Comparison of Corneal Tomography: Repeatability, Precision, Misalignment, Mean Elevation, and Mean Pachymetry.

Simon Schröder1, Stephanie Mäurer1, Timo Eppig1, Berthold Seitz2, Kathrin Rubly1, Achim Langenbucher1.   

Abstract

PURPOSE: To compare corneal tomography and its statistical uncertainty for measurements obtained by three clinically used corneal tomographers: A Scheimpflug camera (Pentacam HR), a swept source optical coherence tomography system (CASIA SS-1000), and Placido ring imaging (TMS-5).
MATERIAL AND METHODS: Repeated measurements with all three devices on 34 normal eyes were used to estimate the repeatability, precision, and mean values of corneal elevation and pachymetry within 8 mm diameter. The repeatability (standard deviation) was calculated for each data point of the corneal elevation data-maps of anterior and posterior cornea as well as for the pachymetry data-maps. Uncertainty on the position of the eye at each measurement might contribute to the differences between elevation data-maps. To take this into account, we defined the precision as the standard deviation for the elevation data-maps of anterior and posterior cornea after correction of misalignment-effects (rotation, translation). The mean elevation and pachymetry data-maps were fitted with Zernike polynomials for interdevice-comparison.
RESULTS: Pentacam HR offered the best repeatability and precision for the anterior corneal elevation (<3 and <1.6 μm, respectively). CASIA SS-1000 offered good repeatability and precision with high resolution for posterior corneal elevation, and the best repeatability for pachymetry (<3 μm). TMS-5 measured anterior elevation with similar repeatability to CASIA SS-1000 (<6 μm). The data-maps of the three tomographers could not be used interchangeably. The largest differences were observed for pachymetry and posterior corneal elevation data-maps.
CONCLUSIONS: Misalignment limited the repeatability of TMS-5 and Pentacam HR, but had little influence on the repeatability of CASIA SS-1000.

Entities:  

Keywords:  Corneal shape; corneal diagnostics; corneal height; corneal thickness; interdevice comparison

Mesh:

Year:  2018        PMID: 29482368     DOI: 10.1080/02713683.2018.1441873

Source DB:  PubMed          Journal:  Curr Eye Res        ISSN: 0271-3683            Impact factor:   2.424


  12 in total

1.  Evaluation of morphological features: femtosecond-LASIK flap vs. SMILE cap, and the effects on corneal higher-order aberrations.

Authors:  Xiaoying He; Qin He; Minjie Yuan; Zhi Fang; Kai Zhang; Tiepei Zhu; Wei Han
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-09-27       Impact factor: 3.535

2.  Repeatability of OCT-Based versus Scheimpflug- and Reflection-Based Keratometry in Patients with Hyperosmolar and Normal Tear Film.

Authors:  Bjørn Gjerdrum; Kjell Gunnar Gundersen; Per Olof Lundmark; Bente Monica Aakre
Journal:  Clin Ophthalmol       Date:  2020-11-18

Review 3.  A review of imaging modalities for detecting early keratoconus.

Authors:  Xuemin Zhang; Saleha Z Munir; Syed A Sami Karim; Wuqaas M Munir
Journal:  Eye (Lond)       Date:  2020-07-16       Impact factor: 3.775

4.  Keratoconic eyes with stable corneal tomography could benefit more from custom intraocular lens design than normal eyes.

Authors:  Simon Schröder; Timo Eppig; Weidi Liu; Jens Schrecker; Achim Langenbucher
Journal:  Sci Rep       Date:  2019-03-05       Impact factor: 4.379

5.  Comparison of corneal elevation and pachymetry measurements made by two state of the art corneal tomographers with different measurement principles.

Authors:  Simon Schröder; Achim Langenbucher; Jens Schrecker
Journal:  PLoS One       Date:  2019-10-16       Impact factor: 3.240

6.  Refractive Precision of Ray Tracing IOL Calculations Based on OCT Data versus Traditional IOL Calculation Formulas Based on Reflectometry in Patients with a History of Laser Vision Correction for Myopia.

Authors:  Bjørn Gjerdrum; Kjell Gunnar Gundersen; Per Olof Lundmark; Bente Monica Aakre
Journal:  Clin Ophthalmol       Date:  2021-02-26

7.  Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography.

Authors:  Sisi Chen; Rongrong Gao; Colm McAlinden; Junming Ye; Yiran Wang; Min Chen; Jinhai Huang; Yong Sun; A-Yong Yu
Journal:  J Ophthalmol       Date:  2020-09-04       Impact factor: 1.909

8.  Reproducibility, and repeatability of corneal topography measured by Revo NX, Galilei G6 and Casia 2 in normal eyes.

Authors:  Adam Wylęgała; Robert Mazur; Bartłomiej Bolek; Edward Wylęgała
Journal:  PLoS One       Date:  2020-04-02       Impact factor: 3.240

9.  Comparison of radius of anterior lens surface curvature measurements in vivo using the anterior segment optical coherence tomography and Scheimpflug imaging.

Authors:  Zhenzhen Liu; Xiaoting Ruan; Wei Wang; Juzi Liu; Yijing Meng; Xiaoxun Gu; Jun Fu; Lixia Luo; Yizhi Liu
Journal:  Ann Transl Med       Date:  2020-03

10.  Effects of intracorneal ring segments implementation technique and design on corneal biomechanics and keratometry in a personalized computational analysis.

Authors:  Niksa Mohammadi Bagheri; Mahmoud Kadkhodaei; Shiva Pirhadi; Peiman Mosaddegh
Journal:  Sci Rep       Date:  2021-07-13       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.