Repeatability and agreement of CCT measurement in myopia using entacam and ultrasound pachymetry.

Yong-Heng Luo1, Qian Zhong, Ping-Bo Ouyang, Xiao-Jian Guo, Xuan-Chu Duan.   

Abstract

AIM: To evaluate the repeatability of central corneal thickness (CCT) measurement by entacam, and agreement of CCT measured by Pentacam and ultrasound pachymetry (USP) in Chinese myopia. Thereby investigate the possibility of Pentacam as a substitute for USP in CCT measurement before refractive surgery. The effects of corneal curvature measured by Pentacam on CCT were also evaluated.
METHODS: One hundred and forty-eight right eyes of 148 individual with myopia were included in this study. Three successive Pentacam CCT measurements followed by 10 successive ultrasound pachymetry were carried out in the 148 eyes. Mean of CCT taken by each device was calculated for comparison. According to the CCT measured by USP, all the 148 eyes were divided into 3 groups: <520µm, 520-560µm, >560µm. For all eyes and each group the CCT obtained by Pentacam and USP were compared. Anterior corneal curvature of the 148 eyes was also adopted for correlation analysis with CCT obtained by ultrasound pachymetry. In addition, CCT measurement using 60 random selected Scheimpflug images was performed by 3 skilled investigators at different time, and this was repeated for 3 times by a forth investigator to assess repeatability of Pentacam CCT measurement using Scheimpflug images.
RESULTS: Intraclass correlation coefficient (ICC) analysis revealed high intraobserver repeatability (ICC=0.994, F=158.60, P<0.001) for CCT measurement by Pentacam. The interobserver (ICC=0.998, F=494.73, P<0.001) and intraobserver (ICC=0.997, F=383.98, P<0.001) repeatability for Pentacam CCT measurements using Scheimpflug images were also excellent. There was high positive correlation between the CCT values measured by Pentacam and ultrasound pachymetry (r=0.963, P<0.001). Bland-altman plots showed that the Pentacam underestimate the CCT by 8.02µm compared with ultrasouond pachymetry. The differences between Pentacam and USP increased as the CCT readings by USP increased (Pentacam vs USP: slope=-0.04, P<0.05). The 95% upper and lower limits of agreement between CCT values obtained from the two devices were +9.33µm and -25.37µm. No significant association could be found between CCT and anterior corneal curvature.
CONCLUSION: Inter- and intraobserver variability for CCT measurements by Pentacam was considerably below clinically significant levels. CCT of myopia obtained by Scheimpflug camera, Pentacam, were highly correlated to that by ultrasound pachymetry. However, the values obtained are not directly interchangeable between Pentacam and ultrasound pachymetry as the 95% limits of agreement are relatively wide. Pentacam can be a useful instrument for measuring CCT in candidates to refractive surgery in clinic.

Entities:  

Keywords:  agreement; central corneal thickness; pentacam; repeatability; ultrasound pachymetry

Year:  2012        PMID: 22773982      PMCID: PMC3388402          DOI: 10.3980/j.issn.2222-3959.2012.03.15

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  27 in total

1.  Central corneal thickness of normal tension glaucoma patients in Japan.

Authors:  L L Wu; Y Suzuki; R Ideta; M Araie
Journal:  Jpn J Ophthalmol       Date:  2000 Nov-Dec       Impact factor: 2.447

Review 2.  Mechanisms of disease: Fuchs' endothelial dystrophy.

Authors:  Sheila Borboli; Kathryn Colby
Journal:  Ophthalmol Clin North Am       Date:  2002-03

3.  Reliability coefficients of three corneal pachymeters.

Authors:  N C Wheeler; C M Morantes; R M Kristensen; T H Pettit; D A Lee
Journal:  Am J Ophthalmol       Date:  1992-06-15       Impact factor: 5.258

4.  Corneal thickness measurements in normal and keratoconic eyes: Pentacam comprehensive eye scanner versus noncontact specular microscopy and ultrasound pachymetry.

Authors:  Omür Ozlenen Uçakhan; Muhip Ozkan; Ayfer Kanpolat
Journal:  J Cataract Refract Surg       Date:  2006-06       Impact factor: 3.351

5.  Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography.

Authors:  Shiro Amano; Norihiko Honda; Yuki Amano; Satoru Yamagami; Takashi Miyai; Tomokazu Samejima; Miyuki Ogata; Kazunori Miyata
Journal:  Ophthalmology       Date:  2006-06       Impact factor: 12.079

6.  Central corneal thickness measurements using Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopia.

Authors:  Thomas Ho; Arthur C K Cheng; Srinivas K Rao; Silvania Lau; Chris K S Leung; Dennis S C Lam
Journal:  J Cataract Refract Surg       Date:  2007-07       Impact factor: 3.351

7.  Anterior segment biometry with the Pentacam: comprehensive assessment of repeatability of automated measurements.

Authors:  Hema Shankar; Deepa Taranath; Chandramalar T Santhirathelagan; Konrad Pesudovs
Journal:  J Cataract Refract Surg       Date:  2008-01       Impact factor: 3.351

Review 8.  Measuring the cornea: the latest developments in corneal topography.

Authors:  Tracy Swartz; Lisa Marten; Ming Wang
Journal:  Curr Opin Ophthalmol       Date:  2007-07       Impact factor: 3.761

Review 9.  Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach.

Authors:  M J Doughty; M L Zaman
Journal:  Surv Ophthalmol       Date:  2000 Mar-Apr       Impact factor: 6.048

10.  Reduction of intraocular pressure and glaucoma progression: results from the Early Manifest Glaucoma Trial.

Authors:  Anders Heijl; M Cristina Leske; Bo Bengtsson; Leslie Hyman; Boel Bengtsson; Mohamed Hussein
Journal:  Arch Ophthalmol       Date:  2002-10
View more
  1 in total

1.  Meta-analysis of Pentacam vs. ultrasound pachymetry in central corneal thickness measurement in normal, post-LASIK or PRK, and keratoconic or keratoconus-suspect eyes.

Authors:  Wenjing Wu; Yan Wang; Lulu Xu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-11-12       Impact factor: 3.117

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.