Giacomo Savini1, Kenneth J Hoffer2,3, Piero Barboni4,5, Nicole Balducci4, Domenico Schiano-Lomoriello1, Pietro Ducoli1.
Abstract
PURPOSE: To investigate the accuracy of a new optical biometer for intraocular lens (IOL) power calculation in eyes undergoing cataract surgery.Entities:
Mesh:
Year: 2017 PMID: 28231267 PMCID: PMC5322924 DOI: 10.1371/journal.pone.0172634
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
| Optical biometry + corneal topography | Ultrasound immersion biometry + corneal topography | Paired t-test | |
|---|---|---|---|
| Corneal power | 43.50 ±1.25 D (range: 40.00–46.07) | 43.41 ±1.27 D (range: 39.80–46.37) | P = 0.0017 |
| Axial length | 24.04 ±1.29 mm (range: 21.09–30.17) | 23.94 ±1.27 mm (range: 20.96–29.84) | P < 0.0001 |
| Short eyes (axial length <22 mm) | 2 (2.7%) | 2 (2.7%) | N/A |
| Long eyes (axial length <26 mm) | 5 (6.6%) | 4 (5.4%) | N/A |
| Target refraction | -0.47 ±1.07 D (range: -3–0) | ||
| Implanted IOL power | 20.66 ±3.25 D (range: 10–35) | ||
Clinical features of the study eyes. D = diopters.
| Hoffer Q (Aladdin) | Hoffer Q (SimK+US) | Holladay 1 (Aladdin) | Holladay 1 (SimK+US) | SRK/T (Aladdin) | SRK/T (SimK+US) | |
|---|---|---|---|---|---|---|
| Mean optimized constant | 5.69 | 5.45 | 1.91 | 1.68 | 119.11 | 118.77 |
| Mean PE | 0.02 ±0.33 D | 0.01 ±0.37 D | 0.02 ±0.34 D | 0.01 ±0.36 D | 0.00 ±0.36 D | -0.01 ±0.39 D |
| MedAE | 0.25 D | 0.16 D | 0.25 D | 0.19 D | 0.25 D | 0.19 D |
| MAE (±SD) | 0.27 ±0.19 D | 0.26 ±0.25 D | 0.28 ±0.19 D | 0.27 ±0.24 D | 0.28 ±0.23 D | 0.28 ±0.27 D |
| PE Variance | 0.04 D | 0.07 D | 0.04 D | 0.06 D | 0.05 D | 0.07 D |
| Eyes with PE within ±0.25 D | 50.68% | 71.23% | 56.16% | 65.75% | 53.42% | 61.64% |
| Eyes with PE within ±0.50 D | 89.04% | 86.30% | 87.67% | 87.67% | 87.67% | 86.30% |
| Eyes with PE within ±1.00 D | 100.00% | 97.26% | 100.00% | 98.63% | 98.63% | 95.89% |
Refractive outcome predictability of IOL power calculation using the Aladdin optical biometer. D = diopters; MedAE = median absolute error; MAE = mean absolute error; SD = standard deviation; PE = prediction error.
Fig 1Mean absolute prediction error by the Hoffer Q, Holladay 1 and SRK/T formulas using the Aladdin measurements.
| Instrument | IOL model | Sample size | Formula | MedAE (D) | MAE (D) | PE ≤ 0.50 D | |
|---|---|---|---|---|---|---|---|
| Present study | Aladdin | Acrysof SN60WF | 73 | Hoffer Q | 0.25 | 0.27 | 89.04% |
| Cooke 2016 | IOLMaster | Acrysof SN60WF | 1079 | Barrett | 0.25 | 0.31 | 80.6% |
| Hoffer 2010 | IOLMaster | Acrysof SN60WF | 50 | Haigis | N/A | 0.46 | 56% |
| Olsen | IOLMaster | AcrySof MA60AC | 461 | Olsen | N/A | 0.43 | 62.5% |
| Reitblat 2015 | IOLMaster | Restor SN6AD1 | 73 | SRK/T | N/A | 0.24 | 91.78% |
| Srivannaboon 2013 | IOLMaster | PY-60AD (Hoya) | 163 | Hoffer Q | N/A | 0.32 | 68% |
| Cooke 2016 | Lenstar | Acrysof SN60WF | 1079 | Olsen | 0.22 | 0.28 | 83.7% |
| Hoffer 2010 | Lenstar | Acrysof SN60WF | 50 | Haigis | N/A | 0.45 | 58% |
| Hoffmann 2013 | Lenstar | 3 IOL models | 308 | Holladay 1 | 0.26 | 0.31 | 79.2% |
| Hoffmann 2013 | Lenstar | Acrysof SN60WF | 82 | Holladay 1 | 0.25 | 0.31 | N/A |
| Reitblat 2015 | Lenstar | Restor SN6AD1 | 73 | SRK/T | N/A | 0.23 | 91.78% |
Accuracy of biometric measurements for intraocular lens power calculation with different optical biometers and formulas (only the results of the best formula are shown).
* Acrysof SN60WF (Alcon), iMics1 (Bausch + Lomb), Tecnis (Abbott Medical Optics).
** the percentage refers only to eyes with an axial length between 22 and 24.5mm.