Raphaël Wakam1,2, Aurélie Benoit1, Kwamivi Bernardin Mawussi3, Caroline Gorin4,5.
Abstract
Attachment systems (AS) enhance retention and stability by anchoring the overdentures to implants. Since 2002, the McGill consensus statement recommends the 2-implant-retained overdentures as the standard choice for edentulous mandible (2-IRMO). Considering the large number of AS available, it remains difficult for a practitioner to make a reasoned choice. A systematic review was conducted in PubMed/Medline and carried out independently by three authors, on retention, wear, and maintenance of AS used clinically or in vitro specifically for 1- or 2-IRMO. The 45 selected studies include 14 clinical and 31 in vitro studies. The risk of bias was evaluated according to the revised Cochrane risk of bias tool for randomized trials (RoB 2). The initial retention force of the cylindrical system is higher than the ball system. The retention loss, related to the wear of the retention device, is responsible for the most common need of maintenance, requiring activation or replacement. Plastic retention devices wear out faster and more significantly than metal ones, implying a worse time behavior of cylindrical systems, but their maintenance rate is similar. Neither system appears categorically superior. Cylindrical systems provide higher initial retention than ball ones; this advantage reduces over time with wear without affecting their need for maintenance.Entities:
Keywords: ball and cylindrical attachment systems; implant-retained overdenture; maintenance; retention force; wear
Year: 2022 PMID: 35269164 PMCID: PMC8911844 DOI: 10.3390/ma15051933
Source DB: PubMed Journal: Materials (Basel) ISSN: 1996-1944 Impact factor: 3.623
Figure 1Design of some ball attachment systems (Dalbo®-Plus, Preci-Clix®, O-Ring®) and cylindrical attachment systems (Locator®, Locator R-Tx®, Novaloc®).
Figure 2Search method for identification of studies using PRISMA guidelines for systematic reviews.
Figure 3Risk of bias graph: author’s assessment of each risk of bias item in proportions for all clinical (A) and in vitro studies (B).
Figure 4Risk of bias summary: authors’ assessment of each risk of bias item for each included clinical (A) and in vitro study (B).
Initial retention force (Newton) of ball and cylindrical attachment systems and the influence of insertion–removal (IRC) and/or chewing cycles (CC) on the loss of retention characterized in in vitro studies. Abbreviations: ns, non-significant; Ti, titanium; TiN, titanium nitride. Negative retention loss corresponds to a gain in retention.
| Studies | 1 or 2-IRMO | Inter-Implant Distance (mm) | Initial and Final Number and Type of Cycles | Cross-Head Speed | Attachments System, Manufacturer | Materials | Retention Setup Before the Test (N) | Retention Force (N) | Retention Loss (%) | ||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Male Part | Retentive Device | Initial | Final | ||||||||
| Branchi et al., 2010 | 1 | - | 1 | 54 | Ball Ø 2.20 mm Sweden & Martina | Ti | Teflon matrix | - | 13.92 | 17.75 | 27 |
| Red O-ring® rubber | - | 10.20 | 2.45 | 75 | |||||||
| Gold | - | 15.30 | 24.42 | −50 | |||||||
| Titanium | - | 13.24 | 4.31 | 68 | |||||||
| Yabul et al., 2018 | 2 | 19 | 10 | 54 | Ball, Biohorizons | Ti | Plastic | - | 34.12 ± 4.99 | 11.19 ± 2.8 | 67.3 |
