Enhanced Penetration for Axial Length Measurement of Eyes with Dense Cataracts Using Swept Source Optical Coherence Tomography: A Consecutive Observational Study.

Nino Hirnschall1, Ralph Varsits1, Birgit Doeller1, Oliver Findl2,3.   

Abstract

INTRODUCTION: The aim of this study was to find cases in which the axial eye length could not be measured with partial coherence interferometry (PCI) technology and to assess if it could be measured using swept source optical coherence tomography (ss-OCT) technology.
METHODS: All patients were measured at their pre-assessment visit 1 week prior to cataract surgery using conventional optical biometry (PCI technology, IOLMaster 500, Carl Zeiss Meditec AG, Jena, Germany). Patients in whom one or both eyes could not be measured using PCI technology were invited to participate in the study and to be measured with the ss-OCT (IOL Master 700, Carl Zeiss Meditec AG, Jena, Germany) device.
RESULTS: Altogether, 1226 eyes of 613 patients were measured consecutively, and 78 eyes were not measured successfully with PCI technology. Among those with unsuccessfully measured eyes, 23 patients were willing to participate in the study, and two of those were also unsuccessfully measured with the ss-OCT device (8.7%, 2/23). However, 91.3% (21/23) of the eyes that were unsuccessfully scanned with PCI technology were measurable with the ss-OCT device. The estimated overall rate of unsuccessful scans with the ss-OCT device was 0.5% (6/1226) ([Formula: see text] < 0.01).
CONCLUSION: ss-OCT technology significantly improves the rate of attainable axial eye length measurements, especially in eyes with posterior subcapsular cataracts, but also in eyes with dense nuclear cataracts, except for white cataracts.

Entities:  

Keywords:  Biometry; Cataract; Swept source optical coherence tomography

Year:  2018        PMID: 29498015      PMCID: PMC5997603          DOI: 10.1007/s40123-018-0122-1

Source DB:  PubMed          Journal:  Ophthalmol Ther


Introduction

The introduction of partial coherence interferometry (PCI) has significantly improved axial eye length measurements in comparison to ultrasound measurements [1, 2]. This is because the accuracy of PCI measurements is significantly higher than the accuracy of applanation ultrasound measurements, and because the results of PCI are less dependent on the examiner [3-5]. However, conventional PCI technology has one important disadvantage: in contrast to ultrasound, dense nuclear and posterior subcapsular cataracts may reduce the signal-to-noise ratio of the measurement, resulting in an unsuccessful measurement [6, 7]. Although the introduction of a composite scan method has reduced this problem, 4.7% of eyes still cannot be measured with PCI technology [8]. Recently, swept source optical coherence tomography (ss-OCT) was introduced for optical biometry [9]. Among the significant differences between PCI and ss-OCT technology are the wavelength and the measurement setup employed in each. The aim of this study was to find cases that could not be measured with PCI technology and to assess if they could be measured using ss-OCT technology instead.

Methods

This study included data from patients who underwent cataract surgery. The data were analyzed retrospectively. Patients with severe corneal scarring or head tremor were excluded from this study. All procedures involving human participants were performed in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards. Informed consent was obtained from all individual participants included in the study. All patients were measured at their pre-assessment visit one week prior to cataract surgery using conventional optical biometry (PCI technology, IOLMaster 500, Carl Zeiss Meditec AG, Jena, Germany). At least five scans were performed with PCI technology in all cases. If the SNR of the composite scan was <2.0, the scan was classified as unsuccessful. Patients in whom one or both eyes could not be measured were invited to participate in the study and to be measured with the ss-OCT (IOLMaster 700, Carl Zeiss Meditec AG) device. As this device was located at another location 40 min away by public transport, patients were asked to come for this additional appointment within the following three days. When successful, only one measurement (consisting of six scans) was performed with the ss-OCT (Fig. 1); otherwise, a maximum of five measurements (i.e., 30 scans in total) were performed to get one successful scan. In contrast to the PCI device, the ss-OCT device does not provide a composite scan or a signal-to-noise ratio. Therefore, each scan was analyzed separately and was defined as successful if the device was able to give a value for axial eye length. Regarding the protocol of the IOLMaster, a scan was only classified as successful if no warning was given by the device.
Fig. 1

Successful (top) and unsuccessful (bottom) longitudinal ss-OCT scans. The arrow depicts the macula

Successful (top) and unsuccessful (bottom) longitudinal ss-OCT scans. The arrow depicts the macula

Specification of the ss-OCT Device

The ss-OCT (IOLMaster 700) device uses a wavelength of 1055 nm and has a scan depth of 44 mm and a scan width of 6 mm. Its resolution in tissue is 22 µm and its measurement speed is 2000 A-scans per second.

