Radiation hardness of the storage phosphor europium doped potassium chloride for radiation therapy dosimetry.

Joseph P Driewer1, Haijian Chen, Andres Osvet, Daniel A Low, H Harold Li.   

Abstract

PURPOSE: An important property of a reusable dosimeter is its radiation hardness, that is, its ability to retain its dosimetric merits after irradiation. The radiation hardness of europium doped potassium chloride (KC1:Eu2+), a storage phosphor material recently proposed for radiation therapy dosimetry, is examined in this study.
METHODS: Pellet-style KCl:Eu2+ dosimeters, 6 mm in diameter, and 1 mm thick, were fabricated in-house for this study. The pellets were exposed by a 6 MV photon beam or in a high dose rate 137Cs irradiator. Macroscopic properties, such as radiation sensitivity, dose response linearity, and signal stability, were studied with a laboratory photostimulated luminescence (PSL) readout system. Since phosphor performance is related to the state of the storage centers and the activator, Eu2+, in the host lattice, spectroscopic and temporal measurements were carried out in order to explore radiation-induced changes at the microscopic level.
RESULTS: KCl:Eu2+ dosimeters retained approximately 90% of their initial signal strength after a 5000 Gy dose history. Dose response was initially supralinear over the dose range of 100-700 cGy but became linear after 60 Gy. Linearity did not change significantly in the 0-5000 Gy dose history spanned in this study. Annealing high dose history chips resulted in a return of supralinearity and a recovery of sensitivity. There were no significant changes in the PSL stimulation spectra, PSL emission spectra, photoluminescence spectra, or luminescence lifetime, indicating that the PSL signal process remains intact after irradiation but at a reduced efficiency due to reparable radiation-induced perturbations in the crystal lattice.
CONCLUSIONS: Systematic studies of KCl:Eu2+ material are important for understanding how the material can be optimized for radiation therapy dosimetry purposes. The data presented here indicate that KCl:Eu2+ exhibits strong radiation hardness and lends support for further investigations of this novel material.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21928642      PMCID: PMC3161504          DOI: 10.1118/1.3611043

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  11 in total

1.  A new needle-crystalline computed radiography detector.

Authors:  P Leblans; L Struye; P Willems
Journal:  J Digit Imaging       Date:  2000-05       Impact factor: 4.056

Review 2.  The physics of computed radiography.

Authors:  J A Rowlands
Journal:  Phys Med Biol       Date:  2002-12-07       Impact factor: 3.609

3.  Theoretical and empirical investigations of KCl : Eu2+ for nearly water-equivalent radiotherapy dosimetry.

Authors:  Yuanshui Zheng; Zhaohui Han; Joseph P Driewer; Daniel A Low; H Harold Li
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

4.  Evaluation of a computed radiography system for megavoltage photon beam dosimetry.

Authors:  Arthur J Olch
Journal:  Med Phys       Date:  2005-09       Impact factor: 4.071

5.  Guidance document on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee.

Authors:  Gary A Ezzell; James M Galvin; Daniel Low; Jatinder R Palta; Isaac Rosen; Michael B Sharpe; Ping Xia; Ying Xiao; Lei Xing; Cedric X Yu
Journal:  Med Phys       Date:  2003-08       Impact factor: 4.071

6.  The energy dependence and dose response of a commercial optically stimulated luminescent detector for kilovoltage photon, megavoltage photon, and electron, proton, and carbon beams.

Authors:  Chester S Reft
Journal:  Med Phys       Date:  2009-05       Impact factor: 4.071

7.  Trapped charge migration in BaFBr:Eu2+: The recuperation of photostimulated luminescence.

Authors: 
Journal:  Phys Rev B Condens Matter       Date:  1996-03-01

8.  Changes in optically stimulated luminescent dosimeter (OSLD) dosimetric characteristics with accumulated dose.

Authors:  Paul A Jursinic
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

9.  Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements.

Authors:  Paul A Jursinic
Journal:  Med Phys       Date:  2007-12       Impact factor: 4.071

10.  Quantitative megavoltage radiation therapy dosimetry using the storage phosphor KCl: Eu2+.

Authors:  Zhaohui Han; Joseph P Driewer; Yuanshui Zheng; Daniel A Low; H Harold Li
Journal:  Med Phys       Date:  2009-08       Impact factor: 4.071

View more
  4 in total

1.  The role of activator concentration and precipitate formation on optical and dosimetric properties of KCl:Eu(2+) storage phosphor detectors.

Authors:  Rachael A Hansel; Zhiyan Xiao; Yanle Hu; Olga Green; Deshan Yang; H Harold Li
Journal:  Med Phys       Date:  2013-09       Impact factor: 4.071

2.  Temporal signal stability of KCl:Eu(2+) storage phosphor dosimeters.

Authors:  Zhiyan Xiao; Rachael Hansel; Haijian Chen; Dongsu Du; Deshan Yang; H Harold Li
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

3.  X ray storage performance of KCl:Eu2+ with high cumulated dose.

Authors:  Rachael A Hansel; Zhiyan Xiao; Lei Zhang; H Harold Li
Journal:  Nucl Instrum Methods Phys Res B       Date:  2014-05-01       Impact factor: 1.377

4.  Two-dimensional high spatial-resolution dosimeter using europium doped potassium chloride: a feasibility study.

Authors:  H Harold Li; Joseph P Driewer; Zhaohui Han; Daniel A Low; Deshan Yang; Zhiyan Xiao
Journal:  Phys Med Biol       Date:  2014-03-20       Impact factor: 3.609

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.