Assessment and reproducibility of anterior chamber depth measurement with anterior segment optical coherence tomography compared with immersion ultrasonography.

Gabor Nemeth1, Attila Vajas, Alexis Tsorbatzoglou, Bence Kolozsvari, Laszlo Modis, Andras Berta.   

Abstract

PURPOSE: To measure anterior chamber depth (ACD) with an anterior segment optical coherence tomography (AS-OCT) and a standard ultrasonic (US) axial scan (A-scan) device using an immersion technique and to assess repeatability, reproducibility, and correlations of the measurements.
SETTING: Department of Ophthalmology, Medical Health and Science Center, University of Debrecen, Debrecen, Hungary.
METHODS: Sixty healthy eyes of 41 patients were enrolled in a study. The central ACD was measured 5 times with AS-OCT (Visante, Carl Zeiss Meditec) using its chamber tool and 5 times with a US A-scan device (UltraScan Imaging System, Alcon Laboratories) using an immersion method. The measurements were performed consecutively by 2 independent observers.
RESULTS: The mean ACD measured with AS-OCT was 3.12 mm +/- 0.33 (SD) by observer 1 and 3.11 +/- 0.33 mm by observer 2 (P = .78). The repeatability was 0.8% +/- 0.4% and 1.9% +/- 1.4%, respectively. The reproducibility was 0.23%. The reliability coefficient with AS-OCT was 99.6%. The mean ACD measured with immersion US A-scan was 2.98 +/- 0.33 mm by observer 1 and 2.95 +/- 0.34 mm by observer 2 (P = .68) . The repeatability was 6.4% +/- 3.8% by observer 1 and 8.5% +/- 4.9% by observer 2. The reproducibility was 0.88%. The reliability coefficient was 87.1% for US A-scan measurements. The difference between ACD values with AS-OCT and values with US A-scan was statistically significant (P = .02). The correlation (r) between AS-OCT and US A-scan was 0.732 (P<.0001) by observer 1 and 0.802 (P<.0001) by observer 2.
CONCLUSIONS: Anterior chamber measurements were significantly deeper with AS-OCT than with US immersion A-scan. Repeatability of ACD measurements was better with AS-OCT than with immersion US, and reproducibility was equal with the 2 methods.

Entities:  

Mesh:

Year:  2007        PMID: 17321395     DOI: 10.1016/j.jcrs.2006.11.013

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  15 in total

1.  Lens thickness assessment: anterior segment optical coherence tomography versus A-scan ultrasonography.

Authors:  Nikoo Hamzeh; Sasan Moghimi; Golshan Latifi; Massood Mohammadi; Nassim Khatibi; Shan C Lin
Journal:  Int J Ophthalmol       Date:  2015-12-18       Impact factor: 1.779

2.  Repeatability assessment of anterior segment biometric measurements under accommodative and nonaccommodative conditions using an anterior segment OCT.

Authors:  Noelia Martínez-Albert; Jose J Esteve-Taboada; Robert Montés-Micó
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-11-02       Impact factor: 3.117

3.  Repeatability and agreement of two anterior segment OCT in myopic patients before implantable collamer lenses implantation.

Authors:  Xi-Fang Zhang; Meng Li; Yan Shi; Xiu-Hua Wan; Huai-Zhou Wang
Journal:  Int J Ophthalmol       Date:  2020-04-18       Impact factor: 1.779

4.  Changes in anterior chamber configuration after cataract surgery as measured by anterior segment optical coherence tomography.

Authors:  Martha Kim; Ki Ho Park; Tae-Woo Kim; Dong Myung Kim
Journal:  Korean J Ophthalmol       Date:  2011-03-11

5.  Comparison of Ocular Morphological Parameters Related to Lens Position by Anterior Segment Optical Coherence Tomography and Ultrasound Biomicroscopy.

Authors:  Zhiying Yu; Fenglei Wang; Fang Dong; Na Li; Dabo Wang; Ling Wang
Journal:  Int J Clin Pract       Date:  2022-03-20       Impact factor: 3.149

Review 6.  Quantitative evaluation of anterior segment parameters in the era of imaging.

Authors:  Syril Dorairaj; Jeffrey M Liebmann; Robert Ritch
Journal:  Trans Am Ophthalmol Soc       Date:  2007

7.  Reproducibility and Agreement Between 2 Spectral Domain Optical Coherence Tomography Devices for Anterior Chamber Angle Measurements.

Authors:  Kenneth M Marion; Jyotsna Maram; Xiaojing Pan; Anna Dastiridou; ZhouYuan Zhang; Alex Ho; Brian A Francis; Srinivas R Sadda; Vikas Chopra
Journal:  J Glaucoma       Date:  2015-12       Impact factor: 2.503

8.  Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth.

Authors:  Jinhai Huang; Weicong Lu; Giacomo Savini; Hao Chen; Chengfang Wang; Xinxin Yu; Fangjun Bao; Qinmei Wang
Journal:  J Ophthalmol       Date:  2016-06-14       Impact factor: 1.909

9.  Application of 5-Fluorouracil-Polycaprolactone Sustained-Release Film in Ahmed Glaucoma Valve Implantation Inhibits Postoperative Bleb Scarring in Rabbit Eyes.

Authors:  Xiu-Zeng Bi; Wei-Hua Pan; Xin-Ping Yu; Zong-Ming Song; Zeng-Jin Ren; Min Sun; Cong-Hui Li; Kai-Hui Nan
Journal:  PLoS One       Date:  2015-11-18       Impact factor: 3.240

10.  Changes in axial length, central cornea thickness, and anterior chamber depth after rhegmatogenous retinal detachment repair.

Authors:  Chunmei Huang; Tonghe Zhang; Jian Liu; Qiang Ji; Ruili Tan
Journal:  BMC Ophthalmol       Date:  2016-07-25       Impact factor: 2.209

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.