| Ball, DTI, Istantbul | Plastic | - | 26.41 ± 5.8 | 10.58 ± 2.96 | 58.9 ± 12.21 | ||||||
| Ball, Straumann | Gold alloy | - | 48.16 ± 6.46 | 43.0 | 10.5 | ||||||
| Ball, Straumann | Titanium | - | 50.39 ± 4.81 | 5.59 | 88.7 | ||||||
| Elsyad et al., 2016 | 2 | 22 | 1 | 50 | Locator®, Zest Anchors | Ti + TiN | Blue | - | 19.64 ± 1.16 | 14.80 ± 0.83 | 24.64 |
| Pink | 51.20 ± 0.83 | 33.60 ± 2.30 | 34.37 | ||||||||
| White | 65.20 ± 1.30 | 39.80 ± 1.48 | 38.95 | ||||||||
| Gamborena et al., 1997 | 1 | - | 1 | 50 | ERA®, Sterngold | Gold alloy | White | 14.91 ± 3.43 | 2.25 | 85 | |
| Orange | 24.71 ± 3.14 | 2.74 | 88 | ||||||||
| Blue | 22.75 ± 2.84 | 3.73 | 85 | ||||||||
| Gray | 24.03 ± 3.53 | 3.43 | 87 | ||||||||
| Stephens et al., 2014 | 2 | 22 | 1 | 60 | Locator®, Zest Anchors | Ti + TiN | Blue | 21.81 ± 7.44 | 15.97 ± 3.96 | 5.84 | |
| Choi et al., 2018 | 2 | 22 | 1 | 50 | Locator®, Zest Anchors | Ti + TiN | Blue | - | 20.1 ± 2.87 | 20.58 ± 3.09 | −2.81 |
| Pink | 24.55 ± 2.14 | 37.42 ± 2.79 | −52.78 | ||||||||
| White | 69.87 ± 5.73 | 42.56 ± 3.27 | 38.81 | ||||||||
| CM Loc®, Cendres & Métaux | Ti | Light Green | 12.45 ± 1.27 | 12.83 ± 1.39 | −3.01 | ||||||
| Red | 17.79 ± 1.70 | 21.48 ± 1.20 | −22.31 | ||||||||
| Green | 39.35 ± 3.45 | 36.99 ± 1.75 | 5.57 | ||||||||
| Blue | 44.07 ± 3.07 | 43.80 ± 2.45 | 0.52 | ||||||||
| Maniewicz et al., 2020 | 2 | 20 | 1 | - | CM Loc®, Cendres & Métaux | Ti | Red | 81.8 | 49.2 | 39.85 | |
| Locator R-Tx®, Zest Anchors | Ti + TiN | Pink | 75.5 | 60.0 | 20.53 | ||||||
| Novaloc®, Straumann AG | Ti + ADLC | Yellow | 57.7 | 59.4 | −2.95 | ||||||
| Yilmaz et al., 2020 | 2 | - | 1 | 50 | Locator®, Zest Anchors | Ti + TiN | Pink | 13.6 | 10.0 | 26.5 | |
| Locator R-Tx®, Zest Anchors | Ti + TiN | Pink | 20.1 | 14.0 | 30.4 | ||||||
| Tehini et al., 2020 | 2 | 22 | 1 | 60 | Locator®, Zest Anchors | Ti + TiN | Blue | 9.95 | 6.37 | −37 | |
| Pink | 15.43 ± 4.08 | 14.00 ± 3.89 | 9 | ||||||||
| White | 41.73 ± 9.29 | 38.20 ± 5.11 | 7 | ||||||||
| Chung et al., 2004 | 2 | - | 1 | 50 | ERA®, Sterngold | Gold alloy | White | 23.76 ± 1.02 | |||
| Grey | 35.24 ± 1.99 | ||||||||||
| Locator®, Zest Anchors | Ti + TiN | Pink | 12.33 ± 1.28 | ||||||||
| White | 28.95 ± 0.78 | ||||||||||
| Jabbour et al., 2014 | 2 | - | 6 months | Clinical conditions | Retentive anchor, Straumann | Ti | Gold matrix | 34.58 * | 24.58 * | 28.92 * | |
| 12 months | 20.27 * | 41.38 * | |||||||||
| 6 months | Locator®, Zest Anchors | Ti + TiN | White | - | 39.27 * | 15.47 * | 60.61 * | ||||
| 12 months | 12.00 * | 66.54 * | |||||||||
| Gonuldas et al., 2018 | 2 | 22 | 1 | 50 | Ball, T.A.G. Medical Products | Ti | Orange plastic cap | 30.7 | 21.64 ± 1.23 | 29.51 | |
| White plastic cap | 47.9 ± 2.62 | 26.24 ± 0.71 | 45.22 | ||||||||
| Locator®, Zest Anchors | Ti + TiN | Blue | 61.39 ± 4.26 | 5.87 | 90.44 | ||||||
| Pink | 79.14 ± 3.63 | 10.4 | 86.86 | ||||||||
| White | 93.75 ± 11.23 | 18.28 ± | 80.50 | ||||||||
| Sultana et al., 2017 | 2 | 22 | 10 | 45 | Ball, Dentsply | Ti | Red plastic cap Clix® | - | 56.2 | 46.0 | 18.1 |
| Locator®, Zest Anchors | Ti + TiN | Pink | 108.9 ± 29.78 | 20.2 | 81.5 | ||||||
| Green | 82.3 ± 14.15 | 17.3 | 79.0 | ||||||||
| Türk et al., 2014 | 2 | 22 | 10 | 50 | Ball, O-Ring®, Biohorizons | Ti | Plastic cap Clix® | - | 32.91 ± 5.30 | 9.70 | 69.43 |
