Comparison of central corneal thickness measurements by Pentacam, noncontact specular microscope, and ultrasound pachymetry in normal and post-LASIK eyes.

Saleh Al-Ageel1, Abdulrahman M Al-Muammar.   

Abstract

PURPOSE: To compare central corneal thickness (CCT) measurements taken with Pentacam, noncontact specular microscope (NCSM), and ultrasound pachymetry (US) in normal and post-laser in situ keratomileusis (LASIK) eyes and to assess the agreement between the three devices.
DESIGN: Prospective clinical trial. PATIENTS AND METHODS: Central corneal thickness (CCT) was measured in 94 eyes of normal volunteer and in 72 eyes of post-LASIK patients. Measurements were made with the three devices.
RESULTS: In normal eyes, the mean (±SD) CCT taken with Pentacam, NCSM, and US was 552.6 ± 36.8 μm, 511.9 ± 38.6 μm, and 533.3 ± 37.9 μm, respectively. The average values of CCT taken with the three instruments were significantly different. In post-LASIK eyes the mean CCT with Pentacam, NCSM, and US was 483.02 ± 6.03 μm, 450.7 ± 5.3 μm, and 469.5 ± 5.8 μm, respectively. The average values of CCT taken were significantly different for Pentacam vs. NCSM (P = 0.046) and Pentacam vs. US (P = 0.02), but not significant for NCSM vs. US (P = 0.352). The Bland and Altman method for assessing clinical agreement between 3 instruments showed that in normal eyes, the mean values and paired differences of the three CCT devices were found to be statistically independent. In post-LASIK eyes, there was significant association between the difference and the mean of the Pentacam and NCSM, and US and NCSM.
CONCLUSION: The three devices cannot be used interchangeably in normal and post-LASIK eyes. Pentacam tends to give significantly thicker reading than ultrasound pachymetry.

Entities:  

Keywords:  Central corneal thickness; Noncontact specular microscopy; Pentacam; Scheimpflug camera; Ultrasound pachymetry

Year:  2009        PMID: 23960858      PMCID: PMC3729515          DOI: 10.1016/j.sjopt.2009.10.002

Source DB:  PubMed          Journal:  Saudi J Ophthalmol        ISSN: 1319-4534


  36 in total

1.  The effects of long-term contact lens wear on corneal thickness, curvature, and surface regularity.

Authors:  Z Liu; S C Pflugfelder
Journal:  Ophthalmology       Date:  2000-01       Impact factor: 12.079

2.  Corneal thickness measurements with contact and noncontact specular microscopic and ultrasonic pachymetry.

Authors:  L Módis; A Langenbucher; B Seitz
Journal:  Am J Ophthalmol       Date:  2001-10       Impact factor: 5.258

3.  Central corneal thickness measurements with partial coherence interferometry, ultrasound, and the Orbscan system.

Authors:  Georg Rainer; Oliver Findl; Vanessa Petternel; Barbara Kiss; Wolfgang Drexler; Christian Skorpik; Michael Georgopoulos; Leopold Schmetterer
Journal:  Ophthalmology       Date:  2004-05       Impact factor: 12.079

4.  Reliability coefficients of three corneal pachymeters.

Authors:  N C Wheeler; C M Morantes; R M Kristensen; T H Pettit; D A Lee
Journal:  Am J Ophthalmol       Date:  1992-06-15       Impact factor: 5.258

5.  Corneal thickness measurements in normal and keratoconic eyes: Pentacam comprehensive eye scanner versus noncontact specular microscopy and ultrasound pachymetry.

Authors:  Omür Ozlenen Uçakhan; Muhip Ozkan; Ayfer Kanpolat
Journal:  J Cataract Refract Surg       Date:  2006-06       Impact factor: 3.351

6.  Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography.

Authors:  Shiro Amano; Norihiko Honda; Yuki Amano; Satoru Yamagami; Takashi Miyai; Tomokazu Samejima; Miyuki Ogata; Kazunori Miyata
Journal:  Ophthalmology       Date:  2006-06       Impact factor: 12.079

Review 7.  Measuring the cornea: the latest developments in corneal topography.

