Repeatability, reproducibility, and age dependency of dynamic Scheimpflug-based pneumotonometer and its correlation with a dynamic bidirectional pneumotonometry device.

Sashia Bak-Nielsen1, Iben Bach Pedersen, Anders Ivarsen, Jesper Hjortdal.   

Abstract

PURPOSE: To assess the repeatability, reproducibility, and age dependency of dynamic Scheimpflug-based pneumotonometry (Corvis)-generated parameters and their correlation with dynamic bidirectional pneumotonometry device [Ocular Response Analyzer (ORA)] measurements [corneal hysteresis (CH) and corneal resistance factor (CRF)].
METHODS: Eyes of 29 healthy participants between 20 and 30 years old were measured using Corvis and ORA. Measurements were performed twice in a row and repeated after 1 week. Repeatability and reproducibility were examined by calculating intraday and interday coefficients of variation (CVs). Corvis measurements were compared with measurements from 19 participants older than 65 years.
RESULTS: The most repeatable and reproducible Corvis-generated parameters, with intraday and interday CVs < 10% were as follows: intraocular pressure (IOP), pachymetry, time to first and second applanation (A1 time, A2 time), radius and amplitude at highest concavity (HC radius and HC deformation amplitude, respectively), time to highest concavity (HC time), deflection length and amplitude at highest concavity [HC deflection length and HC deflection amplitude (millimeter), respectively], and time to highest deflection amplitude [HC deflection amplitude (millisecond)]. However, 7 of 19 analyzed parameters showed CVs > 10%. HC-related parameters showed significant age dependency. The correlation between Corvis and ORA was weak (A2 time and CH, r = 0.36, A1 time and CRF, r = 0.41). Corvis-derived IOP was on average 2.2 ± 2.3 mm Hg higher than cornea-compensated IOP, with large differences in patients with high-average IOP.
CONCLUSIONS: The considerable variation in repeatability and reproducibility should be taken into account when using the Corvis. Corvis parameters showed poor correlation with CH and CRF obtained by ORA measurements.

Entities:  

Mesh:

Year:  2015        PMID: 25393092     DOI: 10.1097/ICO.0000000000000293

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  17 in total

Review 1.  Biomechanical Diagnostics of the Cornea.

Authors:  Vinicius S De Stefano; William J Dupps
Journal:  Int Ophthalmol Clin       Date:  2017

2.  Assessment of corneal topographic, tomographic, densitometric, and biomechanical properties of Fabry patients with ocular manifestations.

Authors:  Veysel Cankurtaran; Kemal Tekin; Ayse Idil Cakmak; Merve Inanc; Faruk Hilmi Turgut
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2020-01-08       Impact factor: 3.117

3.  [Statistical analysis of correlated measurement data in ophthalmology : Tutorial for the application of the linear mixed model in SPSS and R using corneal biomechanical parameters].

Authors:  R Herber; A Kaiser; X Grählert; U Range; F Raiskup; L E Pillunat; E Spörl
Journal:  Ophthalmologe       Date:  2020-01       Impact factor: 1.059

4.  Measurement repeatability of the dynamic Scheimpflug analyzer.

Authors:  Atsuya Miki; Naoyuki Maeda; Tomoko Asai; Yasushi Ikuno; Kohji Nishida
Journal:  Jpn J Ophthalmol       Date:  2017-10-05       Impact factor: 2.447

5.  Dynamic curvature topography for evaluating the anterior corneal surface change with Corvis ST.

Authors:  Chunhong Ji; Jinhua Yu; Tianjie Li; Lei Tian; Yifei Huang; Yuanyuan Wang; Yongping Zheng
Journal:  Biomed Eng Online       Date:  2015-06-04       Impact factor: 2.819

6.  The Relationship between Corvis ST Tonometry Measured Corneal Parameters and Intraocular Pressure, Corneal Thickness and Corneal Curvature.

Authors:  Ryo Asaoka; Shunsuke Nakakura; Hitoshi Tabuchi; Hiroshi Murata; Yoshitaka Nakao; Noriko Ihara; Ulfah Rimayanti; Makoto Aihara; Yoshiaki Kiuchi
Journal:  PLoS One       Date:  2015-10-20       Impact factor: 3.240

7.  Comparison of Two Cap Thickness in Small Incision Lenticule Extraction: 100μm versus 160μm.

Authors:  Miao He; Wei Wang; Hui Ding; Xingwu Zhong
Journal:  PLoS One       Date:  2016-09-21       Impact factor: 3.240

8.  The Relationship of Body Mass Index and Blood Pressure with Corneal Biomechanical Parameters in Healthy Subjects.

Authors:  Mohamad-Reza Sedaghat; Farshad Askarizadeh; Mohsen Nematy; Foroozan Narooie-Noori; Javad Heravian; Tahereh Rakhshandadi; Sattar Rajabi
Journal:  Med Hypothesis Discov Innov Ophthalmol       Date:  2017

9.  Investigation of the effects of Islamic fasting on ocular parameters.

Authors:  Mohamad-Reza Sedaghat; Javad Heravian; Farshad Askarizadeh; Mahmoud Jabbarvand; Mohsen Nematy; Tahereh Rakhshandadi; Maryam-Sadat Amirkalali-Sijavandi; Fatemeh Shahsavan; Foroozan Narooie-Noori
Journal:  J Curr Ophthalmol       Date:  2017-08-18

Review 10.  Dynamic corneal deformation response and integrated corneal tomography.

Authors:  Marcella Q Salomão; Ana Luisa Hofling-Lima; Fernando Faria-Correia; Bernardo Teixeira Lopes; Sandra Rodrigues-Barros; Cynthia J Roberts; Renato Ambrósio
Journal:  Indian J Ophthalmol       Date:  2018-03       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.