| Locator®, Zest Anchors | Ti + TiN | Pink | - | 52.47 ± | 21.70 ± 10.13 | 57.56 | |||||
| Wolf et al., 2009 | 1 | - | 10 | - | Ball, Dalbo Plus®, Cendres & Métaux | Gold alloy | Gold alloy strip | 7 | 8.86 | 2.31 | 77.54 |
| Ti | Gold alloy strip | 7 | 9.87 ± 1.6 | 11.61 ± 3.7 | −5.58 | ||||||
| Ball Ecco®, Unor | Gold alloy | Green plastic cap | 8 | 6.97 | 1.13 | 88.55 | |||||
| Ball Tima®, Unor | Gold alloy | Stainless steel ring | 8 | 11.92 ± 3.1 | 1.55 * ± 1.4 | 89.82 | |||||
| Ball, Pro-Snap®, Metalor Dental | Gold alloy | Green plastic cap | 10 | 8.52 | 3.40 | 54.23 | |||||
| Locator®, Zest Anchors | Ti + TiN | Pink | - | 13.250 ± 6.6 | 2.462 ± 1.8 | 85.66 | |||||
| Alsabeeha et al., 2010 | 1 | - | 10 | 50 | Ball Ø 2.25 mm, Southern | Ti | Gold alloy strip | - | 17.32 ± 3.68 | - | |
| Ball Ø 5.9 mm, Southern | Pure Ti + TiN | Plastic cap | - | 32.06 ± 2.59 | |||||||
| Locator®, Zest Anchors | Ti + TiN | Blue | - | 3.83 | |||||||
| Pink | 9.40 | ||||||||||
| White | 12.39 ± 0.55 | ||||||||||
| Yang et al., 2011 | 1 | - | 1 | 60 | Dal-ro®, Biomet 3i | Ti | Gold alloy strip | - | 6.48 | ||
| Locator, Zest Anchors | Ti + TiN | Blue | - | 15.36 ± 1.4 | |||||||
| Abi Nader et al., 2011 | 2 | 15 | 1 | 15 | Ball Ø 2.25 mm, Nobel Biocare | Ti | Gold alloy | 1 turn | 10.6 | 7.9 | 25.47 |
| Locator®, Zest Anchors | Ti + TiN | White | - | 66.4 | 21.6 | 67.47 | |||||
| Petropoulos et al., 2011 | 2 | - | 1 | 50.8 | Ball Ø 3.5 mm, Nobel Biocare | Ti | Rubber O-ring® | - | 24.3 | ||
| Ball Ø 2.25 mm, Nobel Biocare | Ti | Ti cap/Ti spring | - | 17.8 | |||||||
| Zest®, Zest Anchors | - | - | - | 10.8 | |||||||
| ZAAG®, Zest Anchors | - | - | - | 37.2 | |||||||
| ERA®, Sterngold | - | White | - | 12.7 | |||||||
| Orange | 18.5 | ||||||||||
| Kobayashi et al., 2014 | 2 | 20 | 10 | 4000 | Retentive Anchor, Straumann | Ti | Gold alloy | 7 (0.5 turn) | 40.3 | 67.9 * ± 15.83 | −68.48 (ns) |
| Locator®, Zest Anchors | Ti + TiN | Blue | - | 33.5 | 24.57 ± 12.35 | 26.66 | |||||
| Saheli et al., 2019 | 2 | 19 | 1 | 50 | Ball, Dio Implant, Dio Corp | Ti | - | - | 15.6 | 9.3 | 40.38 |
| 23 | 15.3 | 8.0 | 46.20 | ||||||||
| 29 | 13.8 | 10.2 | 26.07 | ||||||||
| 19 | Locator®, Zest Anchors | Ti + TiN | - | - | 17.1 | 5.2 | 69.59 | ||||
| 23 | 33.5 | 8.0 | 76.12 | ||||||||
| 29 | 33.1 | 6.8 | 79.46 | ||||||||
| Scherer et al., 2014 | 2 | Inter-canine | 1 | 50.8 | Ball, Zimmer Dental | Ti | White cap | - | 35.23 | - | |
| Ball O-ring® Saturno Standard, Zest Anchors | Ti | Rubber ring | 13.13 | ||||||||
| ERA®, Sterngold | - | Orange | 9.36 | ||||||||
| Locator®, Zest Anchors | Ti + TiN | Pink | 26.64 | ||||||||
| Rutkunas et al., 2011 | 1 | 1 | 50 | OP Anchor®, Inoue A. Co. Ltd. | Gold alloy | Rubber ring | - | 3.15 | 3.70 | −17.4 | |
| ERA®, Sterngold | Gold alloy | White | - | 13.12 ± 3.3 | 2.89 | 87 | |||||
| Orange | 12.63 ± 1.4 | 2.86 | 88 | ||||||||
| Locator® Root, Zest Anchors | Stainless steel + TiN | Blue | 16.50 ± 9.4 | 6.24 | 62.18 | ||||||
| Pink | 15.20 ± 6.9 | 11.95 ± 3.5 | 21.38 | ||||||||
| White | 16.61 ± 2.2 | 10.28 ± 3.9 | 38.11 | ||||||||
Prosthodontic maintenance of 1- or 2-implant-retained mandibular overdentures and their ball or cylindrical attachment systems. Abbreviations: nc, non-communicated; H, hauteur; RD, retention device; RCT, randomized clinical trial, Ø, diameter.
