Jae Shin Song1, Do Yeh Yoon1,2, Joon Young Hyon1,2, Hyun Sun Jeon1,3.
Abstract
PURPOSE: To evaluate the agreement in ocular biometry outcomes measured by three different devices, the IOL Master 500, IOL Master 700, and Lenstar LS900, and compare the refractive outcomes after cataract surgery obtained using those three devices.Entities:
Keywords: Biometry; Cataract; Low coherence reflectometry; Swept source optical coherence tomography
Mesh:
Year: 2020 PMID: 32233146 PMCID: PMC7105785 DOI: 10.3341/kjo.2019.0102
Source DB: PubMed Journal: Korean J Ophthalmol ISSN: 1011-8942
Biometry measurements by IOL Master 500, IOL Master 700, and Lenstar LS900
Values are presented as mean ± standard deviation (p-value, compared with the other two devices in order of IOL Master 500, IOL Master 700 and Lenstar LS900).
AL = axial length; Km = mean keratometry; D = diopters; ACD = anterior chamber depth.
*Significantly different at p < 0.05.
Subgroup analysis for biometry measurements by axial length
Values are presented as mean ± standard deviation (p-value, compared with the other two devices in order of IOL Master 500, IOL Master 700, and Lenstar LS900).
*Significantly different at p < 0.05.
Subgroup analysis for biometry measurements by mean keratometry
Values are presented as mean ± standard deviation (p-value, compared to the other two devices in order of IOL Master 500, IOL Master 700, and Lenstar LS900).
D = diopters.
*Significantly different at p < 0.05.
Fig. 1A Bland-Altman plot showing the agreement of the ocular biometry measurements among three different devices. The middle solid line shows the mean difference, and the top and bottom dashed lines show the upper and lower 95% limits of agreement, respectively. (A–C) Axial length (AL), (D–F) mean keratometry (Km), and (G–I) anterior chamber depth (ACD) between (A,D,G) IOL Master 500 and IOL Master 700, (B,E,H) IOL Master 500 and Lenstar LS900, and (C,F,I) IOL Master 700 and Lenstar LS900. D = diopters.
Comparison of PEs among three distinct formulae of the built-in software of IOL Master 500, IOL Master 700, and Lenstar LS900
Values are presented as mean ± standard deviation if not otherwise indicated; Haigis formula for IOL Master 500, SRK/T for IOL Master 700, and Hill-RBF for Lenstar LS900 were used to calcuate.
PE = prediction error; D = diopters.