Statistical Analysis

Excel 12.2.0 was used for descriptive statistical analysis. Frequencies are given as percentages and significance testing of proportions was performed using Fisher’s exact test. p < 0.05 was considered to indicate statistical significance.

Results

Altogether, 1226 eyes of 613 patients were measured consecutively on different measurement days using PCI technology in this study. Among those patients, 65 [78 eyes = 6.4% (78/1226)] were not measured successfully (i.e., 52 patients in whom one eye could not be measured successfully and 13 patients in whom neither eye could be measured successfully). Twenty-three of those 65 patients were willing to participate in the study. The 23 patients were on average 68.4 years of age (SD 14.8; 32–88 years). The female:male distribution was 11:12. Eight patients had a best corrected visual acuity of finger counting or less, 9 patients had a best corrected visual acuity ranging between finger counting and 20/200 Snellen, and 5 patients had a best corrected visual acuity of better than 20/200 Snellen. Mean axial eye length (measured with ss-OCT) was 24.07 mm (SD 1.22 mm; 22.06–26.71). Two of these 23 patients were measured unsuccessfully with ss-OCT technology (2/23; 8.7%). Table 1 presents the reasons for unsuccessful measurement with the PCI device and the ss-OCT device.
Table 1

Reasons for unsuccessful scans using the PCI device and the ss-OCT device

ReasonPCIss-OCT
Nuclear cataract102 (20%)
PSC90 (0%)
Nuclear cataract + PSC40 (0%)
Total232 (8.7%)
Reasons for unsuccessful scans using the PCI device and the ss-OCT device As 91.3% (21/23) of the unsuccessful scans using PCI technology were measurable with the ss-OCT device, the estimated rate of unsuccessful scans with the ss-OCT device was 0.5% (6/1226). This is a significant reduction ( < 0.01) of 92.2% (0.5/6.4) compared to PCI optical biometry.

Discussion

ss-OCT technology was shown to significantly increase the proportion of successful axial eye length measurements compared to PCI technology. In fact, all cases with posterior subcapsular cataract were measured successfully with ss-OCT technology. Twenty percent of the eyes with dense nuclear or white cataracts that were not measurable with PCI technology were also not measurable with the ss-OCT device. In total, the number of unsuccessful scans was reduced from 6.4% to 0.6% when the ss-OCT device was used. Srivannaboon et al. [10] compared the same two devices in a different study setting. They used a much smaller sample size, but performed a direct comparison by measuring all patients with both devices. The main focus of their study was data comparability and the reproducibility of the ss-OCT device, which were found to be good. Similar results were obtained by Shammas et al. [11], Kunert et al. [12], and Kurian et al. [13]. The proportion of unsuccessful scans with PCI technology in Srivannaboon’s study was slightly lower than the proportion in our study (5% vs. 6.4%), possibly due to differences in the study populations or their smaller sample size. Similar to our findings, they observed a significantly higher proportion of successful scans in the ss-OCT group compared to the PCI group. Akman et al. [14] reported a higher proportion of unsuccessful scans (9.0%), but the proportion of eyes with nuclear cataracts and posterior subcapsular cataracts that were unsuccessfully measured by the PCI method but were successfully measured using the ss-OCT device was almost the same as in the present work. The main reason for the higher success rate of ss-OCT technology is the higher wavelength used (1055 nm) compared to PCI technology (780 nm, IOLMaster), as a shorter wavelength results in a reduced penetration depth due to scattering [15]. This phenomenon is commonly known as Rayleigh scattering, which involves an inverse correlation between scattering and wavelength (to the fourth power). The advantageous sensitivity of ss-OCT devices compared to time-domain OCT (basically the sum of multiple PCI scans) was recognized more than 10 years ago, but the wavelength used at that time was higher (1300 nm) [16]. Additionally, the feasibility of using ss-OCT technology for biometric measurements of the human eye was proven more than 10 years ago. However, those devices were prototypes and used a different wavelength (1310 nm) to that employed by ss-OCT devices today [17]. One of the first uses of ss-OCT technology for axial eye length measurements was published in 2009 by Chong et al. [18]. They used a similar wavelength to that of the investigated device (1060 nm) and a tunable filter, but they used a much shorter coherence length (28 mm to measure an axial eye length 20 mm in physical size). They were able to measure pigs’ eyes successfully in their study. There is a good correlation between cataract density and Rayleigh scattering, and it should be noted that because the amount of scattering is inversely proportional to the fourth power of the wavelength, a longer wavelength is significantly less prone to Rayleigh scattering [19]. One drawback of this study is that not all of the eyes that were unsuccessfully measured with PCI technology were measured with ss-OCT technology because the ss-OCT device was situated in another clinic some distance away. The geographical distance to that clinic was an obstacle for several patients—possibly even more so for patients with very dense cataracts, as they tended to be older and/or to have reduced mobility.