| Clinical Studies | Attachment Systems (Male Part/Female Part/Manufacturer) | 1 or 2-IRMO | Number of Patients | Maintenance of the Attachment System | Maintenance of the Overdenture | Total Number of Events | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Activated RD | Replaced RD | Replaced Matrix Housing | Tightened or | Relined Denture | Fracture or Occlusal Adjustment | Replaced Denture | |||||
| Cune et al., 2004 | Titanium ball/Gold alloy RD/Friadent, Mannheim, Germany | 2 | 18 | 7 | - | 1 | 1+4 | nc | nc | nc | 13 |
| Mackie et al., 2011 | Titanium ball Ø 2.25-mm/Gold RD/Straumann, Basel, Switzerland | 2 | 17 | 9.6 | 1.1 | nc | nc | nc | nc | nc | 3.9 ± 2.1 |
| Titanium ball Ø 2.25-mm/Gold RD/Brånemark, Nobel Biocare | 10 | 18.0 | 5.5 | nc | nc | nc | nc | nc | 28.8 ± 12.6 | ||
| Titanium ball Ø 2.25-mm/Gold-platinum RD/Southern Implants | 11 | 11.9 | 2.2 | nc | nc | nc | nc | nc | 16.4 ± 7.5 | ||
| Titanium ball Ø 2.25-mm/Titanium RD/Straumann, Basel, Switzerland | 9 | - | 13.7 | nc | nc | nc | nc | nc | 24.9 ± 10.7 | ||
| Titanium ball Ø 3.95-mm/Plastic RD/Southern Implants | 22 | - | 4.3 | nc | nc | nc | nc | nc | 8.7 ± 4.2 | ||
| Titanium ball Ø 2.2-mm/Rubber ring RD/Steri-Oss (Locator®/Zest Anchors after 6 years) | 21 | - | 29.2 | nc | nc | nc | nc | nc | 32.2 ± 14.5 | ||
| Nogueira et al., 2018 | Nitrite-coated titanium ball/Nylon RD/Neodent, Curitiba, Paraná, Brazil | 1 | 45 | - | 66 | 12+7 | 7+6 | 3 | 23 | nc | 124 |
| Passia et al., 2019 | Titanium ball, Camlog Biotechnologies, Switzerland/Gold RD, Dalbo- Plus Elliptic, Cendres & Métaux, Biel, Switzerland | 1 | 11 | 29 | 23 | nc | 5+4 | 14 | 8 | nc | 83 |
| Walton et al., 2009 | Titanium retentive anchor/Gold RD/Straumann, Basel, Switzerland | 1 | 42 | 37 | 4 | 0 | 5 | 60 | 4+5+2 | nc | 159 |
| 2 | 44 | 34 | 4 | 2 | 1 | 44 | 2+2 | nc | 81 | ||
| Alsabeeha et al., 2011b | Uncoated standard titanium ball Ø 2.25-mm/Dalla Bona–type Gold alloy RD/Southern Implants | 1 | 12 | 13 | 0 | nc | 0+2 | nc | nc | nc | 15 |
| Titanium nitride–coated ball Ø 5.9-mm/Plastic cap RD/Southern Implants | 12 | - | 2 | nc | 0 | nc | nc | nc | 2 | ||
| Titanium nitride–coated abutment/Blue nylon cap RD/Locator®, Zest Anchors, Escondido, CA, USA | 12 | - | 16 | nc | 0 | nc | nc | nc | 16 | ||
| Cristache et al., 2014 | Titanium retentive anchor, H 3.4 mm/Gold RD Elitor®/Straumann, Basel, Switzerland | 2 | 12 | 144 | 8 | nc | 1 | 4 | 2+0 | 2 | 161 |
| Titanium retentive anchor H 3.4 mm/Titanium RD and stainless steel spring, 6.86–10.79 N/Straumann, Basel, Switzerland | 11 | - | 0 | 0 | 2 | 2 | 1+0 | 1 | 6 | ||
| Titanium nitride–coated abutment H 3 mm/Pink nylon cap/Locator®, Zest Anchors, Inc., Escondido, CA, USA | 23 | - | 22 | nc | 0 | 1 | 0 | 1 | 24 | ||
| Kleis et al., 2010 | Dal-Ro®, Biomet 3i Implant Innovations, Palm Beach Gardens, FL, USA | 2 | 25 | 4 | 0 | 0 | 1 | nc | 1 | nc | 6 |
| TG-O-Ring® ball/Rubber ring RD/ | 8 | - | 10 | 3 | 1 | nc | 0 | nc | 14 | ||
| Locator®, Zest Anchors, Escondido, CA, USA | 23 | - | 24 | 4+4 | 2 | nc | 1 | nc | 35 | ||
| Krenmair et al., 2012 | Ball abutment/Gold RD/Camlog, Screw-line, Altatec) | 2 | 10 | 2 | 0 | nc | 1+1 | 5 | 0+1 | 0 | 10 |
| Locator® abutment/Pink RD/Zest Anchors, Inc., Escondido, CA, USA | 9 | - | 4 | nc | 1 | 4+1 | 1 | 0 | 11 | ||
Wear patterns and dimensional changes of ball and cylindrical attachment systems in clinical and in vitro studies. Abbreviations: RCT, randomized clinical trial; SEM, scanning electron microscope; Ø, diameter.
