Corneal thickness and elevation measurements using swept-source optical coherence tomography and slit scanning topography in normal and keratoconic eyes.

Vishal Jhanji1, Bingzhi Yang, Marco Yu, Cong Ye, Christopher K S Leung.   

Abstract

BACKGROUND: To compare corneal thickness and corneal elevation using swept source optical coherence tomography and slit scanning topography.
DESIGN: Prospective study. PARTICIPANTS: 41 normal and 46 keratoconus subjects.
METHODS: All eyes were imaged using swept source optical coherence tomography and slit scanning tomography during the same visit. Mean corneal thickness and best-fit sphere measurements were compared between the instruments. MAIN OUTCOME MEASURES: Agreement of measurements between swept source optical coherence tomography and scanning slit topography was analyzed. Intra-rater reproducibility coefficient and intraclass correlation coefficient were evaluated.
RESULTS: In normal eyes, central corneal thickness measured by swept source optical coherence tomography was thinner compared with slit scanning topography (p < 0.0001) and ultrasound pachymetry (p = < .0001). Ultrasound pachymetry readings had better 95% limits of agreement with swept source optical coherence tomography than slit scanning topography. In keratoconus eyes, central corneal thickness was thinner on swept source optical coherence tomography than slit scanning topography (p = 0.081) and ultrasound pachymetry (p = 0.001). There were significant differences between thinnest corneal thickness, and, anterior and posterior best-fit sphere measurements between both instruments (p < 0.05 for all). Overall, reproducibility coefficients and intraclass correlation coefficients were significantly better with swept source optical coherence tomography for measurement of central corneal thickness, anterior best-fit sphere and, posterior best-fit sphere (all p < 0.001).
CONCLUSIONS: Corneal thickness and elevation measurements were significantly different between swept source optical coherence tomography and slit scanning topography. With better reproducibility coefficients and intraclass correlation coefficients, swept source optical coherence tomography may provide a reliable alternative for measurement of corneal parameters.
© 2013 The Authors. Clinical and Experimental Ophthalmology © 2013 Royal Australian and New Zealand College of Ophthalmologists.

Entities:  

Keywords:  corneal topography; keratoconus; topography

Mesh:

Year:  2013        PMID: 23566209     DOI: 10.1111/ceo.12113

Source DB:  PubMed          Journal:  Clin Exp Ophthalmol        ISSN: 1442-6404            Impact factor:   4.207


  12 in total

1.  Measurements of anterior segment parameters using three different non-contact optical devices in keratoconus patients.

Authors:  Ahmet Taylan Yazıcı; Gökhan Pekel; Ercüment Bozkurt; Yusuf Yıldırım; Evre Pekel; Ahmet Demirok; Omer Faruk Yılmaz
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

2.  A novel Hartman Shack-based topography system: repeatability and agreement for corneal power with Scheimpflug+Placido topographer and rotating prism auto-keratorefractor.

Authors:  Gaurav Prakash; Dhruv Srivastava; Sounak Choudhuri
Journal:  Int Ophthalmol       Date:  2015-03-28       Impact factor: 2.031

3.  Topographic characteristics of keratoconus among a sample of Jordanian patients.

Authors:  Mohammed Ali Abu Ameerh; Nathalie Bussières; Ghada Ismail Hamad; Muawyah Dawoud Al Bdour
Journal:  Int J Ophthalmol       Date:  2014-08-18       Impact factor: 1.779

Review 4.  A review of imaging modalities for detecting early keratoconus.

Authors:  Xuemin Zhang; Saleha Z Munir; Syed A Sami Karim; Wuqaas M Munir
Journal:  Eye (Lond)       Date:  2020-07-16       Impact factor: 3.775

5.  Imaging shear stress distribution and evaluating the stress concentration factor of the human eye.

Authors:  S Joseph Antony
Journal:  Sci Rep       Date:  2015-03-10       Impact factor: 4.379

6.  Comparative Study of Anterior Eye Segment Measurements with Spectral Swept-Source and Time-Domain Optical Coherence Tomography in Eyes with Corneal Dystrophies.

Authors:  Anna K Nowinska; Sławomir J Teper; Dominika A Janiszewska; Anita Lyssek-Boron; Dariusz Dobrowolski; Robert Koprowski; Edward Wylegala
Journal:  Biomed Res Int       Date:  2015-09-17       Impact factor: 3.411

7.  Longitudinal Evaluation of Cornea With Swept-Source Optical Coherence Tomography and Scheimpflug Imaging Before and After Lasik.

Authors:  Tommy C Y Chan; Sayantan Biswas; Marco Yu; Vishal Jhanji
Journal:  Medicine (Baltimore)       Date:  2015-07       Impact factor: 1.889

8.  Comparability and repeatability of pachymetry in keratoconus using four noncontact techniques.

Authors:  Mukesh Kumar; Rohit Shetty; Chaitra Jayadev; Debarun Dutta
Journal:  Indian J Ophthalmol       Date:  2015-09       Impact factor: 1.848

9.  Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography.

Authors:  Sisi Chen; Rongrong Gao; Colm McAlinden; Junming Ye; Yiran Wang; Min Chen; Jinhai Huang; Yong Sun; A-Yong Yu
Journal:  J Ophthalmol       Date:  2020-09-04       Impact factor: 1.909

10.  Comparison of corneal thickness in patients with dry eye disease using the Pentacam rotating Scheimpflug camera and anterior segment optical coherence tomography.

Authors:  Keiichi Fujimoto; Takenori Inomata; Yuichi Okumura; Nanami Iwata; Kenta Fujio; Atsuko Eguchi; Ken Nagino; Hurramhon Shokirova; Maria Karasawa; Akira Murakami
Journal:  PLoS One       Date:  2020-02-03       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.