The role of activator concentration and precipitate formation on optical and dosimetric properties of KCl:Eu(2+) storage phosphor detectors.

Rachael A Hansel1, Zhiyan Xiao, Yanle Hu, Olga Green, Deshan Yang, H Harold Li.   

Abstract

PURPOSE: The activator ion (Eu(2+) in KCl:Eu(2+)) plays an important role in the photostimulated luminescence (PSL) mechanism of storage phosphor radiation detectors. In order to design an accurate, effective, and robust detector, it is important to understand how the activator ion concentration affects the structure and, consequently, radiation detection properties of KCl:Eu(2+).
METHODS: Potassium chloride pellets were fabricated with various amounts of europium dopant (0.01-5.0 mol.% Eu(2+)). Clinical radiation doses were given with a 6 MV linear accelerator. Radiation doses larger than 100 Gy were given with a (137)Cs irradiator. Dose response curves, radiation hardness, and temporal signal stability were measured using a laboratory PSL readout system. The crystal structure of the material was studied using x ray diffraction and luminescence spectroscopy.
RESULTS: The most intense PSL signal was from samples with 1.0 mol.% Eu. However, samples with concentrations higher than 0.05 mol.% Eu exhibited significant degradation in PSL intensity for cumulated doses larger than 3000 Gy. Structural and luminescence spectroscopy showed clear evidence of precipitate phases within the KCl lattice, especially for high activator concentrations. Analysis of PL emission spectra showed that interactions between Eu-Vc dipoles and Eu-Vc trimers could explain trends in PSL sensitivity and radiation hardness observations.
CONCLUSIONS: The concentration of the activator ion (Eu(2+)) significantly affects radiation detection properties of the storage phosphor KCl:Eu(2+). An activator concentration between 0.01 and 0.05 mol.% Eu in KCl:Eu(2+) storage phosphor detectors is recommended for linear dose response, good PSL sensitivity, predictable temporal stability, and high reusability for megavoltage radiation detection.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24007173      PMCID: PMC3765241          DOI: 10.1118/1.4817522

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  A new needle-crystalline computed radiography detector.

Authors:  P Leblans; L Struye; P Willems
Journal:  J Digit Imaging       Date:  2000-05       Impact factor: 4.056

Review 2.  The physics of computed radiography.

Authors:  J A Rowlands
Journal:  Phys Med Biol       Date:  2002-12-07       Impact factor: 3.609

3.  Theoretical and empirical investigations of KCl : Eu2+ for nearly water-equivalent radiotherapy dosimetry.

Authors:  Yuanshui Zheng; Zhaohui Han; Joseph P Driewer; Daniel A Low; H Harold Li
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

4.  Characterization of noise sources for two generations of computed radiography systems using powder and crystalline photostimulable phosphors.

Authors:  Alistair Mackenzie; Ian D Honey
Journal:  Med Phys       Date:  2007-08       Impact factor: 4.071

5.  Precipitation-induced quenching of Eu2+ luminescence in potassium halides.

Authors: 
Journal:  Phys Rev B Condens Matter       Date:  1990-12-15

6.  Participation of Eu-aggregates in the photoluminescence, afterglow and thermoluminescence of UV-irradiated KCl:Eu at 20 K.

Authors:  S Alvarez-García; T M Piters
Journal:  J Phys Condens Matter       Date:  2004-12-10       Impact factor: 2.333

7.  Imaging plate illuminates many fields.

Authors:  Y Amemiya; J Miyahara
Journal:  Nature       Date:  1988-11-03       Impact factor: 49.962

8.  Radiation hardness of the storage phosphor europium doped potassium chloride for radiation therapy dosimetry.

Authors:  Joseph P Driewer; Haijian Chen; Andres Osvet; Daniel A Low; H Harold Li
Journal:  Med Phys       Date:  2011-08       Impact factor: 4.071

9.  Changes in optically stimulated luminescent dosimeter (OSLD) dosimetric characteristics with accumulated dose.

Authors:  Paul A Jursinic
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

10.  Quantitative megavoltage radiation therapy dosimetry using the storage phosphor KCl: Eu2+.

Authors:  Zhaohui Han; Joseph P Driewer; Yuanshui Zheng; Daniel A Low; H Harold Li
Journal:  Med Phys       Date:  2009-08       Impact factor: 4.071

View more
  2 in total

1.  The role of oxygen in the photostimulation luminescence process of europium doped potassium chloride.

Authors:  Zhiyan Xiao; Thomas R Mazur; Joseph P Driewer; H Harold Li
Journal:  Radiat Meas       Date:  2015-05-01       Impact factor: 1.898

2.  Two-dimensional high spatial-resolution dosimeter using europium doped potassium chloride: a feasibility study.

Authors:  H Harold Li; Joseph P Driewer; Zhaohui Han; Daniel A Low; Deshan Yang; Zhiyan Xiao
Journal:  Phys Med Biol       Date:  2014-03-20       Impact factor: 3.609

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.