| Studies | Attachment System, Manufacturer | Number of AS | Assessed Parameters/Methods | Results and Location | |
|---|---|---|---|---|---|
| Patrix | Matrix | ||||
| Alsabeeha et al., 2011b | Titanium nitride–coated ball Ø 5.9-mm/Plastic cap RD/Southern Implants | 5/12 | Wear patterns/SEM | Unaffected | Slight signs of wear |
| Uncoated standard titanium ball Ø 2.25-mm/Dalla Bona–type Gold alloy RD/Southern Implants | 5/12 | Extensive material loss and abrasion along the path of insertion–removal and across the circumference | Extensive plastic deformation, material flaking and sloughing | ||
| Titanium nitride–coated abutment/Blue nylon cap RD/Locator®, Zest Anchors | 5/12 | Unaffected | Surface rupture and material loss (central core) | ||
| Fromentin et al., 2011a | Titanium ball anchor/Gold alloy RD Elitor®/Straumann | 144 (male and female parts) | Wear patterns/SEM | Year 1: Slight scratches only at the equator | Year 1: Roughening, and material loss in the form of flakes |
| Fromentin et al., 2011b | Titanium ball anchor (Ø 2262 to 2267 µm)/Gold alloy RD/Straumann, Basel, Switzerland | 69 patrices | Measure of ball Ø, calculation of Ø loss | Year 1 (24 AS): Ø 5 to 7 µm (20–23%). Significant more loss in BL but no significant difference in Ø reduction | |
| Fromentin et al., 2012 | Titanium ball anchor, Straumann AG, Basel, Switzerland/Gold alloy RD (Ø 2973/2214/2300 µm, E 380/336 µm), Cendres+Metaux, Biel, Switzerland | 70 matrices + | Measure of the external, internal upper and internal lower matrix Ø and deviations from circularity. Calculation of the upper and lower thickness (E) and the thickness loss of the lamellae tip/Coordinate measuring machine with a touch trigger probe | Year 1 (26 AS): Ø 2989/2232/2298 µm; E 373/342 µm, loss 7 µm. Significant increase of the internal upper Ø, no difference of deviation of circularity in the 3 different areas | |
| Jabbour et al., 2014 | Retentive Anchor/Gold RD/Straumann, Burlington, ON, Canada | 48 | Wear patterns/SEM | No significant scratches, minor flattening (equatorial zone) | nc |
| Locator® abutment/White nylon RD/Zest Anchors, Escondido, CA, USA | 48 | Wear patterns/SEM + high-resolution µCT | No significant scratches and spots | Significant wear and plastic deformation (peripheral notch edge, central core edge) | |
| Abi Nader et al., 2011 | Titanium ball/Gold lamellae RD/Nobel Biocare | 4/16 | Wear patterns/SEM | Discrete wear | Slight wear and discrete deformation, probably due to the activator tool (internal surfaces, corners of retentive lamellae) |
| Titanium nitride-coated abutment/nylon RD/Locator®, Zest Anchors, Escondido, CA, USA | 4/16 | Deformation possibly caused by the specific tool (inner surface of the retentive area) | Severe wear | ||
| Choi et al., 2018 | Pure titanium abutment/Green, Red, Light Green and Blue PEEK RD/CM Loc®, Cendres & Métaux | 1/20 | Wear patterns/SEM | No noticeable abrasion | Slight wear, probably from tools used for placement (top and along the vertical split) |
| Titanium nitride–coated abutment/Blue, Pink and White nylon cap RD/Locator®, Zest Anchors | 1/20 | No noticeable abrasion | Severe wear, plastic deformation, surface irregularities, loss of materials more than PEKK RD (retention area, top of central core and periphery) | ||
| Gamborena et al., 1997 | White, Orange, Blue and Grey RD/ERA®, APM-Sterngold | 3/5 | Wear patterns and dimensional Ø changes (top and middle part of central core, metal inner ring of matrix housing)/Traveling | No difference between new and worn matrix housing. Ø loss of central core: −1.80 to +3.54%. The difference (statistical) in Ø between the matrix housing and the middle portion of the RD ranged from 72 to 126 µm | |
