| Literature DB >> 30678658 |
Youngju An1, Eun-Kyoung Kang2, Hyojin Kim3, Min-Ji Kang4, Yong-Soo Byun4, Choun-Ki Joo5,6.
Abstract
BACKGROUND: To evaluate the accuracy of biometric measurements by a swept-source optical coherence tomography (SS-OCT) based biometry for intraocular lens (IOL) power calculation.Entities:
Keywords: Cataract; Intraocular lens power calculation; Partial coherence interferometry; Swept-source optical coherence tomography
Mesh:
Year: 2019 PMID: 30678658 PMCID: PMC6346505 DOI: 10.1186/s12886-019-1036-y
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Comparison of lens opacity
| LOCS ΙΙΙ | Group A (n = 128) | Group B (n = 36) | |
|---|---|---|---|
| Nuclear opalescence | 2.83 ± 0.64 (3.00) | 2.69 ± 1.09 (2.50) | 0.122 |
| Nuclear color | 2.80 ± 0.66 (3.00) | 2.64 ± 1.13 (2.00) | 0.103 |
| Cortical | 2.08 ± 1.41 (3.00) | 1.89 ± 0.90 (2.00) | 0.158 |
| Posterior subcapsular | 0.52 ± 1.03 (0.00) | 3.72 ± 1.06 (4.00) |
Values are presented as mean ± standard deviation (median)
Group A; PCI feasible, Group B; PCI unfeasible
LOCS ΙΙΙ = Lens Opacities Classification System ΙΙΙ
*p–value is for Mann–Whitney U test
Comparison of corneal power and axial length
| SS–OCT biometer | PCI | A–scan US | ||
|---|---|---|---|---|
| Group A (n = 128) | ||||
| Corneal power (D) | 44.14 ± 1.47 | 44.16 ± 1.50 | 44.04 ± 1.50‡ | 0.811* |
| Axial length (mm) | 24.56 ± 2.16 | 24.58 ± 2.22 | 24.49 ± 2.19 | 0.935* |
| Group B (n = 36) | ||||
| Corneal power (D) | 44.01 ± 1.92 | N/A | 44.03 ± 1.88‡ | 0.924† |
| Axial length (mm) | 25.18 ± 2.86 | N/A | 25.02 ± 2.82 | |
Values are presented as mean ± standard deviation
Group A; PCI feasible, Group B; PCI unfeasible
D diopters, SS–OCT swept–source optical coherence tomography, PCI partial coherence interferometry, US ultrasonography
*p–value is for Kruskal–Wallis test
†p–value is for Wilcoxon signed rank test
††Measurement using manual keratometry
Comparison of refractive outcomes of intraocular lens power calculation using the three devices in SRK/T formula
| Group A (n = 128) | Group B (n = 36) | ||||
|---|---|---|---|---|---|
| SS–OCT biometer | PCI | A–scan US | SS–OCT biometer | A–scan US | |
| Optimized constant | 118.02 | 118.04 | 117.90 | 118.02 | 117.90 |
| PE (D) | 0.00 ± 0.44 | 0.00 ± 0.46 | −0.02 ± 0.61 | 0.00 ± 0.47 | −0.01 ± 0.66 |
| MAE (D) | 0.36 ± 0.27† | 0.39 ± 0.30† | 0.47 ± 0.39 | 0.36 ± 0.31 | 0.64 ± 0.36 |
| 0.027* | |||||
| MedAE (D) | 0.31 | 0.32 | 0.40 | 0.32 | 0.60 |
| Eye within (%) | |||||
| ≤ ±0.25 D | 41.44 | 42.97 | 36.72 | 47.22 | 22.22 |
| ≤ ±0.50 D | 71.09 | 67.19 | 60.16 | 72.22 | 47.22 |
| ≤ ±1.00 D | 95.31 | 95.31 | 90.63 | 94.44 | 83.33 |
Values are presented as mean ± standard deviation
Group A; PCI feasible, Group B; PCI unfeasible
D diopters, PE prediction error, MAE mean absolute error, MedAE median absolute error, SS-OCT swept-source optical coherence tomography, PCI partial coherence interferometry, US ultrasonography
*p–value is for Kruskal–Wallis test
†Same letters indicate no statistical significance based on Bonferroni’s method
††p–value is for Wilcoxon signed rank test
Fig. 1The distribution of absolute prediction error using A-scan US, PCI, and SS–OCT biometer in SRK/T formulas. Group A; PCI feasible, Group B; PCI unfeasible (US = Ultrasonography, PCI = Partial coherence interferometry, SS–OCT = swept-source optical coherence tomography, NS = not significant) *p–value is for Mann–Whitney U test.