Literature DB >> 28079690

Noncontact Evaluation of Corneal Grafts: Swept-Source Fourier Domain OCT Versus High-Resolution Scheimpflug Imaging.

Eszter Szalai1, Gábor Németh, Ziad Hassan, László Módis.   

Abstract

PURPOSE: To study the reliability and repeatability of anterior segment measurements in patients who underwent penetrating keratoplasty with swept light source Fourier domain anterior segment optical coherence tomography (SS-OCT) in comparison with a high-resolution Scheimpflug camera.
METHODS: Keratometry (K) and elevation data of the anterior and posterior corneal surfaces, central corneal thickness, and anterior chamber depth were determined in 56 postkeratoplasty eyes (mean age: 59.6 ± 16.5 yrs) with SS-OCT (CASIA SS-1000; Tomey) followed by Scheimpflug photography (Pentacam High Resolution; Oculus). Examinations were also performed on 48 right eyes of 48 age-matched healthy subjects (mean age: 66.3 ± 7.0 yrs).
RESULTS: A statistically significant difference was detected between both tomographers in anterior flat K [1.7 ± 4.92 diopters (D)], posterior astigmatism (1.32 ± 5.7 D) and elevation (6.65 ± 14.07 μm), and central corneal thickness (27.39 ± 90.81 μm) and anterior chamber depth values (0.14 ± 0.66 mm) in patients who had undergone corneal transplantation (P < 0.05). In the postkeratoplasty group, repeatability of SS-OCT was higher for every anterior segment parameter in comparison with the Pentacam. In corneal grafts, the variability between 3 consecutive measurements on the same subject was remarkably lower for every parameter while using SS-OCT.
CONCLUSIONS: In patients who had undergone corneal transplantation, SS-OCT provided more reliable and repeatable anterior segment evaluations when compared with the rotating Scheimpflug system. Our findings suggested that the tested high-resolution swept-source OCT system is a comprehensive and precise diagnostic tool for postoperative follow-up of patients who have undergone penetrating keratoplasty.

Entities:  

Mesh:

Year:  2017        PMID: 28079690     DOI: 10.1097/ICO.0000000000001133

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  4 in total

1.  The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging.

Authors:  Yune Zhao; Ding Chen; Giacomo Savini; Qing Wang; Hongfang Zhang; Yili Jin; Benhao Song; Rui Ning; Jinhai Huang; Chenyang Mei
Journal:  Eye Vis (Lond)       Date:  2020-06-09

2.  Comparative analysis of two optical biometry devices: high wavelength swept source OCT versus partial coherence interferometry.

Authors:  Eszter Szalai; Adrienne Csutak
Journal:  Int Ophthalmol       Date:  2021-10-11       Impact factor: 2.029

3.  Corneal irregularity and visual function using anterior segment optical coherence tomography in TGFBI corneal dystrophy.

Authors:  Yuito Abe; Takashi Omoto; Kohdai Kitamoto; Tetsuya Toyono; Junko Yoshida; Ryo Asaoka; Satoru Yamagami; Takashi Miyai; Tomohiko Usui
Journal:  Sci Rep       Date:  2022-08-12       Impact factor: 4.996

4.  Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography.

Authors:  Sisi Chen; Rongrong Gao; Colm McAlinden; Junming Ye; Yiran Wang; Min Chen; Jinhai Huang; Yong Sun; A-Yong Yu
Journal:  J Ophthalmol       Date:  2020-09-04       Impact factor: 1.909

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.