| Rabbani et al., 2015 | Blue, Pink and White RD/Locator®, Zest Anchors, Inc., Escondido, CA | 30 | Wear patterns/SEM | nc | Significant wear (retention area, central core, periphery) |
| Rutkunas et al., 2011 | Gold alloy ball/rubber ring/OP Anchor®, Inoue Attachments Co., Tokyo, Japan | 2/5 | Wear patterns and dimensional changes/SEM | No Ø loss: 0.52% | |
| Gold alloy abutment/nylon RD (Orange, White)/ERA®, Sterngold, Attleboro, USA | 2/5 | Ø loss: 0 to 2.48%, smooth surfaces (central core and inner peripherical surface) | |||
| Stainless steel TiN coated abutment/nylon RD (Blue, Pink, White), Locator® Root, Zest Anchors, Escondido, USA | 2/5 | Ø loss: 0.22 to 5.34%, surface particle loss and irregular surface (central core and inner peripherical) | |||
| Stephens et al., 2014 | Blue RD/Locator®, Zest Anchors, Inc., Escondido, CA | 20 | Wear patterns/SEM | No visible wear | Severe wear (central core > matrix housing) |
| Türk et al., 2014 | Ball abutment/Rubber ring RD/O-Ring®, Biohorizons | 10 | Dimensional changes (outer and inner diameters of the abutments)/SEM | nc | Ø loss: outer (0.14 ± 0.07mm) and inner (0.08 ± 0.08 mm) |
| Locator®, Zest Anchors, Inc., Escondido, CA | 10 | nc | Ø loss: outer (0.17 ± 0.11 mm), inner: 0.11 ± 0.14 mm. No significant difference between the two AS | ||
| Wolf et al., 2009 | Precious alloy ball/Titanium matrix housing with precious alloy RD/Dalbo-Plus®, Cendres & Metaux | 1 | Wear patterns/SEM | Noticeable signs of abrasion | nc |
| Titanium ball/Titanium matrix housing with precious alloy RD/Straumann | 1 | Slight grooves at the equator without any measurable changes in Ø | Minor wear at the tips of the metal lamellae | ||
| Precious alloy ball/Titanium matrix housing with Green plastic RD/Ecco®, Unor | 1 | Large grooves at the equator without any Ø loss | Obvious damages | ||
| Precious alloy ball/Titanium matrix housing with stainless steel spring/Tima®, Unor | 1 | Extensive signs of wear at the equators with Ø loss | Fractures of the retention springs | ||
| Precious alloy ball/Titanium matrix housing with Red plastic RD/Pro-Snap®, Metalor | 1 | Little signs of abrasion at the equator | nc | ||
| Titanium-nickel coating abutment/Stainless steel housing with Pink RD/Locator®, Zest Anchor | 1 | Little abrasion at the equator | Considerable signs of wear | ||
| Yabul et al., 2018 | Ball/Gold RD/Straumann AG, Basel, Switzerland | 24 | Wear patterns and volumetric loss of ball/Three-dimensional laser scanner | 0.7 ± 0.47% | |
| Ball/Titanium RD/Straumann AG, Basel, Switzerland | 24 | 25.38 ± 5.41% | |||
| Ball/Plastic RD/Biohorizons, Birmingham, Alabama | 24 | 12.94 ± 1% | |||
| Ball/Plastic RD/DTI, Istanbul, Turkey | 24 | 10.47 ± 1.7% | |||
Figure 5Overdenture with two parallel attachments on (A) two parallel implants, and (B) two non-parallel implants. Green axis: axis of insertion–removal of the denture. Black axis: axis of the implant. α: implant angulation. β: inter-implant angulation.
Influence of implant angulation on the initial retention force and after insertion–removal (IRC) and/or chewing cycles (CC) of selected ball and cylindrical attachment systems. Negative retention loss corresponds to a gain in retention.
| In Vitro Studies | 1 or 2 IRMO | Inter-implant Distance (mm) | Angulation | Initial and Final Number and Type of Cycles | Cross-Head Speed (mm/min) | Medium | Attachment Systems/Retentive Device | Implant Angulation α (°) | Retention Force (N) | Retention Loss (%) | |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Initial | Final | ||||||||||
