Rachael A Hansel; Zhiyan Xiao; Lei Zhang; H Harold Li
Journal:  Nucl Instrum Methods Phys Res B       Date:  2014-05-01       Impact factor: 1.377

7.  Temperature dependence of the photostimulated luminescence in KCl:Eu2+

Authors:  Zhiyan Xiao; Rachael Hansel; Lei Zhang; H Harold Li
Journal:  Nucl Instrum Methods Phys Res B       Date:  2014-05-01       Impact factor: 1.377

8.  Two-dimensional high spatial-resolution dosimeter using europium doped potassium chloride: a feasibility study.

Authors:  H Harold Li; Joseph P Driewer; Zhaohui Han; Daniel A Low; Deshan Yang; Zhiyan Xiao
Journal:  Phys Med Biol       Date:  2014-03-20       Impact factor: 3.609

9.  Performance of KCl:Eu2+ storage phosphor dosimeters for low-dose measurements.

Authors:  H Harold Li; Zhiyan Xiao; Rachael Hansel; Nels Knutson; Deshan Yang
Journal:  Phys Med Biol       Date:  2013-06-05       Impact factor: 3.609

10.  Dual-storage phosphor proton therapy dosimetry: Simultaneous quantification of dose and linear energy transfer.

Authors:  Jufri Setianegara; Thomas R Mazur; Deshan Yang; H Harold Li
Journal:  Med Phys       Date:  2021-02-19       Impact factor: 4.071

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.