The effect of denture cleansing solutions on the retention of pink locator attachments after multiple pulls: an in vitro study.

Wenguang You1, Radi Masri, Elaine Romberg, Carl F Driscoll, Tao You.   

Abstract

PURPOSE: The effect of denture cleansing solutions and multiple pulls on the retention of pink Locator patrices was studied.
MATERIALS AND METHODS: Five groups of pink Locator attachments (3.0 lb. Light Retention replacement patrix attachments; five in each group) were soaked for the equivalent of 6 months of clinical use in the following solutions: water (control), Efferdent, Polident Overnight, 6.15% sodium hypochlorite (NaOCL, 1:10 dilution), and Listerine mouthwash. A universal testing machine set at a 2 in/min crosshead speed was used to perform 548 pulls (548 cycles of insertion and removal). The reduction in load to dislodgement (retention) after the initial pull and the final pull and the percent reduction in retention after 6 months were compared between the groups using a one-way ANOVA followed by Tukey's Honestly Significant Difference (HSD) Test (α= 0.05).
RESULTS: Denture cleansing solutions significantly reduced the retentive values of pink Locator attachments after the initial pull (F = 17.435, p < 0.0001). The retentive values of Efferdent, Listerine, Polident Overnight, and water were significantly higher than the retentive value of the attachments soaked in NaOCl. After 6 months of simulated use (548 pulls), the four denture cleansing solutions had significant effects on the retentive values of pink Locator attachments (F = 5.855, p = 0.003). The retentive values for attachments soaked in NaOCl (7.29 ± 1.0 N) were significantly lower than those of attachments soaked in Listerine (15.82 ± 4.7 N) and in Polident Overnight (14.41 ± 3.6 N). These cleansing solutions also had a significant effect on the percentage of retention lost (F = 3.271, p = 0.032). The loss of retention in attachments soaked in Listerine (29 ± 9%) was significantly lower than attachments soaked in water (53 ± 12%). The loss of retention in attachments soaked in Efferdent was 49 ± 9%; in Polident Overnight, 34 ± 18%; and in NaOCl, 42 ± 11%. There was no significant difference in the percentage of retention loss between water, Efferdent, NaOCl, and Polident Overnight. There was also no significant difference in the percentage of retention loss between Efferdent, NaOCl, Polident Overnight, and Listerine.
CONCLUSION: NaOCl significantly decreased the retentive value of Locators. Therefore, it should not be routinely recommended for use as a denture cleanser. Listerine significantly increased the retention of the Locator attachments; however, it is premature to recommend Listerine for use as a denture cleanser.
© 2011 by The American College of Prosthodontists.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21777332     DOI: 10.1111/j.1532-849X.2011.00722.x

Source DB:  PubMed          Journal:  J Prosthodont        ISSN: 1059-941X            Impact factor:   2.752


  5 in total

1.  Stud attachments for the mandibular implant-retained overdentures: Prosthetic complications. A literature review.

Authors:  Elie E Daou
Journal:  Saudi Dent J       Date:  2013-02-06

2.  The Effects of Denture Cleansing Solutions on the Retention of Attachments of Implant Supported Overdentures.

Authors:  Reza Derafshi; Mina Mohaghegh; Maryam Saki; Anahita Safari; Mohammad Rabee Haghighi
Journal:  J Dent (Shiraz)       Date:  2015-03

Review 3.  Biomaterial aspects: A key factor in the longevity of implant overdenture attachment systems.

Authors:  Elie E Daou
Journal:  J Int Soc Prev Community Dent       Date:  2015 Jul-Aug

Review 4.  Evaluation of Retention, Wear, and Maintenance of Attachment Systems for Single- or Two-Implant-Retained Mandibular Overdentures: A Systematic Review.

Authors:  Raphaël Wakam; Aurélie Benoit; Kwamivi Bernardin Mawussi; Caroline Gorin
Journal:  Materials (Basel)       Date:  2022-03-04       Impact factor: 3.623

5.  The Effect of Denture Cleansing Solutions on the Retention of Precision Attachments: An In Vitro Study.

Authors:  Gonca Deste Gokay; Serhat Emre Ozkir; Thomas Gerhard Wolf; Gulsum Gokcimen; Nergiz Rona; Mehmet Bicer; Burak Yilmaz
Journal:  Int J Environ Res Public Health       Date:  2022-04-05       Impact factor: 3.390

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.