| Literature DB >> 19503280 |
Yoshiaki Yasuno, Violeta Dimitrova Madjarova, Shuichi Makita, Masahiro Akiba, Atsushi Morosawa, Changho Chong, Toru Sakai, Kin-Pui Chan, Masahide Itoh, Toyohiko Yatagai.
Abstract
A two- and three-dimensional swept source optical coherence tomography (SS-OCT) system, which uses a ready-to-ship scanning light source, is demonstrated. The light source has a center wavelength of 1.31 mum, -3 dB wavelength range of 110 nm, scanning rate of 20 KHz, and high linearity in frequency scanning. This paper presents a simple calibration method using a fringe analysis technique for spectral rescaling. This SS-OCT system is capable of realtime display of two-dimensional OCT and can obtain three-dimensional OCT with a measurement time of 2 s. In vivo human anterior eye segments are investigated two- and three-dimensionally. The system sensitivity is experimentally determined to be 114 dB. The three-dimensional OCT volumes reveal the structures of the anterior eye segments, which are difficult to observe in two-dimensional OCT images.Entities:
Year: 2005 PMID: 19503280 DOI: 10.1364/opex.13.010652