Evaluation of the Lenstar LS 900 non-contact biometer.

L P J Cruysberg1, M Doors, F Verbakel, T T J M Berendschot, J De Brabander, R M M A Nuijts.   

Abstract

BACKGROUND/AIMS: Evaluation of the reproducibility of the Lenstar LS 900 non-contact biometer, and a comparison with the Visante anterior-segment optical coherence tomography (AS-OCT) and the IOLMaster.
METHODS: Central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), keratometry (K) values, corneal diameter (CD) and axial length (AL) were obtained in 38 healthy volunteers (76 eyes) to determine the reproducibility of the Lenstar. CCT, ACD, CD, K values and AL measurements measured with the Lenstar were compared with the AS-OCT and IOLMaster. Intraocular lens (IOL) power calculations were done to study the significance of the difference between AL measurements.
RESULTS: The reproducibility of the Lenstar was better than 0.9% for CCT, ACD, LT, K values and AL measurements. Although all correlations were highly significant (p<0.001), all comparisons showed a significant difference, except for the comparison of CD measurements using the Lenstar and IOLMaster (p = 0.175). The differences in IOL power calculations for an AL of 20, 25 and 30 mm with a mean difference between Lenstar and IOLMaster AL measurements of 0.03 mm, were 0.13 D, 0.10 D and 0.08 D, respectively.
CONCLUSION: The reproducibility of the Lenstar was excellent. Small but significant differences exist between the Lenstar, Visante OCT and the IOLMaster. Therefore, measurements of the Lenstar, AS-OCT and IOLMaster are not interchangeable. Despite the significant difference between AL measurements, there is no clinically significant difference in the IOL power calculation results.

Entities:  

Mesh:

Year:  2009        PMID: 19692383     DOI: 10.1136/bjo.2009.161729

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  40 in total

1.  Comparison of optical low-coherence reflectometry and applanation ultrasound biometry on intraocular lens power calculation.

Authors:  Mirjana Bjeloš Rončević; Mladen Bušić; Ivan Cima; Biljana Kuzmanović Elabjer; Damir Bosnar; Daliborka Miletić
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-09-18       Impact factor: 3.117

2.  Intraobserver and interobserver repeatability of ocular components measurement in cataract eyes using a new optical low coherence reflectometer.

Authors:  Mirjana Bjeloš Rončević; Mladen Bušić; Ivan Cima; Biljana Kuzmanović Elabjer; Damir Bosnar; Daliborka Miletić
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-10-28       Impact factor: 3.117

3.  Agreement analysis of LENSTAR with other techniques of biometry.

Authors:  S Jasvinder; T F Khang; K K S Sarinder; V P Loo; V Subrayan
Journal:  Eye (Lond)       Date:  2011-03-11       Impact factor: 3.775

4.  [Comparison of IOL-Master 500 vs. Lenstar LS900 concerning the calculation of target refraction: A retrospective analysis].

Authors:  M Stattin; C Zehetner; N E Bechrakis; L Speicher
Journal:  Ophthalmologe       Date:  2015-05       Impact factor: 1.059

5.  Agreement of keratometric readings measured using rotating Scheimpflug imaging, auto-refractokeratometer, and biograph in eyes with keratoconus.

Authors:  Meltem Guzin Altınel; Hasim Uslu
Journal:  Int Ophthalmol       Date:  2021-02-05       Impact factor: 2.031

6.  Assessment of eye length changes in accommodation using dynamic extended-depth OCT.

Authors:  Yu-Cherng Chang; Keke Liu; Carolina de Freitas; Alex Pham; Florence Cabot; Siobhan Williams; Ethan Adre; Giovanni Gregori; Marco Ruggeri; Sonia H Yoo; Arthur Ho; Jean-Marie Parel; Fabrice Manns
Journal:  Biomed Opt Express       Date:  2017-04-26       Impact factor: 3.732

7.  Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter.

Authors:  Hasan Ali Bayhan; Seray Aslan Bayhan; Izzet Can
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

8.  Precision of 5 different keratometry devices.

Authors:  Jonas Vejvad Nørskov Laursen; Peter Jeppesen; Thomas Olsen
Journal:  Int Ophthalmol       Date:  2015-03-29       Impact factor: 2.031

9.  Comparison of keratometric measurements obtained by the Verion Image Guided System with optical biometry and auto-keratorefractometer.

Authors:  Leyla Asena; Sirel Gür Güngör; Ahmet Akman
Journal:  Int Ophthalmol       Date:  2016-06-07       Impact factor: 2.031

10.  Scheimpflug camera combined with placido-disk corneal topography and optical biometry for intraocular lens power calculation.

Authors:  Ahmet Kirgiz; Kurşat Atalay; Havva Kaldirim; Kubra Serefoglu Cabuk; Mehmet Orcun Akdemir; Muhittin Taskapili
Journal:  Int Ophthalmol       Date:  2016-09-01       Impact factor: 2.031

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.