The in vitro effect of different implant angulations and cyclic dislodgement on the retentive properties of an overdenture attachment system.

Suhail Ali Al-Ghafli1, Konstantinos X Michalakis, Hiroshi Hirayama, Kiho Kang.   

Abstract

STATEMENT OF PROBLEM: The retentive capacity of ball attachments may be altered by a change in implant angulations.
PURPOSE: The purpose of this in vitro study was to investigate the effect of cyclic dislodgement on the retention of an overdenture attachment system when 2 implants were placed at angulations of 0, 5, 10, 15, and 20 degrees.
MATERIAL AND METHODS: Twelve acrylic resin blocks were fabricated and divided into 6 groups of 2 pairs each. In each of the 6 groups, 1 acrylic resin block was used to house the implants (block A), while the other (block B) was used to house the overdenture attachments. Two implants positioned at 0/0 degrees, with a standard plastic component(white) designed for 0-degree angulations, served as a control (CTRL), while the other 5 pairs of implants were placed in 5 different angulations: 0D: 0/0 degrees, 5D: 5/5 degrees, 10D: 10/10 degrees, 15D: 15/15 degrees, 20D: 20/20degrees (n=5). The extended range (green color) attachment was used for all groups except the control group. Implants(4.3 mm x 13 mm, internally hexed) were placed in blocks B. All angled implants were mesially tilted. Thirty pairs of attachments (Locator) were used. Dislodging cycles were applied to the overdenture attachment system. The initial retentive forces among the groups were not identical. The cycles required for the retentive forces of the attachments to decrease from the initial values to 60 N, and then to 40 N and 20 N, were recorded for standardization purposes.One-way ANOVA and Tukey HSD tests were used to analyze the difference in retention loss among the 6 groups(alpha=.05). A regression analysis (alpha=.05) was also performed to investigate the relationship between the implant angulation,the retentive force, and the logarithm of the number of cycles required for ball attachment retention decrease.
RESULTS: The 1-way ANOVA and the Tukey HSD tests revealed significant differences for the number of cycles required by different implant angulation groups for the initial retentive values to decrease to 60 N, 40 N, and 20 N (P<.001).The 0D and 5D groups required the longest time for retention loss, while 20D and CTRL groups demonstrated the shortest time for retention loss. The results of the regression analysis of the logarithmic number of cycles on retentive force and implant angulation demonstrated a significant effect (P<.001).
CONCLUSIONS: Implant angulations negatively affect attachment retention longevity. (J Prosthet Dent 2009;102:140-147)

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19703620     DOI: 10.1016/S0022-3913(09)60134-7

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  17 in total

1.  Stud attachments for the mandibular implant-retained overdentures: Prosthetic complications. A literature review.

Authors:  Elie E Daou
Journal:  Saudi Dent J       Date:  2013-02-06

2.  The influences of implant angulations in one and two directions on the retentive properties of overdenture attachments: an in vitro study.

Authors:  Parsa Atashrazm; Mohammad Hossein Dashti; Leila Zamani Alavijeh; Samira Azarmaeh; Sayena Mahdizadeh
Journal:  J Indian Prosthodont Soc       Date:  2013-04-12

Review 3.  Biomaterial aspects: A key factor in the longevity of implant overdenture attachment systems.

Authors:  Elie E Daou
Journal:  J Int Soc Prev Community Dent       Date:  2015 Jul-Aug

Review 4.  Rehabilitation with implant-supported overdentures in total edentulous patients: A review.

Authors:  Juan F Martínez-Lage-Azorín; Gustavo Segura-Andrés; Joan Faus-López; Rubén Agustín-Panadero
Journal:  J Clin Exp Dent       Date:  2013-12-01

5.  A prospective cross-over study to evaluate the effect of two different occlusal concepts on the masseter muscle activity in implant-retained mandibular overdentures.

Authors:  Ahmed M Abdelhamid; Kenda I Hanno; Mohamed H Imam
Journal:  Int J Implant Dent       Date:  2015-12-22

6.  A comparative evaluation of fatigue resistance of two different implant overdenture stud attachments with two different denture base materials: An in vitro study.

Authors:  Namrata Agrawal; Sumeet Jain; Deepak Agrawal
Journal:  J Indian Prosthodont Soc       Date:  2018 Jan-Mar

7.  Managing maxillary proclination with novel designed angulated implants.

Authors:  Swati Ahuja; Nicholas Egbert; Vinay Jain; David Cagna
Journal:  J Indian Prosthodont Soc       Date:  2017 Apr-Jun

8.  Effect of implant- and occlusal load location on stress distribution in Locator attachments of mandibular overdenture. A finite element study.

Authors:  Angel Alvarez-Arenal; Ignacio Gonzalez-Gonzalez; Hector deLlanos-Lanchares; Elena Martin-Fernandez; Aritza Brizuela-Velasco; Joseba Ellacuria-Echebarria
Journal:  J Adv Prosthodont       Date:  2017-10-16       Impact factor: 1.904

9.  Strains around distally inclined implants retaining mandibular overdentures with Locator attachments: an in vitro study.

Authors:  Moustafa Abdou Elsyad; Fathi Abo Setta; Ahmed Samir Khirallah
Journal:  J Adv Prosthodont       Date:  2016-04-21       Impact factor: 1.904

10.  Effect of abutment angulation in the retention and durability of three overdenture attachment systems: An in vitro study.

Authors:  Carlos Aroso; Antonio Sergio Silva; Raul Ustrell; Jose Manuel Mendes; Ana Cristina Braga; Esther Berastegui; Tomas Escuin
Journal:  J Adv Prosthodont       Date:  2016-02-23       Impact factor: 1.904

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.