Literature DB >> 24694582

Anterior segment biometry using spectral-domain optical coherence tomography.

Florent Aptel, Christophe Chiquet, Aurélie Gimbert, Jean-Paul Romanet, Gilles Thuret, Philippe Gain, Nelly Campolmi.   

Abstract

PURPOSE: To compare the anterior chamber and anterior chamber angle measurements obtained with spectral-domain anterior segment optical coherence tomography (AS-OCT) and time-domain AS-OCT.
METHODS: The anterior chamber of healthy participants was imaged with spectral-domain AS-OCT (Casia SS-1000; Tomey, Nagoya, Japan) and time-domain AS-OCT (Visante; Carl Zeiss Meditec, Dublin, CA). Central corneal thickness (CCT), anterior chamber depth (ACD), angle opening distance at 500 and 750 μm (AOD 500 and AOD 750), trabecular iris space area at 500 and 750 μm (TISA 500 and TISA 750), and scleral spur angle were measured. The intraclass correlation coefficients (ICCs) were calculated. The Pearson correlation test was used to evaluate the correlation between the measurements and Bland-Altman plots to evaluate the agreement.
RESULTS: One hundred one eyes of 101 healthy participants were analyzed. Excellent repeatability was found with both devices for CCT, AOD 500, AOD 750, TISA 750, and scleral spur angle (ICC = 0.90 to 0.98 and 0.89 to 0.97, respectively) and excellent to moderate repeatability was found for TISA 500 (ICC = 0.68 to 0.97 and 0.70 to 0.93, respectively). For all parameters, Casia and Visante measurements were significantly correlated (r = 0.76 to 0.98; P < .02). ACD measured with the Casia was significantly larger (mean difference = 0.12 ± 0.08 mm; P < .0001), and the scleral spur angle measured with the Casia was significantly lower (mean difference = 4.85° ± 5.30°; P < .01). There were nonsignificant differences between the devices for the other parameters (P > .06).
CONCLUSIONS: Both Casia and Visante AS-OCT demonstrate high repeatability. Except for the ACD and scleral spur angles, the two devices show good agreement. Copyright 2014, SLACK Incorporated.

Entities:  

Mesh:

Year:  2014        PMID: 24694582     DOI: 10.3928/1081597X-20140326-01

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  19 in total

1.  Measurement of early changes in anterior chamber morphology after cataract extraction measured by anterior segment optical coherence tomography.

Authors:  Kozue Kasai; Genichiro Takahashi; Koichi Kumegawa; Murat Dogru
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-08-21       Impact factor: 3.117

2.  Correlation between Intraocular Pressure and Angle Configuration Measured by OCT: The Chinese American Eye Study.

Authors:  Benjamin Y Xu; Bruce Burkemper; Juan Pablo Lewinger; Xuejuan Jiang; Anmol A Pardeshi; Grace Richter; Mina Torres; Roberta McKean-Cowdin; Rohit Varma
Journal:  Ophthalmol Glaucoma       Date:  2018-09-29

3.  Anterior Segment Optical Coherence Tomography: Applications for Clinical Care and Scientific Research.

Authors:  Jing Shan; Charles DeBoer; Benjamin Y Xu
Journal:  Asia Pac J Ophthalmol (Phila)       Date:  2019-04-25

4.  Evaluation of anterior chamber parameters with spectral-domain optical coherence tomography.

Authors:  Isabel Pinilla Lozano; Carmen López de la Fuente; Francisco Segura; Elvira Orduna Hospital; Ana Sánchez-Cano
Journal:  Jpn J Ophthalmol       Date:  2017-11-29       Impact factor: 2.447

5.  Comparison of ultrasound biomicroscopy and spectral-domain anterior segment optical coherence tomography in evaluation of anterior segment after laser peripheral iridotomy.

Authors:  Xiao-Yun Ma; Dan Zhu; Jun Zou; Wen-Jie Zhang; Yi-Lin Cao
Journal:  Int J Ophthalmol       Date:  2016-03-18       Impact factor: 1.779

6.  Reproducibility and Agreement of Anterior Segment Parameter Measurements Obtained Using the CASIA2 and Spectralis OCT2 Optical Coherence Tomography Devices.

Authors:  Benjamin Y Xu; Derek D Mai; Rafaella C Penteado; Luke Saunders; Robert N Weinreb
Journal:  J Glaucoma       Date:  2017-11       Impact factor: 2.503

7.  The change of anterior segment parameters after cataract surgery in normal-tension glaucoma.

Authors:  Wonseok Lee; Hyoung Won Bae; Chan Yun Kim; Gong Je Seong
Journal:  Int J Ophthalmol       Date:  2017-08-18       Impact factor: 1.779

8.  Diurnal Variation of Optical Coherence Tomography Measurements of Static and Dynamic Anterior Segment Parameters.

Authors:  Benjamin Y Xu; Rafaella C Penteado; Robert N Weinreb
Journal:  J Glaucoma       Date:  2018-01       Impact factor: 2.503

9.  The effect of corneal irregularity on astigmatism measurement by automated versus ray tracing keratometry.

Authors:  Hyun Cheol Roh; Roy S Chuck; Jimmy K Lee; Choul Yong Park
Journal:  Medicine (Baltimore)       Date:  2015-04       Impact factor: 1.889

10.  Comparative Study of Anterior Eye Segment Measurements with Spectral Swept-Source and Time-Domain Optical Coherence Tomography in Eyes with Corneal Dystrophies.

Authors:  Anna K Nowinska; Sławomir J Teper; Dominika A Janiszewska; Anita Lyssek-Boron; Dariusz Dobrowolski; Robert Koprowski; Edward Wylegala
Journal:  Biomed Res Int       Date:  2015-09-17       Impact factor: 3.411