Sisi Chen1, Rongrong Gao1, Colm McAlinden1,2, Junming Ye1, Yiran Wang1, Min Chen1, Jinhai Huang1, Yong Sun3, A-Yong Yu1.
Abstract
PURPOSE: To compare central corneal thickness (CCT), aqueous depth (AQD), and anterior chamber depth (ACD) measurements using the swept-source (CASIA SS-1000, Tomey, Japan) and time-domain (Visante, Carl Zeiss Meditec, USA) anterior segment optical coherence tomographers (OCT) in normal eyes.Entities:
Year: 2020 PMID: 32953169 PMCID: PMC7487088 DOI: 10.1155/2020/9739878
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Intraobserver repeatability outcomes provided by SS-OCT and TD-OCT devices.
| Parameter | Observer | Mean ± SD |
| CoR | CoV (%) | ICC (95% CI) |
|
| ||||||
| CCT ( | SS-OCT | 535.59 ± 30.31 | 0.85 | 2.34 | 0.16 | 0.999 (0.999 to 0.999) |
| TD-OCT | 531.04 ± 30.46 | 2.22 | 6.16 | 0.42 | 0.995 (0.992 to 0.997) | |
| AQD (mm) | SS-OCT | 3.24 ± 0.27 | 0.02 | 0.05 | 0.61 | 0.995 (0.992 to 0.996) |
| TD-OCT | 3.23 ± 0.25 | 0.03 | 0.09 | 0.97 | 0.985 (0.977 to 0.990) | |
| ACD (mm) | SS-OCT | 3.78 ± 0.26 | 0.02 | 0.06 | 0.53 | 0.994 (0.991 to 0.996) |
| TD-OCT | 3.76 ± 0.25 | 0.03 | 0.09 | 0.83 | 0.984 (0.977 to 0.990) | |
SD = standard deviation; Sw = within-subject standard deviation; CoR = coefficient of repeatability (2.77 Sw); CoV = within-subject coefficient of variation; ICC = intraclass correlation coefficient; CCT = central corneal thickness; AQD = aqueous depth; ACD = anterior chamber depth.
The mean difference, paired t test (P value), and 95% LoA for differences between the SS-OCT and TD-OCT devices.
| Device pairings | Mean ± SD |
| 95% LoA |
|
| |||
| CCT ( | 4.55 ± 2.62 | <0.001 | −0.59 to 9.69 |
| AQD (mm) | 0.01 ± 0.05 | 0.111 | −0.10 to 0.12 |
| ACD (mm) | 0.02 ± 0.05 | 0.022 | −0.09 to 0.12 |
SD = standard deviation; CCT = central corneal thickness; AQD = aqueous depth; ACD = anterior chamber depth.
Figure 1Bland–Altman plot showing the agreement for CCT measurements between the CASIA SS-OCT and the Visante TD-OCT. The solid line indicates the mean difference (bias), and the dotted lines indicate 95% limits of agreement (LoA).
Figure 2Bland–Altman plot showing the agreement for AQD (corneal endothelium to the anterior lens) measurements between the CASIA SS-OCT and the Visante TD-OCT. The solid line indicates the mean difference (bias), and the dotted lines indicate 95% limits of agreement (LoA).
Figure 3Bland–Altman plot showing the agreement for ACD (from the corneal epithelium to the anterior lens) measurements between the CASIA SS-OCT and the Visante TD-OCT. The solid line indicates the mean difference (bias), and the dotted lines indicate 95% limits of agreement (LoA).