| Ortégon et al., 2009 | 2 | 20 | Distal | 1 | 50 | nc | Ball Astra-Tech/PreciClix® RD | 0 (0 RD) | nc | 20.11 ± 2.51 | nc |
| 10 (0 RD) | nc | 21.31 ± 1.79 | nc | ||||||||
| 15 (0 RD) | nc | 18.73 ± 4.14 | nc | ||||||||
| 10 (10 RD) | nc | 19.93 ± 1.38 | nc | ||||||||
| 15 (15 RD) | nc | 16.84 ± 1.77 | nc | ||||||||
| Elsyad et al., 2018 | 2 | 22 | Mesial | 1 | 50 | nc | Locator®/Blue | 0 | 20.63 ± 0.70 | 3.52 ± 0.46 | 82.94 |
| 5 | 15.90 ± 0.65 | 24.02 ± 0.97 | 51.07 | ||||||||
| 10 | 28.07 ± 1.01 | 26.28 ± 0.62 | 63.77 | ||||||||
| 20 | 47.44 ± 0.51 | 30.56 ± 0.51 | 35.58 | ||||||||
| Locator®/Pink | 0 | 40.07 ± 0.90 | 19.15 ± 0.78 | 52.21 | |||||||
| 5 | 30.13 ± 0.82 | 31.23 ± 0.68 | −3.65 | ||||||||
| 10 | 31.57 ± 0.51 | 16.16 ± 1.04 | 48.81 | ||||||||
| 20 | 56.18 ± 0.75 | 36.45 ± 1.50 | 35.12 | ||||||||
| Locator®/White | 0 | 49.20 ± 0.72 | 32.01 ± 0.01 | 34.94 | |||||||
| 5 | 40.26 ± 1.10 | 41.14 ± 1.03 | −2.18 | ||||||||
| 10 | 41.06 ± 1.00 | 19.05 ± 0.93 | 53.60 | ||||||||
| 20 | 57.28 ± 0.63 | 19.99 ± 1.00 | 65.10 | ||||||||
| Elsyad et al., 2019 | 2 | 22 | Distal | 1 | 50 | nc | Locator®/Blue | 0 | 20.63 ± 0.70 | 3.52 ± 0.47 | 82.94 |
| 5 | 15.29 ± 0.61 | 9.13 ± 0.81 | 40.29 | ||||||||
| 10 | 30.16 ± 1.25 | 2.97 ± 1.05 | 90.15 | ||||||||
| 20 | 17.08 ± 0.88 | 30.26 ± 0.65 | −77.17 | ||||||||
| Locator®/Pink | 0 | 40.07 ± 0.90 | 19.15 ± 0.78 | 51.40 | |||||||
| 5 | 34.60 ± 0.53 | 16.31 ± 1.14 | 52.86 | ||||||||
| 10 | 47.10 ± 0.85 | 9.11 ± 1.02 | 80.66 | ||||||||
| 20 | 39.12 ± 1.02 | 21.10 ± 1.01 | 46.06 | ||||||||
| Locator®/White | 0 | 48.20 ± 0.72 | 32.02 ± 1.00 | 33.57 | |||||||
| 5 | 43.13 ± 1.20 | 44.01 ± 1.00 | −2.04 | ||||||||
| 10 | 49.25 ± 1.39 | 40.18 ± 1.05 | 18.42 | ||||||||
| 20 | 40.30 ± 1.13 | 41.32 ± 1.50 | −2.53 | ||||||||
| Locator®/Green | 20 | 27.50 ± 0.50 | 15.42 ± 0.52 | 43.93 | |||||||
| Locator®/Red | 20 | 38.23 ± 1.08 | 20.14 ± 1.03 | 47.32 | |||||||
| Rabbani et al., 2015 | 2 | 23 | Mesial | 1 | nc | nc | Locator®/Blue | 0/0 | 77 ± 13.5 | 25.8 ± 5.2 | 65.5 ± 10.2 |
| 0/10 | 66.4 ± 26.7 | 14.7 ± 7.9 | 70.6 ± 22.9 | ||||||||
| 5/5 | 73.7 ± 10.1 | 18.4 ± 3.7 | 74.1 ± 7.9 | ||||||||
| Locator®/Pink | 0/0 | 72.7 ± 1.3 | 27.7 ± 8.2 | 62.1 ± 10.7 | |||||||
| 0/10 | 74.8 ± 6.7 | 31.3 ± 4.1 | 35.1 ± 2.4 | ||||||||
| 5/5 | 71.4 ± 3.4 | 29.4 ± 2.7 | 30.2 ± 2.7 | ||||||||
| Locator®/White | 0/0 | 83.8 ± 10.9 | 32.0 ± 10.7 | 62.1 ± 9.6 | |||||||
| 0/10 | 101.32 ± 12.0 | 35.1 ± 2.4 | 65.1 ± 4.2 | ||||||||
| 5/5 | 89.5 ± 15.7 | 30.2 ± 2.7 | 65.1 ± 9.4 | ||||||||
| Stephens et al., 2014 | 2 | 22 | Distal | 1 | 60 | Artificial saliva | Locator®/Blue | 0 | 21.81 ± 7.44 | 15.97 ± 3.96 | 26.78 |
| 5 | 30.03 ± 6.24 | 15.43 ± 1.59 | 48.62 | ||||||||
| 10 | 24.75 ± 6.83 | 14.22 ± 2.43 | 42.54 | ||||||||
| Al-Ghafi et al., 2009 | 2 | 15 | nc | 1 | nc | nc | Locator®/Green | 0 | 81.75 | nc | nc |
| 5 | 91.74 | nc | nc | ||||||||
| 10 | 104.72 | nc | nc | ||||||||
| 15 | 84.86 | nc | nc | ||||||||
| 20 | 78.04 | nc | nc | ||||||||
| Passia et al., 2016 | 1 | - | nc | 10 | nc | Water | Locator®/Green | 0 | 21.5 | nc | 66 |
| 20 | 24.4 | nc | 75 | ||||||||
| CM Loc®/Green | 0 | 22.5 | nc | 32 | |||||||
| 20 | 27.4 | nc | 47 | ||||||||
| Choi et Jeong, 2018 | 2 | 22 | Distal | 1 | 50 | Deionized water | Locator®/Blue | 0 | 20.1 ± 2.87 | 20.58 ± 3.09 | −2.81 ± 4.07 |
| 10 | 22.94 ± 1.48 | 25.03 ± 2.59 | −8.93 ± 5.99 | ||||||||
| Locator®/Pink | 0 | 24.55 ± 2.14 | 37.42 ± 2.79 | −52.78 ± 7.78 | |||||||
| 10 | 47.13 ± 8.96 | 30.33 ± 4.18 | 34.77 ± 6.82 | ||||||||
| Locator®/White | 0 | 69.87 ± 5.73 | 42.56 ± 3.27 | 38.81 ± 5.32 | |||||||
| 10 | 56.86 ± 4.44 | 39.55 ± 2.95 | 30.19 ± 5.50 | ||||||||
