Central and peripheral corneal thickness measurement with Orbscan II and topographical ultrasound pachymetry.

Jose Manuel González-Méijome1, Alejandro Cerviño, Eva Yebra-Pimentel, Manuel A Parafita.   

Abstract

PURPOSE: To compare thickness measurements of the central 6.0 mm of the cornea obtained with the Orbscan(R) II topography system and topographical ultrasound pachymetry.
SETTING: School of Optometry, University of Santiago de Compostela, Galicia, Spain.
METHODS: In 24 right eyes, pachymetric measurements were taken at the center and 1.2 mm and 3.0 mm on the superior and inferior hemimeridians. A 1-sample t test was applied to assess the significance of the relationship between Orbscan II and ultrasound methods. The relationship between the 2 was assessed by analyzing regression and plotting the differences against the mean corneal thickness. Orbscan II data were analyzed in 3 ways: (1) without the application of an acoustic equivalent correction factor; (2) with a correction factor of 0.92, as recommended by the manufacturer; (3) with correction using the equations derived in this study. The data were systematically compared with those of ultrasound pachymetry.
RESULTS: Before the correction factor was applied, the Orbscan II overestimated the corneal thickness at all locations, with the mean difference (48.15 microm +/- 33.74 [SD]) significantly different from zero (P <.001). Differences increased toward the periphery, and the reliability of Orbscan II readings seemed to decrease with thicker corneas. After the acoustic equivalent was applied, the differences were significantly less; however, this effect did not seem clinically significant as large differences remained. When specific corrective equations were applied for each corneal location, the level of agreement between Orbscan II and ultrasound pachymetry improved substantially; the mean (-0.11 +/- 15.22 microm) was not statistically different from zero (P >.05).
CONCLUSIONS: The acoustic equivalent correction factor proposed by the manufacturer to obtain corneal thickness measurements with the Orbscan II compared to those from ultrasound pachymetry was not valid for all corneal topography positions. Orbscan II measurements agreed better with those of ultrasound pachymetry when equations for the central and each peripheral location across the topography were applied.

Mesh:

Year:  2003        PMID: 12551679     DOI: 10.1016/s0886-3350(02)01815-1

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  24 in total

1.  Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis.

Authors:  K Kawana; T Tokunaga; K Miyata; F Okamoto; T Kiuchi; T Oshika
Journal:  Br J Ophthalmol       Date:  2004-04       Impact factor: 4.638

2.  Comparison of ultrasonic pachymetry with Orbscan in corneal haze.

Authors:  Rana Altan-Yaycioglu; Aysel Pelit; Yonca Aydin Akova
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2007-06-22       Impact factor: 3.117

3.  Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry.

Authors:  Hani S Al-Mezaine; Saleh A Al-Amro; Dustan Kangave; Abdulkareem Sadaawy; Taher A Wehaib; Saleh Al-Obeidan
Journal:  Int Ophthalmol       Date:  2007-09-26       Impact factor: 2.031

4.  Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography.

Authors:  Mohammad Naderan; Saeed Shoar; Morteza Naderan; Mohammad Amin Kamaleddin; Mohammad Taher Rajabi
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

5.  Multimodal Assessment of Corneal Thinning Using Optical Coherence Tomography, Scheimpflug Imaging, Pachymetry, and Slit-Lamp Examination.

Authors:  Julius T Oatts; Jeremy D Keenan; Tova Mannis; Tom M Lietman; Jennifer Rose-Nussbaumer
Journal:  Cornea       Date:  2017-04       Impact factor: 2.651

6.  Corneal pachymetry mapping with high-speed optical coherence tomography.

Authors:  Yan Li; Raj Shekhar; David Huang
Journal:  Ophthalmology       Date:  2006-05       Impact factor: 12.079

7.  Day to Day Clinically Relevant Corneal Elevation, Thickness, and Curvature Parameters Using the Orbscan II Scanning Slit Topographer and the Pentacam Scheimpflug Imaging Device.

Authors:  Hassan Hashemi; Shiva Mehravaran
Journal:  Middle East Afr J Ophthalmol       Date:  2010-01

8.  Comparison of artemis 2 ultrasound and Visante optical coherence tomography corneal thickness profiles.

Authors:  Roxana Ursea; Matthew Feng; Raksha Urs; Arindam RoyChoudhury; Ronald H Silverman
Journal:  J Refract Surg       Date:  2012-11-30       Impact factor: 3.573

9.  Decreased central corneal thickness in ankylosing spondylitis.

Authors:  Huseyin Ortak; Ahmet Inanır; Selim Demir; Alper Uysal; Şafak Şahin; Mustafa Sağcan; Yalçın Önder; Sait Alim; Ayşe Kevser Demir
Journal:  Int Ophthalmol       Date:  2013-07-12       Impact factor: 2.031

10.  Central corneal thickness measured by the Orbscan II system, contact ultrasound pachymetry, and the Artemis 2 system.

Authors:  Tania Paul; Mira Lim; Christopher E Starr; Harriet O Lloyd; D Jackson Coleman; Ronald H Silverman
Journal:  J Cataract Refract Surg       Date:  2008-11       Impact factor: 3.351

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.