Performance of three biometry devices in patients with different grades of age-related cataract.

Georgios Mylonas1, Stefan Sacu, Wolf Buehl, Markus Ritter, Michael Georgopoulos, Ursula Schmidt-Erfurth.   

Abstract

PURPOSE: The new Lenstar biometry device was compared in a typical clinical setting to the IOL-Master and Visante-OCT.
METHODS: Fifty-one eyes of 51 patients with age-related cataract were examined with Lenstar LS900 (Haag Streit AG) biometer, IOL-Master V.5 (Carl Zeiss Meditec AG) and Visante-OCT (Carl Zeiss Meditec AG) before cataract surgery. Central corneal thickness (CCT), anterior chamber depth (ACD), keratometry readings of flattest and steepest meridian (K), corneal radius (R) and axial length (AL) values were correlated. Cataracts were graded according to the Lens Opacities Classification System III (LOCS) regarding nuclear colour (NC), nuclear opalescence (NO), cortical (C) and posterior subcapsular (P) cataract.
RESULTS: Mean values and standard deviations for AL, K and R was 23.66 ± 1.23 mm and 23.67 ± 1.26 mm, 43.24 ± 1.69 dpt and 43.16 ± 1.71 dpt, 7.68 ± 0.29 mm and 7.70 ± 0.28 mm with the IOL-Master and with the LS900, respectively (r = 0.99 and p = 0.76, r = 0.99 and p = 0.029, r = 0.89 and p = 0.14, respectively). Visante-OCT demonstrated highest values of three devices regarding to ACD followed by Lenstar LS900 and IOLMaster. Axial length measurements were unfeasible in 10% of the cases (five patients) and this significantly correlated with the presence of posterior subcapsular cataract of LOCS III grade 4.0 or higher.
CONCLUSIONS: IOL-Master, Lenstar LS900 and AC–OCT proved to be excellent non-contact measurement methods in eyes with age-related cataract. Nevertheless, ultrasound biometry is still required for cases with dense posterior subcapsular cataract.
© 2011 The Authors. Acta Ophthalmologica © 2011 Acta Ophthalmologica Scandinavica Foundation.

Entities:  

Mesh:

Year:  2011        PMID: 21310011     DOI: 10.1111/j.1755-3768.2010.02042.x

Source DB:  PubMed          Journal:  Acta Ophthalmol        ISSN: 1755-375X            Impact factor:   3.761


  9 in total

1.  [Comparison of IOL-Master 500 vs. Lenstar LS900 concerning the calculation of target refraction: A retrospective analysis].

Authors:  M Stattin; C Zehetner; N E Bechrakis; L Speicher
Journal:  Ophthalmologe       Date:  2015-05       Impact factor: 1.059

Review 2.  Optimising biometry for best outcomes in cataract surgery.

Authors:  R Sheard
Journal:  Eye (Lond)       Date:  2013-12-06       Impact factor: 3.775

3.  Reproducibility of a long-range swept-source optical coherence tomography ocular biometry system and comparison with clinical biometers.

Authors:  Ireneusz Grulkowski; Jonathan J Liu; Jason Y Zhang; Benjamin Potsaid; Vijaysekhar Jayaraman; Alex E Cable; Jay S Duker; James G Fujimoto
Journal:  Ophthalmology       Date:  2013-06-04       Impact factor: 12.079

4.  Comparison of ocular biometry measurements by applanation and immersion A-scan techniques.

Authors:  Dupe S Ademola-Popoola; Donald A Nzeh; Sadiat E Saka; Lateefat B Olokoba; Tokunbo S Obajolowo
Journal:  J Curr Ophthalmol       Date:  2016-02-09

5.  Comparison of the biometric measurements calculated with Zeiss IOL-Master and WaveLight OB-820.

Authors:  Georgios Labiris; Panagiota Ntonti; Konstantinos Ntitsos; Andreas Katsanos; Haris Sideroudi; Vasilios Kozobolis
Journal:  Clin Ophthalmol       Date:  2017-04-21

6.  Meta-analysis of optical low-coherence reflectometry versus partial coherence interferometry biometry.

Authors:  Jinhai Huang; Colm McAlinden; Yingying Huang; Daizong Wen; Giacomo Savini; Ruixue Tu; Qinmei Wang
Journal:  Sci Rep       Date:  2017-02-24       Impact factor: 4.379

7.  Comparison of two swept-source optical coherence tomography biometers and a partial coherence interferometer.

Authors:  Chan Min Yang; Dong Hui Lim; Hyo Jeong Kim; Tae-Young Chung
Journal:  PLoS One       Date:  2019-10-11       Impact factor: 3.240

8.  The impact of cataract progression on accuracy of intraocular lens power measurement.

Authors:  Lin Leng; Honglei Li; Min Yin; Han Gao; Ting Shao; Keli Long
Journal:  PLoS One       Date:  2021-02-10       Impact factor: 3.240

Review 9.  Update on Intraocular Lens Formulas and Calculations.

Authors:  Tina Xia; Christine E Martinez; Linda M Tsai
Journal:  Asia Pac J Ophthalmol (Phila)       Date:  2020 May-Jun
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.