| CM Loc®/Green (extralight) | 0 | 12.45 ± 1.27 | 12.83 ± 1.39 | −3.01 ± 1.92 | |||||||
| 10 | 12.11 ± 1.28 | 15.52 ± 1.41 | −30.52 ± 23.84 | ||||||||
| CM Loc/Red (light) | 0 | 17.79 ± 1.70 | 21.48 ± 1.20 | −22.31 ± 17.18 | |||||||
| 10 | 27.63 ± 2.28 | 26.74 ± 2.42 | 3.25 ± 2.69 | ||||||||
| CM Loc®/Green (medium) | 0 | 39.35 ± 3.45 | 36.99 ± 1.75 | 5.57 ± 5.28 | |||||||
| 10 | 46.96 ± 1.70 | 43.84 ± 1.99 | 6.55 ± 5.01 | ||||||||
| CM Loc®/Blue (high) | 0 | 44.07 ± 3.07 | 43.80 ± 2.45 | 0.52 ± 1.92 | |||||||
| 10 | 45.73 ± 4.29 | 39.44 ± 8.23 | 14.69 ± 11.22 | ||||||||
| Manie | 2 | 20 | Mesial | 1 | nc | Artificial saliva | CM Loc®/Red (light) | 0 | 81.8 ± 18.5 | 49.2 ± 12.6 | 39.85 |
| 10 | 47.1 ± 8.6 | 42.4 ± 13.5 | 9.99 | ||||||||
| 20 | 47.4 ± 8.9 | 35.4 ± 9.6 | 25.32 | ||||||||
| 30 | 44.4 ± 10.8 | 41.2 ± 14.8 | 7.21 | ||||||||
| Novaloc®/Yellow | 0 | 57.7 ± 31.0 | 59.4 ± 16.0 | −2.95 | |||||||
| 10 | 41.8 ± 14.8 | 54.4 ± 15.8 | −30.14 | ||||||||
| 20 | 48.9 ± 13.9 | 52.6 ± 8.4 | −7.57 | ||||||||
| 30 | 76.6 ± 44.5 | 47.7 ± 18.6 | −37.73 | ||||||||
| Locator R-Tx®/Pink (medium) | 0 | 75.5 ± 24.9 | 60.0 ± 19.6 | 20.53 | |||||||
| 10 | 66.3 ± 16.9 | 45.5 ± 11.9 | 31.37 | ||||||||
| 20 | 62.3 ± 22.1 | 33.6 ± 11.4 | 46.07 | ||||||||
| 30 | 78.6 ± 33.3 | 29.9 ± 18.1 | 61.96 | ||||||||
| Yilmaz et al., 2020 | 2 | nc | Distal | 1 | 50 | nc | Locator®/Pink | 0/0 | 13.6 | 10.0 | 26.5 |
| 0/30 | 18.3 | 12.8 | 30.0 | ||||||||
| 30/30 | 50.2 | 21.7 | 56.8 | ||||||||
| Locator R-Tx®/Pink | 0/0 | 20.1 | 14.0 | 30.3 | |||||||
| 0/30 | 17.5 | 10.6 | 39.4 | ||||||||
| 30/30 | 33.3 | 20.0 | 39.9 | ||||||||
| Yang et al., 2011 | 1 | nc | 1 | 60 | nc | Dal-ro®/Gold alloy strip | 0 | 6.48 ± 0.34 | |||
| 15 | 6.25 ± 0.2 | ||||||||||
| 30 | 5.76 ± 0.16 | ||||||||||
| 45 | 4.75 ± 0.92 | ||||||||||
| Locator®/Blue | 0 | 15.36 ± 1.4 | |||||||||
| 15 | 14.67 ± 0.74 | ||||||||||
| 30 | 13.3 ± 1.7 | ||||||||||
| 45 | 6.58 ± 0.34 | ||||||||||
| Sultana et al., 2017 | 2 | 22 | Distal | 10 | 50 | Dry condition | Ball Dentsply/plastic Red Clix® | 0 | 56.2 ± 6.12 | 46.0 ± 4.74 | 18.1 |
| 20 | 45.7 ± 8.03 | 40.7 ± 2.88 | 10.9 | ||||||||
| Locator®/Pink | 0 | 108.9 ± 29.78 | 20.2 ± 5.74 | 81.5 | |||||||
| Locator®/Green | 20 | 82.3 ± 14.15 | 17.3 ± 3.73 | 79.0 | |||||||
Effects of denture cleansing solutions on the initial retention (N) and the retention loss (%) of Locator® attachment system.
| In Vitro Studies | Cross-Head Speed | Immersion Time, Number and Type of Cycles | RD | Tap Water | Artificial Saliva | NaCl 0.9% | Cool Mint Listerine | NaOCl 6.15% | Aktident | Efferdent | Protefix | Corega | Polident | Polident Overnight |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Sodium Chloride | Menthol Derivatives | Sodium Hypochlorite | Sodium | Sodium Bicarbonate, Perborate, or Percarbonate, Sodium Carbonate, Potassium Monopersulfate, Citric Acid, EDTA… | ||||||||||
| Ayyıldız et al., 2020 | 2 | 12 M | Blue | 41.1 | 33.3 | 44.3 | 52.5 | |||||||
| Pink | 58.7 | 39.7 | 58.5 | 58.3 | ||||||||||
| White | 76.7 | 52.3 | 89.0 | 93.7 | ||||||||||
| Kürkcüoglu et al., 2016 | 50 | 6 M | Blue | 22.1 | 10.4 | 7.7 | 13.8 | |||||||
| Pink | 27.3 | 29.2 | 25.5 | 27.6 | ||||||||||
| White | 36.7 | 38.3 | 23.5 | 33.6 | ||||||||||
| Nguyen et al., 2010 | 50 | 6 M | Pink | 45.3 | 51.1 | 7.8 | 40.8 | 45.0 | 45.0 | |||||
| Srinivasan et al., 2016 | 120 | 10 IRC | Blue | 35.6 | 43.3 | |||||||||
| You et al., 2011 | 50 | 6 M | Pink | 22.2 | 22.3 | 12.6 | 21.5 | 21.8 | ||||||
* Statistically significant retention loss. Abbreviations: CC, chewing cycles; IRC, insertion–removal cycles; M, months; RD, retention device.