Mohammad-Reza Sedaghat1, Hamed Momeni-Moghaddam2,3, AbbasAli Yekta3,4,5, Ahmed Elsheikh6,7,8, Mehdi Khabazkhoob5,9, Renato Ambrósio10, Nasim Maddah3, Zeynab Danesh3.
Abstract
PURPOSE: To compare the biomechanically-corrected intraocular pressure (bIOP) measured by the Corvis ST (Oculus, Wetzlar, Germany) with IOP measurements made by other commonly used tonometers; and to test the correlations between IOP measures and central corneal thickness.Entities:
Keywords: corneal biomechanics; intraocular pressure; tonometry
Year: 2019 PMID: 31802962 PMCID: PMC6802558 DOI: 10.2147/OPTO.S220776
Source DB: PubMed Journal: Clin Optom (Auckl) ISSN: 1179-2752
Mean And Standard Deviation (SD) Of Measured IOP Using Different Tonometers (n=94 Eyes)
| Methods | Mean±SD (mmHg) | Range (mmHg) | Pairwise Comparisons |
|---|---|---|---|
| Icare (a) | 14.59±3.40 | 6.00–22.00 | a, d: 1.000 |
| GAT (b) | 13.08±2.64 | 8.00–20.00 | |
| Corvis, bIOP (c) | 16.11±1.66 | 13.00–23.40 | |
| ORA, IOPcc (d) | 15.01±2.55 | 7.80–20.90 |
Abbreviations: IOP, Intraocular pressure; 95% CI, 95% Confidence interval; GAT, Goldmann applanation tonometer; bIOP, Biomechanically-corrected IOP; ORA, Ocular response analyzer; IOPcc, Cornea-compensated IOP.
Figure 1Agreement of the bIOP with the IOP measured using Icare (upper), GAT (middle) and IOPcc (lower). (n=94 eyes) The mean difference is shown by a solid horizontal line, while the 95% limits of agreement, zero mean difference and the regression line are indicated by dashed horizontal and oblique lines, respectively.
Mean And Range Of Differences Between bIOP And Other IOP Measurements In mmHg, And 95% Limits Of Agreement (n= 94 Eyes)
| Methods | Mean Difference±SD | Range | 95% Limits Of Agreement | ICC | Spearman Correlation |
|---|---|---|---|---|---|
| Icare - bIOP | −1.51±2.95 | −9.30, 7.00 | −7.29, 4.27 | 0.507 | 0.482 |
| GAT - bIOP | −3.02±2.60 | −10.40, 5.00 | −8.11, 2.07 | 0.279 | 0.351 |
| IOPcc - bIOP | −1.09±1.96 | −6.20, 3.70 | −4.93, 2.75 | 0.686 | 0.557 |
Abbreviations: IOP, Intraocular pressure; 95% CI, 95% Confidence interval; GAT, Goldmann applanation tonometer; bIOP, Biomechanically-corrected IOP; IOPcc, cornea-compensated IOP; ICC, intra-class correlation coefficient.
Mean And Standard Deviation Of IOP (mmhg) In Small (<540 µm) And Large (≥540 µm) Central Corneal Thickness (n=94)
| CCT | <540 µm (n=54) | ≥540 µm (n=40) | P-value |
|---|---|---|---|
| Mean±SD | Mean±SD | ||
| Icare | 13.70±3.12 | 15.80±3.43 | 0.001 |
| GAT | 12.81±2.89 | 13.45±2.25 | 0.211 |
| Corvis, bIOP | 15.77±1.02 | 16.55±2.20 | 0.165 |
| ORA, IOPcc | 14.67±2.60 | 15.48±2.44 | 0.129* |
Note: *Parametric statistics.
Abbreviations: IOP, Intraocular pressure; GAT, Goldmann applanation tonometer; bIOP, Biomechanically-corrected IOP; ORA, Ocular response analyzer; IOPcc, cornea-compensated IOP; CCT, Central Corneal Thickness.