Authors:  Tracy Swartz; Lisa Marten; Ming Wang
Journal:  Curr Opin Ophthalmol       Date:  2007-07       Impact factor: 3.761

8.  Epithelial and corneal thickness measurements by in vivo confocal microscopy through focusing (CMTF).

Authors:  H F Li; W M Petroll; T Møller-Pedersen; J K Maurer; H D Cavanagh; J V Jester
Journal:  Curr Eye Res       Date:  1997-03       Impact factor: 2.424

9.  Corneal stress test for extended wear.

Authors:  O D Solomon
Journal:  CLAO J       Date:  1996-01

10.  Comparison of four corneal pachymetry technique;s in corneal refractive surgery.

Authors:  Jaime Javaloy; Maria Teresa Vidal; Jose R Villada; Alberto Artola; Jorge L Alió
Journal:  J Refract Surg       Date:  2004 Jan-Feb       Impact factor: 3.573

View more
  8 in total

1.  Measurement of central corneal thickness in health and disease.

Authors:  Imtiaz A Chaudhry
Journal:  Saudi J Ophthalmol       Date:  2009-10

2.  Measurements of central corneal thickness and endothelial parameters with three different non-contact specular microscopy devices.

Authors:  Ozgur Cakici; Remzi Karadag; Huseyin Bayramlar; Efe Koyun
Journal:  Int Ophthalmol       Date:  2016-05-24       Impact factor: 2.031

3.  Repeatability and interobserver reproducibility of Artemis-2 high-frequency ultrasound in determination of human corneal thickness.

Authors:  Kelechi C Ogbuehi; Uchechukwu L Osuagwu
Journal:  Clin Ophthalmol       Date:  2012-05-21

4.  Central corneal thickness and its determinants in a geriatric population: a population-based study.

Authors:  Hassan Hashemi; Payam Nabovati; Mohammadreza Aghamirsalim; Abbasali Yekta; Farhad Rezvan; Mehdi Khabazkhoob
Journal:  Eye (Lond)       Date:  2022-02-01       Impact factor: 4.456

5.  Corneal thickness measurements with Scheimpflug and slit scanning imaging techniques in keratoconus.

Authors:  Mohammad Aghazadeh Amiri; Hassan Hashemi; Shahroukh Ramin; Abbasali Yekta; Azadeh Taheri; Payam Nabovati; Mehdi Khabazkhoob
Journal:  J Curr Ophthalmol       Date:  2016-10-08

6.  Which feature influences on-eye power change of soft toric contact lenses: Design or corneal shape?

Authors:  Tamsin Doll; Joshua Moore; Ahmad H Shihab; Bernardo T Lopes; Ashkan Eliasy; Osama Maklad; Richard Wu; Lynn White; Steve Jones; Ahmed Elsheikh; Ahmed Abass
Journal:  PLoS One       Date:  2020-11-25       Impact factor: 3.240

7.  Performance of Zernike polynomials in reconstructing raw-elevation data captured by Pentacam HR, Medmont E300 and Eye Surface Profiler.

Authors:  Yueying Wei; Bernardo T Lopes; Ashkan Eliasy; Richard Wu; Arwa Fathy; Ahmed Elsheikh; Ahmed Abass
Journal:  Heliyon       Date:  2021-12-18

8.  Normal Corneal Thickness and Endothelial Cell Density in Rhesus Macaques (Macaca mulatta).

Authors:  M Isabel Casanova; Laura J Young; Sangwan Park; Soohyun Kim; Karolina Roszak; Brian C Leonard; Andrew Blandino; Monica J Motta; Glenn Yiu; Jennifer Y Li; Ala Moshiri; Sara M Thomasy
Journal:  Transl Vis Sci Technol       Date:  2022-09-01       Impact factor: 3.048

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.