Sara González-Godínez1, Roxana Saucedo-Urdapilleta2, Mariana Mayorquín-Ruiz2, Cecilio Velasco-Barona2, Eduardo Moragrega-Adame2, Irving A Domínguez-Varela3, Roberto Gonzalez-Salinas4.
Abstract
PURPOSE: To assess the axial length (AL) measurement failure rate using partial-coherence interferometry (PCI) and swept-source optical coherence tomography (SS-OCT) in dense cataracts. As a secondary outcome, the SS-OCT biometry was compared to immersion ultrasound.Entities:
Keywords: Dense cataracts; IOL calculation; eye; immersion ultrasound biometry; partial-coherence interferometry; swept-source OCT
Mesh:
Year: 2022 PMID: 34937218 PMCID: PMC8917608 DOI: 10.4103/ijo.IJO_854_21
Source DB: PubMed Journal: Indian J Ophthalmol ISSN: 0301-4738 Impact factor: 1.848
Failure rate of axial length measurement in dense cataracts with PCI and SS-OCT optic biometers
| LOCS III | Number of eyes | PCI failure rate (%) | SS - OCT failure rate (%) | Obtained with PCI | Obtained with SS-OCT |
|
|---|---|---|---|---|---|---|
| NO4 | 13 | 30.77 | 0 | 9 (69.23) | 13 (100) | 0.079 |
| NO5 | 2 | 100 | 0 | 0 | 2 (100) | 0.298 |
| NO6 | 7 | 100 | 57.14 | 0 | 3 (42.86) | 0.193 |
| C4 | 3 | 33.33 | 0 | 2 (66.6) | 3 (100) | 0.999 |
| C5 | 1 | 0 | 0 | 1 (100) | 1 (100) | 0.857 |
| P3 | 9 | 33.33 | 0 | 6 (66.67) | 9 (100) | 0.326 |
| P4 | 9 | 88.89 | 11.11 | 1 (11.11) | 8 (88.89) | 0.352 |
| P5 | 6 | 83.33 | 16.66 | 1 (16.67) | 6 (100) | 0.517 |
| N + P | 8 | 75 | 12.5 | 2 (25.0) | 7 (87.5) | - |
| C + P | 5 | 100 | 40 | 0 | 3 (60.0) | - |
| Intumescent | 7 | 100 | 100 | 0 | 0 | - |
NO=Nuclear opalescence, C=Cortical, NO=Nuclear Opalescence, P=Posterior subcapsular, PCI=partial coherence interferometry, SS-OCT=swept-source OCT. *Chi-square test
Correlation between SS-OCT-based biometry and Immersion ultrasound
| Number of eyes | SS-OCT | Immersion US | Linear regression ( | Correlation ( | * | |
|---|---|---|---|---|---|---|
| ACD | 48 | 3.14±0.44 | 2.98±0.48 | 0.55 | 0.741 | 0.0017 |
| LT | 48 | 4.31±0.51 | 4.59±0.73 | 0.47 | 0.685 | 0.0007 |
| AL | 33 | 24.00±1.75 | 23.91±0.31 | 0.95 | 0.974 | 0.0056 |
SS-OCT=swept-source OCT, US=immersion ultrasound, ACD=Anterior chamber depth, LT=Lens thickness, AL=axial length. *Wilcoxon match-pairs signed-rank test
Figure 1Bland–Altman plot showing the level of agreement for (a) ACD, (b) LT, and (c) AL measurements between the SS OCT optical biometer and A scan immersion ultrasound