Conclusions

In summary, ss-OCT technology significantly improves the rate of attainable axial eye length measurements, especially in eyes with posterior subcapsular cataracts, but also in eyes with dense nuclear cataracts, except for white cataracts. It is possible that the use of an even higher wavelength would increase the number of successful scans, but it may also reduce the scan resolution.
  19 in total

1.  Optical biometry in cataract surgery.

Authors:  Oliver Findl; Wolfgang Drexler; Rupert Menapace; Barbara Kiss; Christoph K Hitzenberger; Adolf F Fercher
Journal:  Dev Ophthalmol       Date:  2002

2.  Three-dimensional and high-speed swept-source optical coherence tomography for in vivo investigation of human anterior eye segments.

Authors:  Yoshiaki Yasuno; Violeta Dimitrova Madjarova; Shuichi Makita; Masahiro Akiba; Atsushi Morosawa; Changho Chong; Toru Sakai; Kin-Pui Chan; Masahide Itoh; Toyohiko Yatagai
Journal:  Opt Express       Date:  2005-12-26       Impact factor: 3.894

3.  Large coherence length swept source for axial length measurement of the eye.

Authors:  Changho Chong; Takuya Suzuki; Kohki Totsuka; Atsushi Morosawa; Tooru Sakai
Journal:  Appl Opt       Date:  2009-04-01       Impact factor: 1.980

4.  Three-dimensional optical coherence tomography at 1050 nm versus 800 nm in retinal pathologies: enhanced performance and choroidal penetration in cataract patients.

Authors:  Boris Povazay; Boris Hermann; Angelika Unterhuber; Bernd Hofer; Harald Sattmann; Florian Zeiler; James E Morgan; Christiane Falkner-Radler; Carl Glittenberg; Susanne Blinder; Wolfgang Drexler
Journal:  J Biomed Opt       Date:  2007 Jul-Aug       Impact factor: 3.170

5.  Repeatability and reproducibility of anterior ocular biometric measurements with 2-dimensional and 3-dimensional optical coherence tomography.

Authors:  Shinichi Fukuda; Keisuke Kawana; Yoshiaki Yasuno; Tetsuro Oshika
Journal:  J Cataract Refract Surg       Date:  2010-11       Impact factor: 3.351

6.  Biometry of cataractous eyes using partial coherence interferometry: clinical feasibility study of a commercial prototype I.

Authors:  Barbara Kiss; Oliver Findl; Rupert Menapace; Matthias Wirtitsch; Wolfgang Drexler; Christoph K Hitzenberger; Adolf F Fercher
Journal:  J Cataract Refract Surg       Date:  2002-02       Impact factor: 3.351

7.  Partial coherence interferometry: a novel approach to biometry in cataract surgery.

Authors:  W Drexler; O Findl; R Menapace; G Rainer; C Vass; C K Hitzenberger; A F Fercher
Journal:  Am J Ophthalmol       Date:  1998-10       Impact factor: 5.258

8.  Clinical comparison of a new swept-source optical coherence tomography-based optical biometer and a time-domain optical coherence tomography-based optical biometer.

