Corneal thickness mapping by 3D swept-source anterior segment optical coherence tomography.

Alberto Neri1, Marianna Malori, Patrizia Scaroni, Rosachiara Leaci, Elisabetta Delfini, Claudio Macaluso.   

Abstract

PURPOSE: To assess accuracy and repeatability of central corneal thickness (CCT) measurements obtained by swept-source anterior segment optical coherence tomography (AS-OCT), spectral-domain retinal OCT with corneal module and ultrasound pachymetry (USP), and to assess repeatability of pachymetric mapping with AS-OCT.
METHODS: 50 healthy volunteers were recruited. A single, experienced operator analysed the right eye of each participant twice in the same session with AS-OCT ('corneal map' routine), retinal OCT and USP. CCT measurements were compared using repeated-measures analysis of variance, Bonferroni test, Pearson correlation and Bland-Altman plots. Repeatability of thickness maps and CCT measurements were assessed using Alpha of Cronbach, intraclass correlation coefficient (ICC) and coefficient of repeatability.
RESULTS: Mean CCT±SD was 540±28.9 μm for AS-OCT, 544±29.5 μm for retinal OCT and 549.3±31.7 μm for USP; the differences were statistically significant (p<0.01). CCT measurements obtained with the three instruments were highly correlated: r was 0.965 for AS-OCT/USP, 0.962 for retinal OCT/USP and 0.984 for AS-OCT/retinal OCT comparison. The repeatability of CCT measurements was higher for AS-OCT than for the other devices (p<0.001). Repeatability of pachymetric maps was excellent (ICC=0.999).
CONCLUSIONS: Pachymetric maps by swept-source AS-OCT showed excellent repeatability. CCT measurements obtained by AS-OCT, USP and retinal OCT were highly correlated although not identical.
© 2012 The Authors. Acta Ophthalmologica © 2012 Acta Ophthalmologica Scandinavica Foundation.

Entities:  

Mesh:

Year:  2012        PMID: 22682316     DOI: 10.1111/j.1755-3768.2012.02453.x

Source DB:  PubMed          Journal:  Acta Ophthalmol        ISSN: 1755-375X            Impact factor:   3.761


  8 in total

1.  Dynamic imaging of accommodation by swept-source anterior segment optical coherence tomography.

Authors:  Alberto Neri; Marco Ruggeri; Alessandra Protti; Rosachiara Leaci; Stefano A Gandolfi; Claudio Macaluso
Journal:  J Cataract Refract Surg       Date:  2015-02-20       Impact factor: 3.351

2.  Repeatability of corneal epithelial thickness measurements using Fourier-domain optical coherence tomography in normal and post-LASIK eyes.

Authors:  Xingxuan Jack Ma; Li Wang; Douglas D Koch
Journal:  Cornea       Date:  2013-12       Impact factor: 2.651

3.  The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging.

Authors:  Yune Zhao; Ding Chen; Giacomo Savini; Qing Wang; Hongfang Zhang; Yili Jin; Benhao Song; Rui Ning; Jinhai Huang; Chenyang Mei
Journal:  Eye Vis (Lond)       Date:  2020-06-09

4.  Comparison of corneal elevation and pachymetry measurements made by two state of the art corneal tomographers with different measurement principles.

Authors:  Simon Schröder; Achim Langenbucher; Jens Schrecker
Journal:  PLoS One       Date:  2019-10-16       Impact factor: 3.240

5.  Reliability of a New Swept-Source Optical Coherence Tomography Biometer in Healthy Children, Adults, and Cataract Patients.

Authors:  Jinhai Huang; Yuyu Zhao; Giacomo Savini; Guanshun Yu; Jinjin Yu; Zhongxing Chen; Ruixue Tu; Yune Zhao
Journal:  J Ophthalmol       Date:  2020-05-15       Impact factor: 1.909

6.  Comparative Study of Anterior Eye Segment Measurements with Spectral Swept-Source and Time-Domain Optical Coherence Tomography in Eyes with Corneal Dystrophies.

Authors:  Anna K Nowinska; Sławomir J Teper; Dominika A Janiszewska; Anita Lyssek-Boron; Dariusz Dobrowolski; Robert Koprowski; Edward Wylegala
Journal:  Biomed Res Int       Date:  2015-09-17       Impact factor: 3.411

7.  Comparison of Repeatability and Agreement between Swept-Source Optical Biometry and Dual-Scheimpflug Topography.

Authors:  Soyeon Jung; Hee Seung Chin; Na Rae Kim; Kang Won Lee; Ji Won Jung
Journal:  J Ophthalmol       Date:  2017-12-10       Impact factor: 1.909

8.  Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography.

Authors:  Sisi Chen; Rongrong Gao; Colm McAlinden; Junming Ye; Yiran Wang; Min Chen; Jinhai Huang; Yong Sun; A-Yong Yu
Journal:  J Ophthalmol       Date:  2020-09-04       Impact factor: 1.909

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.