Comparison of IOL-Master 700 and IOL-Master 500 biometers in ocular biological parameters of adolescents.

Qiang Shi1, Guang-Yan Wang1, Yu-Hong Cheng1, Cheng Pei1.   

Abstract

AIM: To compare the differences and consistency of IOL-Master 700 biometers applying swept optical coherence tomography with the conventional IOL-Master 500 applying partial coherence interference in terms of the ocular biological parameters in adolescents with ametropia.
METHODS: A total of 110 adolescents (110 eyes) with ametropia were collected, including 55 males and 55 females; age 10.69±2.81y. Ocular biological measurements were taken by IOL-Master 700 and IOL-Master 500 respectively to obtain biological parameters including axial length (AL), mean corneal anterior surface keratometry (Km), anterior chamber depth (ACD), and horizontal corneal diameter (WTW). Paired t-test was used to compare the differences between the two instruments. The intra-group correlation coefficient (ICC) and the Bland-Altman analysis were used to evaluate the consistency of parameter measurements between the two instruments for the four biological parameters.
RESULTS: Statistical analysis showed that there was no significant difference in the Km value measured by IOL-Master 700 and IOL-Master 500 (t=-1.644, P=0.116). The average differences of the AL, ACD, and WTW distances between the two instruments are 0.028, 0.101 and 0.064 mm respectively, and the differences are statistically significant (t=2.644, 12.505, 3.911, P<0.001). The consistency study results indicated high correlation in the measurement of AL, Km, ACD and WTW between the two instruments (ICC=0.994, 0.873, 0.927, 0.912).
CONCLUSION: The novel biometric instrument IOL-Master 700 makes no difference with IOL-Master 500 in the measurement of Km. There are some differences in the values of AL, ACD, and WTW. However, the two instruments show good consistency in these four biological measurements. The measured values of Km are interchangeable between the instruments. These two types of biometrics can be used as mutual reference in consideration of that the differences in AL, ACD, and WTW measurements are not sufficient to produce clinically meaningful differences. International Journal of Ophthalmology Press.

Entities:  

Keywords:  IOL-Master 500; IOL-Master 700; adolescents; axial length; myopia

Year:  2021        PMID: 34282385      PMCID: PMC8243180          DOI: 10.18240/ijo.2021.07.08

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  21 in total

1.  Evaluation of total corneal power measurements with a new optical biometer.

Authors:  Mehdi Shajari; Ruven Sonntag; Michaela Ramsauer; Thomas Kreutzer; Efstathios Vounotrypidis; Thomas Kohnen; Siegfried Priglinger; Wolfgang J Mayer
Journal:  J Cataract Refract Surg       Date:  2020-05       Impact factor: 3.351

2.  The relationship of axial length to cycloplegic refraction and keratometry in amblyopic eyes of hyperopic children.

Authors:  Arif O Khan
Journal:  J AAPOS       Date:  2012-02       Impact factor: 1.220

3.  Biometry and intraocular lens power calculation results with a new optical biometry device: comparison with the gold standard.

Authors:  Godefroy Kaswin; Antoine Rousseau; Mohamed Mgarrech; Emmanuel Barreau; Marc Labetoulle
Journal:  J Cataract Refract Surg       Date:  2014-04       Impact factor: 3.351

4.  Reproducibility of optical biometry using partial coherence interferometry : intraobserver and interobserver reliability.

Authors:  A Vogel; H B Dick; F Krummenauer
Journal:  J Cataract Refract Surg       Date:  2001-12       Impact factor: 3.351

5.  Comparison of a new optical biometer using swept-source optical coherence tomography and a biometer using optical low-coherence reflectometry.

Authors:  Kenneth J Hoffer; Peter C Hoffmann; Giacomo Savini
Journal:  J Cataract Refract Surg       Date:  2016-08       Impact factor: 3.351

6.  Electronic devices and myopic refraction among children aged 6-14 years in urban areas of Tianjin, China.

Authors:  Shengxin Liu; Sheng Ye; Wei Xi; Xin Zhang
Journal:  Ophthalmic Physiol Opt       Date:  2019-05-17       Impact factor: 3.117

7.  Correlation between bulbar axis length and retinal ruptures in case of myopia eye.

Authors:  Emina H Alimanović
Journal:  Bosn J Basic Med Sci       Date:  2009-08       Impact factor: 3.363

8.  Multicenter study of optical low-coherence interferometry and partial-coherence interferometry optical biometers with patients from the United States and China.

Authors:  Kenneth J Hoffer; H John Shammas; Giacomo Savini; Jinhai Huang
Journal:  J Cataract Refract Surg       Date:  2016-01       Impact factor: 3.351

Review 9.  Myopia: Anatomic Changes and Consequences for Its Etiology.

Authors:  Jost B Jonas; Kyoko Ohno-Matsui; Songhomitra Panda-Jonas
Journal:  Asia Pac J Ophthalmol (Phila)       Date:  2019 Sep-Oct

10.  Comparison of Ocular Biometry and Refractive Outcomes Using IOL Master 500, IOL Master 700, and Lenstar LS900.

Authors:  Jae Shin Song; Do Yeh Yoon; Joon Young Hyon; Hyun Sun Jeon
Journal:  Korean J Ophthalmol       Date:  2020-04
View more
  6 in total

1.  Masticatory Muscle Thickness and Activity Correlates to Eyeball Length, Intraocular Pressure, Retinal and Choroidal Thickness in Healthy Women versus Women with Myopia.

Authors:  Grzegorz Zieliński; Marcin Wójcicki; Maria Rapa; Anna Matysik-Woźniak; Michał Baszczowski; Michał Ginszt; Monika Litko-Rola; Jacek Szkutnik; Ingrid Różyło-Kalinowska; Robert Rejdak; Piotr Gawda
Journal:  J Pers Med       Date:  2022-04-13

2.  Saving of Time Using a Software-Based versus a Manual Workflow for Toric Intraocular Lens Calculation and Implantation.

Authors:  Barbara S Brunner; Nikolaus Luft; Siegfried G Priglinger; Mehdi Shajari; Wolfgang J Mayer; Stefan Kassumeh
Journal:  J Clin Med       Date:  2022-05-20       Impact factor: 4.964

3.  Commentary: Efficacy of swept-source optical coherence tomography in dense cataract.

Authors:  Robin Malik; Atul Kumar Singh
Journal:  Indian J Ophthalmol       Date:  2022-01       Impact factor: 1.848

4.  The Axial Length of the Eyeball and Bioelectrical Activity of Masticatory and Neck Muscles: A Preliminary Report.

Authors:  Grzegorz Zieliński; Michał Baszczowski; Maria Rapa; Anna Matysik-Woźniak; Magdalena Zawadka; Jacek Szkutnik; Piotr Gawda; Robert Rejdak; Piotr Majcher; Michał Ginszt
Journal:  Pain Res Manag       Date:  2022-08-16       Impact factor: 2.667

5.  Agreement of axial length and anterior segment parameters measured with the MYAH device compared to Pentacam AXL and IOLMaster 700 in myopic children.

Authors:  Huri Sabur; Omer Takes
Journal:  Int Ophthalmol       Date:  2022-07-31       Impact factor: 2.029

6.  A comparison of IOLMaster 500 and IOLMaster 700 in the measurement of ocular biometric parameters in cataract patients.

Authors:  Jianhong Jiang; Xiaojing Pan; Mingming Zhou; Xiaoyun Wang; Hai Zhu; Dongfang Li
Journal:  Sci Rep       Date:  2022-07-27       Impact factor: 4.996

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.