Authors:  Sabong Srivannaboon; Chareenun Chirapapaisan; Pratuangsri Chonpimai; Siriwan Loket
Journal:  J Cataract Refract Surg       Date:  2015-10       Impact factor: 3.351

9.  Partial coherence laser interferometry vs conventional ultrasound biometry in intraocular lens power calculations.

Authors:  M S Rajan; I Keilhorn; J A Bell
Journal:  Eye (Lond)       Date:  2002-09       Impact factor: 3.775

10.  Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500.

Authors:  Ahmet Akman; Leyla Asena; Sirel Gür Güngör
Journal:  Br J Ophthalmol       Date:  2015-12-16       Impact factor: 4.638

View more
  15 in total

1.  Comparison of IOL-Master 700 and IOL-Master 500 biometers in ocular biological parameters of adolescents.

Authors:  Qiang Shi; Guang-Yan Wang; Yu-Hong Cheng; Cheng Pei
Journal:  Int J Ophthalmol       Date:  2021-07-18       Impact factor: 1.779

2.  Comparison of agreement and efficiency of a swept source-optical coherence tomography device and an optical low-coherence reflectometry device for biometry measurements during cataract evaluation.

Authors:  Samuel F Passi; Atalie C Thompson; Preeya K Gupta
Journal:  Clin Ophthalmol       Date:  2018-11-01

3.  Comparison of ocular biometric measurements in patients with cataract using three swept-source optical coherence tomography devices.

Authors:  Richul Oh; Joo Youn Oh; Hyuk Jin Choi; Mee Kum Kim; Chang Ho Yoon
Journal:  BMC Ophthalmol       Date:  2021-01-27       Impact factor: 2.209

4.  Ocular biometry in dense cataracts: Comparison of partial-coherence interferometry, swept-source optical coherence tomography and immersion ultrasound.

Authors:  Sara González-Godínez; Roxana Saucedo-Urdapilleta; Mariana Mayorquín-Ruiz; Cecilio Velasco-Barona; Eduardo Moragrega-Adame; Irving A Domínguez-Varela; Roberto Gonzalez-Salinas
Journal:  Indian J Ophthalmol       Date:  2022-01       Impact factor: 1.848

5.  Repeatability, reproducibility, and comparison of ocular biometry using a new optical coherence tomography-based system and another device.

Authors:  Adam Wylęgała; Bartłomiej Bolek; Robert Mazur; Edward Wylęgała
Journal:  Sci Rep       Date:  2020-09-02       Impact factor: 4.379

6.  Refractive surprise after cataract surgery secondary to smeared optics of swept-source optical coherence tomography biometer: a case report.

Authors:  Francis Carr; Vinod Gangwani
Journal:  BMC Ophthalmol       Date:  2020-08-28       Impact factor: 2.209

7.  Agreement of ocular biometric measurements in young healthy eyes between IOLMaster 700 and OA-2000.

Authors:  Xuan Liao; Yue Peng; Bo Liu; Qing-Qing Tan; Chang-Jun Lan
Journal:  Sci Rep       Date:  2020-02-21       Impact factor: 4.379

8.  Comparison of Ocular Biometry and Refractive Outcomes Using IOL Master 500, IOL Master 700, and Lenstar LS900.

Authors:  Jae Shin Song; Do Yeh Yoon; Joon Young Hyon; Hyun Sun Jeon
Journal:  Korean J Ophthalmol       Date:  2020-04

9.  Development and Verification of an Adjustment Factor for Determining the Axial Length Using Optical Biometry in Silicone Oil-Filled Eyes.

Authors:  Gerd U Auffarth; Tadas Naujokaitis; Louise Blöck; Anna Daghbashyan; Jan Meis; Victor A Augustin; Ramin Khoramnia; Timur M Yildirim
Journal:  Diagnostics (Basel)       Date:  2022-01-11

10.  Comparison Study of the Two Biometers Based on Swept-Source Optical Coherence Tomography Technology.

Authors:  Jing Dong; Jinhan Yao; Shuimiao Chang; Piotr Kanclerz; Ramin Khoramnia; Xiaogang Wang
Journal:  Diagnostics (Basel)       Date:  2022-02-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.